Cook TPNRecalls & reported problems across all models
Cook Incorporated · 17 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 173
- Deaths reported
- 0
- Injuries reported
- 75
- Recalls
- 0
We found 17 models of Cook TPN by Cook Incorporated in the FDA's device registry (GUDID). Together they are named in 173 adverse event reports in the FDA's MAUDE database, received between Oct 17, 2014 and Aug 23, 2022 — classified as 75 injuries, 98 malfunctions.
The most frequently reported specific device problem is “Break” (65 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Most reported Cook TPN problems
Device problems
- Break65
- Material Separation21
- Leak/Splash18
- Fluid/Blood Leak11
- Fitting Problem9
- Hole In Material9
- Material Twisted/Bent7
- Material Rupture6
Patient problems
- No Known Impact Or Consequence To Patient83
- No Code Available62
- No Consequences Or Impact To Patient19
- No Information6
- No Clinical Signs, Symptoms or Conditions4
- Blood Loss3
- No Patient Involvement3
- Sepsis2
Cook TPN models
17 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Cook TPN G04261Catheter,intravascular,therapeutic,short-term less than 30 days30 reports
- Cook TPN G07903Catheter,intravascular,therapeutic,short-term less than 30 days29 reports
- Cook TPN G07902Catheter,intravascular,therapeutic,short-term less than 30 days25 reports
- Cook TPN G07933Catheter,intravascular,therapeutic,short-term less than 30 days16 reports
- Cook TPN G03319Catheter,intravascular,therapeutic,short-term less than 30 days12 reports
- Cook TPN G07935Catheter,intravascular,therapeutic,short-term less than 30 days12 reports
- Cook TPN G07960Catheter,intravascular,therapeutic,short-term less than 30 days9 reports
- Cook TPN G01868CATHETER, PERCUTANEOUS7 reports
- Cook TPN G01880Catheter,intravascular,therapeutic,short-term less than 30 days7 reports
- Cook TPN G02120CATHETER, PERCUTANEOUS7 reports
- Cook TPN G01873Catheter,intravascular,therapeutic,short-term less than 30 days6 reports
- Cook TPN G29464Catheter,intravascular,therapeutic,short-term less than 30 days5 reports
- Cook TPN G01869CATHETER, PERCUTANEOUS4 reports
- Cook TPN G01872Catheter,intravascular,therapeutic,short-term less than 30 days1 report
- Cook TPN G08074Catheter,intravascular,therapeutic,short-term less than 30 days1 report
- Cook TPN G12022Catheter,intravascular,therapeutic,short-term less than 30 days1 report
- Cook TPN G32440Catheter,intravascular,therapeutic,short-term less than 30 days1 report
Recent Cook TPN adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryAug 23, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT HAD A TRANSLUMBAR TUNNELED CENTRAL VENOUS CATHETER (CVC) THAT REQUIRED REPAIR ON (B)(6) 2020. A REPAIR WAS COMPLETED WITH A COOK TPN SINGLE LUMEN CATHETER REPAIR SET (RPN: C-RHCS-6.5, LOT NUMBER 9836148). TWO YEARS LATER, THE SAME CVC REQUIRED ANOTHER REPAIR AT THE SAME LOCATION. DURING THE REPAIR, IT WAS NOTED THAT THE METAL ROD OF THE PREVIOUS REPAIR KIT HAD MIGRATED TOWARDS THE INSERTION POINT OF THE CATHETER AND WAS LOCATED AT THE AREA WHERE THE CVC SECUREMENT DEVICE HAD BEEN PLACED. THE METAL ROD WAS THEN PUSHED BACK TOWARDS THE CATHETER HUB BEFORE THE LUMEN WAS CUT FOR REPAIR. THE LUMEN WAS REPAIRED WITH ANOTHER COOK TPN SINGLE LUMEN CATHETER REPAIR SET, (RPN: C-RHCS-6.5, LOT NUMBER 9836148). THE PATIENT DID NOT EXPERIENCE ANY ADDITIONAL ADVERSE EFFECTS DUE TO THIS OCCURRENCE. INFORMATION REGARDING EVENT DETAILS HAS BEEN REQUESTED, BUT IS CURRENTLY UNAVAILABLE.
