Skip to content

Rosen G01253Recalls & reported problems

Cook Incorporated · Rosen Curved Wire Guide · Class 2

Adverse event reports
21
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 00827002012530), received between Aug 28, 2017 and Apr 2, 2025 — classified as 3 injuries, 18 malfunctions.

The most frequently reported specific device problem is “Unraveled Material” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unraveled Material5
  • Adverse Event Without Identified Device or Use Problem3
  • Break3
  • Material Protrusion/Extrusion3
  • Material Frayed2
  • Contamination /Decontamination Problem1
  • Device Contamination with Chemical or Other Material1
  • Difficult to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient4
  • No Known Impact Or Consequence To Patient4
  • Insufficient Information3
  • No Patient Involvement3
  • Vascular Dissection3
  • Laceration(s)1
  • Renal Impairment1

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, DURING A PROCEDURE INVOLVING ANEURYSM REPAIR AND STENT GRAFT PLACEMENT, DURING WHICH A SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE (REPORTED UNDER PATIENT IDENTIFIER (B)(6) AND A SAFE-T-J AMPLATZ EXTRA-STIFF SUPPORT WIRE GUIDE (REPORTED UNDER PATIENT IDENTIFIER (B)(6) WERE USED, THE PATIENT EXPERIENCED A DISSECTION OF THE LEFT SUBCLAVIAN ARTERY. GENERAL ANESTHESIA WAS ADMINISTERED FOR THE PROCEDURE, AND PERCUTANEOUS ACCESS WAS OBTAINED IN THE LEFT AND RIGHT FEMORAL ARTERIES. CUT-DOWN ACCESS WAS OBTAINED IN THE RIGHT AXILLARY ARTERY. A 9X57 PERIPHERAL STENT WAS SUCCESSFULLY PLACED IN THE LEFT SUBCLAVIAN ARTERY, WHICH WAS RE-VASCULARIZED WITH A FENESTRATED BRANCHED DEVICE. STENT PATENCY WAS MAINTAINED, WITH NORMAL END ARTERY PERFUSION. THE WIRES USED DURING THE PROCEDURE SUCCESSFULLY ACCESSED THE TARGET VASCULATURE AND PERFORMED AS INTENDED, WITHOUT ANY NOTED DEVICE DEFICIENCIES. A LEFT SUBCLAVIAN ARTERY INJURY (DISSECTION) WAS REPORTED DURING THE PROCEDURE; HOWEVER, THE INJURY DID NOT REQUIRE TREATMENT. PER THE REPORTER, THE INJURY WAS CAUSALLY RELATED TO A STENT, AS THE DISSECTION OCCURRED AFTER PLACEMENT OF THE BRIDGING STENT, WHILE POSITIONING THE DEVICE. PER THE REP

    Patient problems: Vascular Dissection

    MAUDE report 21753761 ↗
  • InjuryApr 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, SIX DAYS AFTER AN UNSPECIFIED PROCEDURE, DURING WHICH FOUR SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDES WERE USED, A DISSECTION OF THE RIGHT RENAL ARTERY WAS NOTED, WHICH WAS TREATED CONSERVATIVELY. PER THE REPORTER, THE DISSECTION OCCURRED AFTER PLACEMENT OF BRIDGING STENTS AND WAS CONSIDERED POSSIBLY RELATED TO A COOK WIRE GUIDE, AS THE EVENT ¿COULD HAVE OCCURRED¿ DURING CANNULATION OF THE VESSEL. THE EVENT WAS CONSIDERED CAUSALLY RELATED TO THE PROCEDURE AND POSSIBLY RELATED TO THE TREATED DISEASE AND ¿PATHOLOGICAL VESSEL WALLS¿. PER THE USER/REPORTER, THE EVENT DID NOT OCCUR DUE TO A DEVICE DEFICIENCY. THE USER COULD NOT SPECIFY WHAT DEVICE WAS INVOLVED EXACTLY. THERE HAS BEEN NO REPORT THAT THE PATIENT REQUIRED ANY ADDITIONAL PROCEDURES OR EXPERIENCED ANY ADVERSE EFFECTS DUE TO THIS EVENT.

