Cook TPN G01880Recalls & reported problems
Cook Incorporated · Cook TPN, Double Lumen Catheter Set With 8Fr Peel-Away, Introducer Set · Class 2 · Implant
All models of Cook TPN (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00827002018808), received between Jul 27, 2016 and Jul 13, 2018 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Device Operates Differently Than Expected1
- Fitting Problem1
- Fracture1
- Leak/Splash1
- Material Frayed1
Patient problems
- No Known Impact Or Consequence To Patient6
- No Code Available1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 13, 2018Break; Break
IT WAS REPORTED A COOK TPN DOUBLE LUMEN TPN CATHETER SET BROKE WHERE THE HUB AND SILICONE CATHETER CONNECT. TO DATE, ADDITIONAL INFORMATION HAS BEEN REQUESTED THREE TIMES, BUT NO ADDITIONAL INFORMATION IS AVAILABLE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7686216 ↗MalfunctionSep 26, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
THE INTERNATIONAL CUSTOMER REPORTED THAT THE CLAMP ON THE COOK TPN DOUBLE LUMEN TPN CATHETER WAS OPEN, AND WHEN THE OPERATOR ATTEMPTED TO CLOSE THE CLAMP, IT WOULD SPONTANEOUSLY REOPEN AND NOT STAY CLOSED. THE CLAMP HAD TO BE "OVER-FLEXED" IN ORDER TO KEEP IT CLOSED. THE PATIENT WAS REPORTEDLY UNAFFECTED, AND THE PRODUCT IS NOT AVAILABLE FOR RETURN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6893684 ↗MalfunctionAug 7, 2017Material Frayed
UPON OPENING THE PACKAGE OF THE DEVICE PRIOR TO DEVICE MANIPULATION, THE PHYSICIAN NOTICED THAT THE WIRE GUIDE TIP WAS FRAYED. ANOTHER DEVICE WAS USED INSTEAD. THE DEVICE DID NOT MAKE PATIENT CONTACT, AND THERE WERE NO INJURIES OR ADDITIONAL MEDICAL INTERVENTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6770278 ↗InjuryApr 13, 2017Break; Break
INTERNATIONAL CUSTOMER REPORTED THAT A COOK TOTAL PARENTERAL NUTRITION (TPN) DOUBLE LUMEN CATHETER WAS IMPLANTED INTO A PATIENT ON (B)(6) 2016. THE EXTENSION TUBING OF THE CATHETER REPORTEDLY FRACTURED ON (B)(6) 2017. (1820334-2017-00680) THE CONDITIONS AND CIRCUMSTANCES SURROUNDING THE EVENT ARE NOT KNOWN. A KIT WAS OBTAINED TO REPAIR THE DEVICE; THE REPAIR KIT DID NOT APPEAR TO BE THE PROPER SIZE FOR THE NEEDED REPAIR. (1820334-2017-00772) A SECOND REPAIR KIT WAS OBTAINED AND THE CATHETER SUCCESSFULLY REPAIRED. THE REPAIRED CATHETER THEN FRACTURED SEVERAL HOURS LATER AND WAS SUBSEQUENTLY EXPLANTED AND REPLACED WITH A NEW TPN CATHETER (1820334-2017-00772) ON (B)(6) 2017. NO FURTHER EVENT OR PATIENT INFORMATION WAS PROVIDED. THE DEVICE THAT IS THE SUBJECT OF THIS REPORT IS REPORTEDLY AVAILABLE FOR EVALUATION, HOWEVER IT HAS NOT BEEN RECEIVED BY THE MANUFACTURER AS OF THE DATE OF THIS REPORT. THIS IS ONE OF THREE SEPARATE REPORTS BEING SUBMITTED AS THREE EVENTS WERE REPORTED. SEE MDR NUMBERS 1820334-2017-00680, 1820334-2017-00771 AND 1820334-2017-00772 FOR ALL ASSOCIATED EVENTS. THE SAME PATIENT IS INVOLVED IN ALL EVENTS.