Patient problems: Failure of Implant
MAUDE report 15276329 ↗InjuryAug 6, 2021Fluid/Blood Leak; Fluid/Blood Leak
NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12286877 ↗InjuryJul 9, 2021Fluid/Blood Leak; Fluid/Blood Leak
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12143352 ↗InjuryJun 11, 2021Crack; Crack
IT WAS REPORTED A MALE PATIENT REQUIRED A REDO DOUBLE LUMEN TPN CATHETER SET FOR AN UNKNOWN PROCEDURE. SOME TIME AFTER THE PROCEDURE, THE DEVICE WAS NOTED TO HAVE CRACKED. AN ADDITIONAL SURGERY WAS REQUIRED FOR THE "CONSTRUCTION" OF A NEW DEVICE ON THE LEFT SIDE OF THE PATIENT. THE PATIENT REPORTEDLY "SUFFERED FROM REQUIRED INPATIENT CARE AND PSYCHOSOCIAL STRESS". ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: Distress; Scar Tissue
MAUDE report 11983790 ↗InjuryFeb 9, 2021Fluid/Blood Leak
IN ADDITIONAL INFORMATION RECEIVED ON 19FEB2021, IT WAS REPORTED THAT THE CATHETER WAS PLACED IN THE RIGHT JUGULAR VEIN FOR PARENTERAL NUTRITION IN WHICH INFUSIONS OF TRAVASOL AND LIPIDS WERE BEING ADMINISTERED. THE DEVICE WAS SECURED TO THE PATIENT WITH THE SPECIALIZED DRESSING TEGADERM ADVANCE. THERE WAS NO DIFFICULTY ATTACHING OR DISCONNECTING ANCILLARY DEVICES SUCH AS EXTENSION WITH NEEDLELESS CONNECTORS AND DELIVERY TUBING TO THE CATHETER. PATIENT ACTIVITY LEVEL WAS DESCRIBED AS "QUITE RESTLESS." THE DEVICE FAILED DURING HYPERALIMENTATION TREATMENT. THE FIRST BREAK OCCURRED AT HOME WHERE THE CATHETER WAS HELD IN PLACE WITH KITE LOCK. THE SECOND BREAK HAD A CONTINUOUS INFUSION OF SOLUTE. THERE WAS A NEED FOR RELOCATION OF ANOTHER CENTRAL LINE. THE PATIENT WAS HOSPITALIZED OR REQUIRED PROLONGED HOSPITALIZATION. THE DEVICE IS NOT AVAILABLE TO BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 11301124 ↗InjuryFeb 9, 2021Fluid/Blood Leak; Fluid/Blood Leak
IN ADDITIONAL INFORMATION RECEIVED ON 19FEB2021, IT WAS REPORTED THAT THE CATHETER WAS PLACED IN THE RIGHT JUGULAR VEIN FOR PARENTERAL NUTRITION IN WHICH INFUSIONS OF TRAVASOL AND LIPIDS WERE BEING ADMINISTERED. THE DEVICE WAS SECURED TO THE PATIENT WITH THE SPECIALIZED DRESSING TEGADERM ADVANCE. THERE WAS NO DIFFICULTY ATTACHING OR DISCONNECTING ANCILLARY DEVICES SUCH AS EXTENSION WITH NEEDLELESS CONNECTORS AND DELIVERY TUBING TO THE CATHETER. PATIENT ACTIVITY LEVEL WAS DESCRIBED AS "QUITE RESTLESS." THE DEVICE FAILED DURING HYPERALIMENTATION TREATMENT. THE FIRST BREAK OCCURRED AT HOME WHERE THE CATHETER WAS HELD IN PLACE WITH KITE LOCK. THE SECOND BREAK HAD A CONTINUOUS INFUSION OF SOLUTE. THERE WAS A NEED FOR RELOCATION OF ANOTHER CENTRAL LINE. THE PATIENT WAS HOSPITALIZED OR REQUIRED PROLONGED HOSPITALIZATION. THE DEVICE IS NOT AVAILABLE TO BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Code Available
MAUDE report 11300867 ↗InjuryDec 3, 2020Material Separation; Material Separation
IN THE DEVICE FAILURE ANALYSIS SIGNED ON (B)(6) 2021, IT WAS FOUND THAT THE RED HUB PLUNGER WOULD NOT RELEASE FLUID. ENCRUSTATION WAS NOTED ON THE DEVICE. WHILE FLUID WAS ABLE TO EXIT THE DISTAL TIP OF THE BLUE HUB, A LOT OF RESISTANCE WAS NOTED IN DOING SO.
