Skip to content

Cook G00655Recalls & reported problems

Cook Incorporated · Fixed Core Wire Guide Straight · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002006553), received between Sep 19, 2018 and Sep 27, 2018 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1
  • Entrapment of Device1
  • Material Twisted/Bent1
  • Stretched1

Patient problems

  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 27, 2018Break; Material Twisted/Bent

    IT WAS REPORTED THE END OF THE STRAIGHT EXCHANGE FIXED CORE WIRE GUIDE WAS LOOPED ON ITSELF WHEN REMOVED FROM THE PACKAGING, AND IT WAS LATER FOUND THAT THE DISTAL WELD CONNECTION HAD BROKEN. THE DEVICE DID NOT MAKE PATIENT CONTACT, AND THERE WERE NO INJURIES OR ADDITIONAL PROCEDURES REPORTED.

    Patient problems: No Patient Involvement

    MAUDE report 7914211 ↗
  • MalfunctionSep 19, 2018Entrapment of Device; Stretched; Detachment of Device or Device Component

    THE COMPLAINANT REPORTED THAT THE STRAIGHT EXCHANGE FIXED CORE WIRE GUIDE WAS STUCK INSIDE OF ANOTHER MANUFACTURER'S BALLOON DURING PATIENT CONTACT. THE WIRE DID NOT COME APART UNTIL IT WAS OUTSIDE OF THE PATIENT'S BODY. IT WAS FURTHER NOTED THAT THE COIL NEAR THE DISTAL END OF THE WIRE WAS ELONGATED, AND THE DISTAL WELD CONNECTION SEPARATED FROM THE SAFETY WIRE AND COIL TO RESULT IN COIL ELONGATION. THE PROCEDURE WAS COMPLETED WITH A NEW SHEATH, AND THE PRODUCT PROBLEM DID NOT RESULT IN ANY REPORTED SERIOUS INJURIES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7889750 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page