    Patient problems: Renal Impairment; Vascular Dissection

    MAUDE report 21738026 ↗
  • InjuryNov 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, A SIXTY-NINE-YEAR-OLD MALE WITH A TYPE IV THORACOABDOMINAL ANEURYSM UNDERWENT A PROCEDURE INVOLVING FENESTRATED ENDOVASCULAR AORTIC ANEURYSM REPAIR WITH STENT PLACEMENT WITHIN THE CELIAC ARTERY (CA), SUPERIOR MESENTERIC ARTERY (SMA), RIGHT RENAL ARTERY (RRA), AND LEFT RENAL ARTERY (LRA), ANGIOPLASTY OF THE LEFT EXTERNAL ILIAC ARTERY (EIA), AND INTRAVENOUS ULTRASOUND. THE RENAL ARTERIES WERE ¿RELATIVELY SMALL¿ IN DIAMETER. THE LENGTH FROM THE OSTIUM OF THE LEFT RENAL ARTERY TO THE FIRST MAJOR BIFURCATION WAS 41.2-MILLIMETERS, AND THE DIAMETER OF THE LRA AT THE LOCATION OF INTENDED DISTAL LANDING ZONE WAS 4.0-MILLIMETERS. THE LRA WAS LESS THAN 50% STENOSED, AND THERE WAS A POTENTIAL EXCESSIVE GAP AT THE LEVEL OF THE LRA. THE RIGHT AND LEFT COMMON AND EXTERNAL ILIAC ARTERIES WERE MODERATELY CALCIFIED, WITH MULTIPLE AREAS OF FOCAL STENOSIS. THE FEMORAL ARTERIES WERE MILDLY CALCIFIED. THE ILIAC ARTERIES WERE NOT TORTUOUS. THE ANGLE OF THE LEFT RENAL ARTERY WITH RESPECT TO THE AORTA WAS SIXTY-DEGREES. GENERAL ANESTHESIA WAS ADMINISTERED, AND ENDOTRACHEAL INTUBATION WAS PERFORMED. ANTIBIOTICS WERE ADMINISTERED. OVERLAY FLUOROSCOPY WAS OBTAINED TO PROVIDE ON-SCREEN MARKINGS DU

    Patient problems: Thrombosis/Thrombus; Vascular Dissection

    MAUDE report 20696544 ↗
  • MalfunctionAug 19, 2024Unraveled Material; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20000401 ↗
  • MalfunctionFeb 20, 2023Unraveled Material; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT. UPON RETURN AND EVALUATION OF THE COMPLAINT DEVICE, THE WIRE WAS NOT UNRAVELED AS INITIALLY REPORTED. THE WIRE WAS BENT IN A CORKSCREW SHAPE, BUT NO OTHER DAMAGE WAS NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16405656 ↗
  • MalfunctionOct 7, 2022Break; Break

    AS REPORTED, THE TIP OF A SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE WAS "BROKEN". ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 15564186 ↗
  • MalfunctionSep 9, 2022Break; Unraveled Material; Break; Unraveled Material

    AS REPORTED, DURING A THROMBECTOMY PROCEDURE FOR TREATMENT OF A DEEP VEIN THROMBOSIS, A SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE UNRAVELED. ACCESS WAS OBTAINED IN THE RIGHT INTERNAL JUGULAR VEIN AND THE WIRE WAS ADVANCED TO THE DISTAL VENA CAVA. THE PATIENT WAS FLIPPED FOR ACCESS INTO THE POPLITEAL VEIN, AS THE USER INTENDED TO ADVANCE THE WIRE THROUGH ANOTHER 16 FRENCH SHEATH. THE WIRE WAS SNARED AND EVERYTHING APPEARED "FINE"; HOWEVER, THE WIRE UNRAVELED IN THE MIDDLE OF THE FEMORAL VEIN BEFORE REACHING THE SHEATH. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED. A PHOTO PROVIDED BY THE CUSTOMER SHOWS THAT THE WIRE UNRAVELED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15390607 ↗
  • MalfunctionAug 17, 2022Break; Unraveled Material; Break; Unraveled Material