Patient problems: No Code Available
MAUDE report 10941399 ↗InjuryNov 19, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IN ADDITIONAL INFORMATION RECEIVED ON 25NOV2020, IT WAS REPORTED THAT ANTIBIOTICS WERE USED TO TREAT THE OCCURRENCE OF SEPSIS. THE ONLY PRE-EXISTING CONDITION THAT WOULD HAVE MADE THE PATIENT PRE-DISPOSED FOR TO A HIGHER RISK OF INFECTION WAS SHORT BOWEL SYNDROME. THE PATIENT WAS DESCRIBED AS SELF CARING AT THE TIME SEPSIS OCCURRED. THE DEVICE WAS PLACED IN THE LEFT INTERNAL JUGULAR FOR TPN ADMINISTRATION. THE FAILURE OCCURRED DURING TREATMENT. WHEN NOT IN USE, THE CATHETER WAS FLUSHED WITH AND LOCKED AN 0.9% SALINE SOLUTION. A PRESUMED 10ML SYRINGE WAS USED FOR FLUSHING. A NEEDLELESS INJECTOR CAP WAS USED, HOWEVER NO ANTI-TAMPER DEVICE WAS USED. THE DEVICE WAS INITIALLY SECURED WITH SUTURES, HOWEVER GRIPLOK WAS LATER USED. PATIENT ACTIVITY LEVEL WAS DESCRIBED AS "ACTIVE."
Patient problems: No Code Available; Sepsis
MAUDE report 10868375 ↗InjuryNov 4, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT A REDO SINGLE LUMEN TPN CATHETER HAS SPLIT. THE PATIENT HAS REPORTEDLY GONE THROUGH "SEVERAL" LINES AND HAS PRESENTED WITH A SPLIT LINE TWICE IN (B)(6) 2020. THE SECOND SPLIT IS ALSO REPORTED UNDER AN MDR WITH PATIENT IDENTIFIER (B)(6). DUE TO THIS OCCURRENCE, IT WAS REPORTED THE PATIENT REQUIRED "MORE VENOUS INTERVENTION" AND X-RAY IMAGING. ADDITIONAL INFORMATION REGARDING PATIENT AND EVENT INFORMATION HAS BEEN REQUESTED BUT IS NOT CURRENTLY AVAILABLE.
