Skip to content

Cook G00507Recalls & reported problems

Cook Incorporated · Fixed Core Wire Guide Safe-T-J, Curved · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002005075), received between Nov 20, 2019 and Jul 9, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Frayed” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 16, 2022Open, Classified

    Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

    Z-1833-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Frayed2
  • Material Protrusion/Extrusion1
  • Material Twisted/Bent1
  • Peeled/Delaminated1
  • Stretched1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 9, 2024Peeled/Delaminated; Material Twisted/Bent; Peeled/Delaminated; Material Twisted/Bent

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19700462 ↗
  • MalfunctionJun 14, 2023Stretched; Unraveled Material; Material Protrusion/Extrusion; Stretched; Unraveled Material; Material Protrusion/Extrusion

    UPON RETURN AND INITIAL EVALUATION OF THE DEVICE, THE WIRE WAS UNRAVELED, AND THE SAFETY WIRE AND MANDRIL PROTRUDED AND WERE EXPOSED. THE WIRE WAS STRETCHED AND ELONGATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17128501 ↗
  • MalfunctionNov 12, 2020Material Frayed; Material Frayed

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10828444 ↗
  • MalfunctionNov 20, 2019Material Frayed; Material Frayed

    ADDITIONAL INFORMATION RECEIVED 05DEC2019. AS REPORTED, THE DEVICE WAS USED WITH A COOK ULT 7.0 CATHETER. THE PROCEDURE PERFORMED WAS A PERCUTANEOUS TRANSHEPATIC CHOLANGIOGRAPHY DRAINAGE. THE PATIENT DID NOT HAVE ANY NOTED VESSEL TORTUOSITY OR CALCIFICATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9346902 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page