    AS REPORTED, DURING AN UNSPECIFIED PROCEDURE, THE TIP OF A SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE "FELL APART" AND SEPARATED FROM THE INNER WIRE WHILE THE USER ATTEMPTED TO STRAIGHTEN THE TIP. PER THE CUSTOMER, THIS OCCURRED ON THE BACK TABLE AFTER THE DEVICE WAS REMOVED FROM THE PATIENT. THE WIRE WAS NOT ALTERED PRIOR TO USE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15239576 ↗
  • MalfunctionApr 27, 2022Material Protrusion/Extrusion; Material Protrusion/Extrusion

    AS REPORTED, FOLLOWING AN UNSPECIFIED PROCEDURE, A SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE FRAYED. THE INNER MANDRIL PROTRUDED APPROXIMATELY 3-4CM FROM THE TIP OF THE DEVICE AND CUT THE FINGER OF THE PHYSICIAN HANDLING THE DEVICE. THE PHYSICIAN'S FINGER WAS REPORTED TO BE "FINE". A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL INTERVENTION DUE TO THIS OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. UPON RETURN EVALUATION OF THE DEVICE, THE WELD WAS BROKEN FREE OF THE SAFETY WIRE BUT WAS STILL ATTACHED TO THE COIL.

    Patient problems: Laceration(s)

    MAUDE report 14220466 ↗
  • MalfunctionJul 1, 2021Flaked; Flaked

    NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12103580 ↗
  • MalfunctionJan 20, 2021Unraveled Material; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11201017 ↗
  • MalfunctionJan 7, 2021Material Frayed; Stretched; Material Frayed; Stretched

    NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11133910 ↗
  • MalfunctionOct 26, 2020Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10733034 ↗
  • MalfunctionJun 6, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Patient Involvement

    MAUDE report 8677245 ↗
  • MalfunctionFeb 6, 2019Material Protrusion/Extrusion; Material Protrusion/Extrusion

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Patient Involvement

    MAUDE report 8311907 ↗
  • MalfunctionFeb 5, 2019Material Integrity Problem; Material Integrity Problem

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8306481 ↗
  • MalfunctionSep 28, 2018Material Protrusion/Extrusion

    IT WAS REPORTED THAT THE SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE HAD ITS INNER WIRE PROTRUDING FROM THE COIL UPON OPENING THE PACKAGE. THERE WERE NO INJURIES OR ADDITIONAL PROCEDURES REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7919524 ↗
  • MalfunctionSep 26, 2018Difficult to Advance; Material Twisted/Bent; Physical Resistance/Sticking

    THE COMPLAINANT REPORTED THAT THE SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE MET RESISTANCE DURING ADVANCEMENT. THE COILING WAS CURLED UP AND STICKING OUT OF THE WIRE, SO ANOTHER SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE WAS USED TO COMPLETE THE PROCEDURE. THE PRODUCT PROBLEM DID NOT RESULT IN ANY REPORTED SERIOUS INJURIES TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7909877 ↗
  • MalfunctionMay 10, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT A HAIR WAS VISUALIZED INSIDE THE PACKAGE OF THE UNOPENED SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE. THE DEVICE WAS NOT USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7502492 ↗
  • MalfunctionDec 21, 2017Fracture; Fracture

    IT WAS REPORTED THAT DURING A PROCEDURE, THE TIP OF THE SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE WAS ABOUT TO SNAP OFF. ADDITIONAL INFORMATION REGARDING EVENT DETAILS, PATIENT ANATOMY AND OUTCOME HAS BEEN REQUESTED, BUT IS NOT AVAILABLE AT THIS TIME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7140402 ↗
  • MalfunctionAug 28, 2017Material Frayed; Material Frayed

    A SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE WAS USED IN A TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVI), PERFORMED BY RIGHT FEMORAL APPROACH. IT WAS REPORTED, WHEN THE WIRE GUIDE REACHED THE TRICUSPID VALVE, UNDER FLUOROSCOPE, THE PHYSICIAN NOTED THE SHAPE OF THE TIP WAS ABNORMAL. THE WIRE GUIDE WAS REMOVED FROM THE PATIENT AND THE OUTER SPRING COIL WAS LOOSENED/FRAYED AT THE TIP. NO RESIDUES OF THE WIRE GUIDE REMAINED IN THE VESSEL. ANOTHER WIRE GUIDE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6824459 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.