Skip to content

Cook G02058Recalls & reported problems

Cook Incorporated · Femoral Artery Catheter Set · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002020580), received between Apr 18, 2025 and Apr 18, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 18, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT THE CATHETER INCLUDED IN THE FEMORAL ARTERY PRESSURE MONITORING SET FRACTURED AT THE HUB. UPON ARRIVAL IN THE ICU POST-SURGERY, IT WAS NOTED THAT THE FEMORAL ARTERIAL LINE HAD BEEN SHEARED OFF. THE SURGEON WAS PROMPTLY NOTIFIED, AND THE PATIENT WAS TRANSFERRED TO THE CARDIOVASCULAR INTERVENTIONAL LABORATORY (CVIL) FOR REMANGIOGRAPHY. ON (B)(6) 2025, THE PATIENT WAS RECEIVED IN CVIL FOLLOWING AN OFF-PUMP CORONARY ARTERY BYPASS GRAFTING (OPCABCG) PROCEDURE INVOLVING THREE GRAFTS. THE FEMORAL CANNULA WAS IN SITU IN THE LEFT FEMORAL ARTERY AND SUTURED TO THE SKIN. THE RIGHT GROIN WAS BRUISED AND SHOWED SIGNS OF HEMATOMA FROM A CORONARY ANGIOGRAM PERFORMED ON (B)(6)2025. THE PATIENT WAS INTUBATED AND VENTILATED. AN ULTRASOUND OF THE LEFT GROIN REVEALED A FOREIGN BODY, WHICH WAS VISIBLE ON THE ULTRASOUND-GUIDED ACCESS. A 4FR SHEATH WAS INITIALLY INSERTED INTO THE RIGHT FEMORAL ARTERY (RFA), WHICH WAS LATER UPGRADED TO A 6FR 45CM SHEATH. THE FOREIGN BODY WAS SUCCESSFULLY RETRIEVED USING A SNARE. POST-ANGIOGRAPHY REVEALED TRAUMA TO THE VESSEL, BUT THERE WAS NO BLEEDING FROM THE RIGHT GROIN. A COMPETITOR DEVICE (5FR) WAS USED TO CLOSE THE RIGHT GROIN. THE PATIENT WAS TH

    Patient problems: Foreign Body In Patient

    MAUDE report 21866136 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page