Patient problems: No Code Available
MAUDE report 6490336 ↗MalfunctionApr 13, 2017Fitting Problem; Fitting Problem
INTERNATIONAL CUSTOMER REPORTED THAT A COOK TOTAL PARENTERAL NUTRITION (TPN) DOUBLE LUMEN CATHETER WAS IMPLANTED INTO A PATIENT ON (B)(6) 2016. THE EXTENSION TUBING OF THE CATHETER REPORTEDLY FRACTURED ON (B)(6) 2017. THE CONDITIONS AND CIRCUMSTANCES SURROUNDING THE EVENT ARE NOT KNOWN. (1820334-2017-00680) A KIT WAS OBTAINED TO REPAIR THE DEVICE; THE REPAIR KIT DID NOT APPEAR TO BE THE PROPER SIZE FOR THE NEEDED REPAIR (1820334-2017-00771). A SECOND REPAIR KIT WAS OBTAINED AND THE CATHETER SUCCESSFULLY REPAIRED. THE REPAIRED CATHETER THEN FRACTURED SEVERAL HOURS LATER (1820334-2017-00772) AND WAS SUBSEQUENTLY EXPLANTED AND REPLACED WITH A NEW TPN CATHETER ON (B)(6) 2017. NO FURTHER EVENT OR PATIENT INFORMATION WAS PROVIDED. THE DEVICE IS NOT AVAILABLE FOR EVALUATION. THIS IS ONE OF THREE SEPARATE REPORTS BEING SUBMITTED AS THREE EVENTS WERE REPORTED. SEE MDR NUMBERS 1820334-2017-00680, 1820334-2017-00771 AND 1820334-2017-00772 FOR ALL ASSOCIATED EVENTS. THE SAME PATIENT IS INVOLVED IN ALL EVENTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6490315 ↗MalfunctionApr 13, 2017Fracture; Fracture
INTERNATIONAL CUSTOMER REPORTED THAT A COOK TOTAL PARENTERAL NUTRITION (TPN) DOUBLE LUMEN CATHETER WAS IMPLANTED INTO A PATIENT ON (B)(6) 2016. THE EXTENSION TUBING OF THE CATHETER REPORTEDLY FRACTURED ON (B)(6) 2017 (1820334-2017-00680). THE CONDITIONS AND CIRCUMSTANCES SURROUNDING THE EVENT ARE NOT KNOWN. A KIT WAS OBTAINED TO REPAIR THE DEVICE; THE REPAIR KIT DID NOT APPEAR TO BE THE PROPER SIZE FOR THE NEEDED REPAIR.( 1820334-2017-00771). A SECOND REPAIR KIT WAS OBTAINED AND THE CATHETER SUCCESSFULLY REPAIRED. THE REPAIRED CATHETER THEN FRACTURED SEVERAL HOURS LATER (1820334-2017-00772), AND WAS SUBSEQUENTLY EXPLANTED AND REPLACED WITH A NEW TPN CATHETER ON (B)(6) 2017. NO FURTHER EVENT OR PATIENT INFORMATION WAS PROVIDED. THE DEVICE WAS RECEIVED FOR EVALUATION. THIS IS ONE OF THREE SEPARATE REPORTS BEING SUBMITTED AS THREE EVENTS WERE REPORTED. SEE MDR NUMBERS , 1820334-2017-00680, 1820334-2017-00771 AND 1820334-2017-00772 FOR ALL ASSOCIATED EVENTS. THE SAME PATIENT IS INVOLVED IN ALL EVENTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6489970 ↗MalfunctionJul 27, 2016Leak/Splash; Leak/Splash
IT WAS REPORTED THAT A CATHETER, PLACED IN THE EXTERNAL JUGULAR VEIN OF THE (B)(6) OLD FEMALE PATIENT ON (B)(6) 2016, WAS LEAKING FLUID FROM THE AREA NEAR THE BASE OF THE MANIFOLD. THE LEAK WAS TEMPORARILY COVERED WITH REINFORCING TAPE ON (B)(6) 2016. ON (B)(6) 2016 THE DEVICE WAS RETRIEVED OUT OF THE PATIENT, SINCE THERAPY WAS COMPLETED AT THE SAME TIME NO ADDITIONAL DEVICE WAS PLACED IN THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5826130 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page