Cook G02092Recalls & reported problems
Cook Incorporated · Radial Artery Catheter Set · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00827002020924), received between Apr 13, 2017 and Jan 26, 2024 — classified as 5 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Complete Blockage” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Complete Blockage2
- Device Contaminated During Manufacture or Shipping2
- Unable to Obtain Readings2
- Break1
- Delivered as Unsterile Product1
- Obstruction of Flow1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- No Patient Involvement4
- No Clinical Signs, Symptoms or Conditions3
- No Code Available3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryJan 26, 2024Unable to Obtain Readings; Obstruction of Flow; Unable to Obtain Readings; Obstruction of Flow
ADDITIONAL INFORMATION WAS RECEIVED (B)(6) 2024. THE DEVICE WAS SAID TO BE IN PLACE FOR ONE HOUR OR MORE PRIOR TO THE FAILURE. THE CATHETER WAS FLUSHED INTRAOPERATIVELY WITH SALINE SOLUTION WITHOUT HEPARIN BEFORE PUNCTURING THE ARTERY. HOWEVER, JUST FLUSHING THE CATHETER WAS NOT ENOUGH; A MANUAL FLUSH WITH 0.9% SALINE SOLUTION IN A SYRINGE WAS NECESSARY TO UNCLOG THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18586702 ↗InjurySep 3, 2021Complete Blockage; Unable to Obtain Readings; Complete Blockage; Unable to Obtain Readings
IT WAS REPORTED IN ADDITIONAL INFORMATION RECEIVED ON 06OCT2021 THAT THERE WAS RESISTANCE WHEN ADVANCING THE CATHETER OVER THE WIRE. IT WAS NOTED THAT THERE WAS BENDING BETWEEN THE CATHETER AND THE "FIXATION BUTTERFLY", CAUSING OBSTRUCTION "WITH VERY HIGH FREQUENCY", DESPITE THE USE OF PROPER INSERTION AND FIXATION TECHNIQUES. "SLIP" (AS INITIALLY REPORTED) REFERRED TO THE TIP OF THE CATHETER OPENING WHEN SLIDING THROUGH THE GUIDE AND ENCOUNTERING RESISTANCE IN THE SKIN. SELDINGER TECHNIQUE WAS USED TO INSERT THE DEVICE. THE CATHETER WAS REPLACED WITH ONE FROM ANOTHER MANUFACTURER. THERE WAS NO DIFFICULTY PLACING THE REPLACEMENT DEVICE. THE PATIENT DID NOT REQUIRE HOSPITALIZATION/PROLONGED HOSPITALIZATION DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12421085 ↗InjuryFeb 1, 2021Complete Blockage; Complete Blockage
IT WAS REPORTED THAT A RADIAL ARTERY PRESSURE MONITORING CATHETER WAS OBSTRUCTED, PREVENTING A BLOOD PRESSURE READING FROM BEING OBTAINED. A (B)(6) YEAR OLD FEMALE PATIENT IN POSTOPERATIVE SURGERY HAD A CATHETER PLACED IN THE RIGHT RADIAL ARTERY IN THE UPPER LIMB. THE DEVICE WAS SECURED TO THE PATIENT WITH A CATHETER BUTTERFLY. AN ARTERIAL PRESSURE MONITORING LINE WAS ATTACHED TO THE CATHETER. IMAGING WAS NOT USED TO AID IN THE PLACEMENT OF THE DEVICE. WHEN PERFORMING THE SANITIZATION AND REPOSITIONING OF THE LIMB, IT WAS VERIFIED THAT INVASIVE PRESSURE WAS NOT BEING READ. SEVERAL WASHING MANEUVERS WERE PERFORMED TO UNBLOCK THE DEVICE BUT WERE UNSUCCESSFUL. IT WAS REPORTED THAT THE DEVICE WAS REPLACED. UPON REMOVAL, NO DEFECT OF THE DEVICE WAS NOTED..
Patient problems: No Clinical Signs, Symptoms or Conditions; No Code Available
MAUDE report 11259575 ↗MalfunctionNov 4, 2019Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
NO ADDITIONAL PATIENT/EVENT INFORMATION HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Patient Involvement
MAUDE report 9273629 ↗MalfunctionSep 10, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Patient Involvement
MAUDE report 8983352 ↗MalfunctionJul 8, 2019Delivered as Unsterile Product; Delivered as Unsterile Product
NO ADDITIONAL INFORMATION REGARDING THE PATIENT OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Patient Involvement
MAUDE report 8768422 ↗MalfunctionDec 21, 2018Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
NO NEW EVENT DESCRIPTION INFORMATION TO REPORT AT THIS TIME.
Patient problems: No Patient Involvement
MAUDE report 8192293 ↗MalfunctionDec 18, 2018Break
THE COMPLAINANT REPORTED THAT THE PACKAGE FOR THE SINGLE LUMEN DISTAL AORTIC PRESSURE MONITORING SET WAS NOTED TO BE BROKEN DURING INSPECTION. THIS PRODUCT PROBLEM WAS NOTED PRIOR TO PATIENT CONTACT, AND IT DID NOT RESULT IN ANY REPORTED PATIENT IMPACT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8176654 ↗InjuryApr 13, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT UNDERWENT PLACEMENT OF A RADIAL ARTERY PRESSURE MONITORING CATHETER ON AN UNSPECIFIED DATE. THE PATIENT COMPLAINED OF UPPER LEFT LIMB PAIN AND EDEMA AND A "LOSS PAM." IT IS OPINED DUE TO A LANGUAGE BARRIER THAT THE "LOSS PAM" REFERS TO A LOSS OF PERIPHERAL ARTERIAL MONITORING. NO FURTHER PATIENT OR EVENT INFORMATION WAS PROVIDED. IT IS OPINED THAT THE LOSS OF MONITORING RESULTED IN A NEED TO EXPLANT THE DEVICE AND PERFORM A REPLACEMENT. HE DEVICE IS NOT AVAILABLE FOR EVALUATION.
Patient problems: No Code Available
MAUDE report 6489570 ↗InjuryApr 13, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT UNDERWENT PLACEMENT OF A RADIAL ARTERY PRESSURE MONITORING CATHETER ON (B)(6) 2017 VIA A LEFT RADIAL ARTERY PUNCTURE ROUTE. RADIAL ARTERY PRESSURE MONITORING THROUGH THE DEVICE WAS LOST ON (B)(6) 2017; ONE DAY FOLLOWING PLACEMENT OF THE DEVICE AND INITIATION OF MONITORING. IT IS OPINED THAT THE DEVICE WAS EXPLANTED AND A REPLACEMENT ARTERIAL MONITORING CATHETER WAS PLACED. NO FURTHER INFORMATION WAS PROVIDED. THE DEVICE IS NOT AVAILABLE FOR EVALUATION.
Patient problems: No Code Available
MAUDE report 6489558 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page