Patient problems: No Code Available
MAUDE report 10786389 ↗InjuryJul 16, 2020Material Separation; Material Separation
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Code Available
MAUDE report 10284116 ↗InjuryJul 10, 2020Break; Break
IT WAS REPORTED THAT A COOK TPN TRIPLE LUMEN TPN CATHETER SET WAS FOUND TO BE BROKEN AND LEAKING. THE DEVICE WAS PLACED IN THE JUGULARIS INTERNA RIGHT ON (B)(6) 2020 AND SECURED TO THE PATIENT USING SUTURES AND A BANDAGE. WHEN NOT IN USE, THE CATHETER WAS MAINTAINED WITH A HEPARIN LOCK. THE PATIENT WAS REPORTED TO BE BEDRIDDEN. DATE OF EVENT WAS ORIGINALLY REPORTED AS (B)(6) 2020. ADDITIONAL INFORMATION WAS RECEIVED STATING ON 03JUN2020, THE NURSE NOTICED THAT THE BED WAS WET WHERE THE THREE LUMENS OF THE CENTRAL VENOUS CATHETER WERE LOCATED. IT WAS NOTED THAT THE CATHETER HAD BEEN "PIERCED" JUST BEFORE THE CATHETER DIVIDED INTO THREE LUMENS. THE DEVICE HAD BEEN INDWELLING FOR 8 WEEKS. NO PROBLEMS WITH FLUSHING OR DRAWING FLUIDS FROM THE CATHETER HAD BEEN NOTICED. THE CATHETER WAS SUBSEQUENTLY REMOVED AND REPLACED UNDER FULL ANESTHESIA. ABOUT SIX HOURS POST-PROCEDURE, THE PATIENT WAS ADMITTED TO THE INTENSIVE CARE UNIT DUE TO SEPSIS. THE PATIENT WAS TREATED WITH ANTIBIOTICS AND SPENT TWO AND A HALF WEEKS IN THE ICU. THE PATIENT STILL HAS INPATIENT STATUS.
Patient problems: No Code Available; Sepsis
MAUDE report 10260325 ↗MalfunctionJul 10, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
ADDITIONAL INFORMATION WAS RECEIVED ON 17JUL2020. IT WAS REPORTED THE CATHETER WAS FIRST PLACED ON "30DEC2020" IN THE PATIENT'S THORAX. IT IS LIKELY THE CATHETER WAS ACTUALLY PLACED ON 30DEC2019. IT WAS REPAIRED ONCE ON (B)(6) 2020 AND AGAIN ON (B)(6) 2020. AFTER EACH REPAIR THE PATIENT WAS HOSPITALIZED AS NECESSARY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10259967 ↗InjuryJul 10, 2020Break; Break
NO ADDITIONAL PATIENT/EVENT INFORMATION HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Code Available
MAUDE report 10259751 ↗InjuryJul 10, 2020Break; Material Rupture; Break; Material Rupture
IT WAS REPORTED THAT A COOK TPN TRIPLE LUMEN TPN CATHETER SET WAS FOUND TO BE BROKEN AND LEAKING. THE DEVICE WAS PLACED IN THE JUGULARIS INTERNA RIGHT AND SECURED TO THE PATIENT USING SUTURES AND A BANDAGE. WHEN NOT IN USE, THE CATHETER WAS MAINTAINED WITH A HEPARIN LOCK. THE PATIENT WAS REPORTED TO BE ACTIVE. ON (B)(6) 2020, WHILE FLUSHING THE DEVICE, BLOOD WAS FOUND TO BE SLOWLY SEEPING OUT OF THE CATHETER, JUST BEFORE IT SPLITS INTO THREE LUMENS. NO OTHER PROBLEMS WERE NOTED WITH FLUSHING OR DRAWING FLUIDS. THE DEVICE HAD BEEN INDWELLING FOR 8 WEEKS. THE CATHETER WAS SUBSEQUENTLY REMOVED AND REPLACED UNDER FULL ANESTHESIA. NO COMPLICATIONS WERE NOTED DURING THE REPLACEMENT PROCEDURE. THE PATIENT NOW HAS OUTPATIENT STATUS.
Patient problems: No Code Available
MAUDE report 10259154 ↗MalfunctionMay 12, 2020Break; Fitting Problem; Break; Fitting Problem
A DEVICE FAILURE ANALYSIS WAS CONDUCTED ON 01JUN2020 FOR THE RETURNED DEVICE AND FOUND THAT IN ADDITION TO THE BREAK IN THE LINE, THE SILICONE SLEEVE HAS COME LOOSE FROM THE WINGED HUB, WHICH WAS NOT REPORTED BY THE CUSTOMER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10048335 ↗
Other Cook Incorporated products
All Cook Incorporated devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.