Skip to content

Cook G02433Recalls & reported problems

Cook Incorporated · Lock Pericardiocentesis Set · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00827002024335), received between Nov 16, 2016 and Feb 26, 2026 — classified as 2 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Unraveled Material3
  • Material Frayed2
  • Material Separation2
  • Coiled1
  • Detachment Of Device Component1
  • Device Contaminated During Manufacture or Shipping1
  • Device Contamination with Chemical or Other Material1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • No Code Available3
  • No Clinical Signs, Symptoms or Conditions2
  • No Patient Involvement2
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2026Unraveled Material

    IT WAS REPORTED THAT THE WIRE GUIDE FROM A COOK LOCK PERICARDIOCENTESIS CATHETER SET WAS FOUND TO BE STRETCHED AND SEPARATED. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24455270 ↗
  • MalfunctionNov 24, 2025Fluid/Blood Leak; Gas/Air Leak; Fluid/Blood Leak; Gas/Air Leak

    IT WAS REPORTED IN ADDITIONAL INFORMATION THAT NOTHING WAS ATTACHED TO THE DEVICE; THE DEVICE WAS TO BE USED BY ITSELF UNDER IMAGE GUIDANCE. PRIOR TO USING THE PRODUCT, AIR LEAKAGE WAS OBSERVED FROM THE HUB OF THE CATHETER, "THE PROXIMAL DISTAL".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23638594 ↗
  • MalfunctionOct 18, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED BY THE DISTRIBUTOR THAT THERE WAS A HAIR INSIDE A STERILE LOCK PERICARDIOCENTESIS CATHETER SET. NO PATIENT CONTACT WAS MADE.

    Patient problems: No Patient Involvement

    MAUDE report 9209242 ↗
  • MalfunctionMar 25, 2019Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8449664 ↗
  • MalfunctionApr 23, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    THE (B)(4) DISTRIBUTOR REPORTED THAT BACK HAIR WAS OBSERVED INSIDE THE PRIMARY PACKAGE DURING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7450091 ↗
  • MalfunctionApr 6, 2018Unraveled Material; Unraveled Material

    THE INTERNATIONAL CUSTOMER REPORTED THAT THE LOCK PERICARDIOCENTESIS CATHETER SET, DURING THE IMPLANT OF A PERICARDIAL DRAIN, WHEN RETRIEVING THE GUIDE, THERE WAS A SENSE OF RESISTANCE: IN FACT, THE GUIDE WAS UNRAVELED. THE DRAIN AND WIRE GUIDE WERE REMOVED AT THE SAME TIME BECAUSE OF THE RESISTANCE AND THEN ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. PER THE CUSTOMER, THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT. THE DEVICE IS REPORTEDLY UNAVAILABLE FOR RETURN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7404233 ↗
  • MalfunctionMar 28, 2018Break; Coiled; Material Frayed

    THE FOLLOWING INFORMATION WAS REPORTED: THE LOCK PERICARDIOCENTESIS CATHETER SET WAS USED. AS THE PHYSICIANS WERE PULLING THE WIRE OUT OF THE PIGTAIL CATHETER AFTER BEING PLACED IN THE CORRECT POSITION, THE WIRE BEGAN TO GET A THINNER CALIBER (FRAYED) AND COILED. PART OF THE WIRE BROKE. IT CONTINUED TO STRETCH AND THEY STOPPED WHEN THEY MET RESISTANCE AND FOUND PART OF THE WIRE BROKE. IT WAS BEING HELD TOGETHER BY THE SMALL THIN COILED WIRE. THEY OPTED TO REMOVE THE PIG TAIL CATHETER AND WIRE TO PLACE A NEW CHEST TUBE USING A NEW SET. A CHEST X-RAY WAS ALSO DONE TO CONFIRM PLACEMENT AND TO MAKE SURE NO FOREIGN BODY WAS FOUND INSIDE THE PATIENT. NO PARTS OF THE DEVICE REMAINED IN THE PATIENT. THERE WERE NO ADDITION AFFECTS TO THE PATIENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7379110 ↗
  • InjuryMar 28, 2018Break; Break

    THE INTERNATIONAL CUSTOMER REPORTED DURING THE USE OF THE LOCK PERICARDIOCENTESIS CATHETER SET, USING THE SELDINGER TECHNIQUE FOR DRAIN PLACEMENT TO DRAIN FLUID IN LEFT LOWER QUADRANT. THE CUSTOMER STATED THEY NO THIS IS CONSIDERED AS OFF LABEL. THE STRAW COLORED FLUID WAS DRAINED ON FIRST PASS OF NEEDLE AT FIVE CENTIMETERS (CM). DILATION WAS DONE AND ALLOWED TO DRAINED CATHETER OVER WIRE. IT WAS REPORTED ON ATTEMPT TO REMOVE WIRE, THERE WAS SIGNIFICANT RESISTANCE NOTED AND THE WIRE APPEARED TO BE SHEARING. THE DRAIN WAS REMOVED COMPLETELY AND NO WIRE REMAINED IN SITU WAS CONFIRMED BY ABDOMINAL ULTRASOUND. IT WAS REPORTED THE PRODUCT WAS USED FOR THIS PROCEDURE MANY TIMES WITHOUT ADVERSE EFFECT, AND AGREED IT MAY HAVE BEEN USER ERROR. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. THE CUSTOMER REPORTED THE DEVICE IS AVAILABLE FOR RETURN, BUT HAS NOT BEEN RECEIVED FOR EVALUATION AS OF THE DATE OF THIS REPORT.

    Patient problems: No Code Available

    MAUDE report 7377264 ↗
  • MalfunctionMar 28, 2018Unraveled Material

    THE COMPLAINANT REPORTED THAT THE PERICARDIAL DRAIN WAS USED AND THE GUIDE WIRE UNRAVELED (THE MESH DISASSEMBLED) DURING WITHDRAWAL. THERE WERE NO REPORTED ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE PRODUCT PROBLEM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7377065 ↗
  • InjuryMar 27, 2018Break; Detachment Of Device Component

    IT WAS REPORTED THAT WHILE USING A LOCK PERICARDIOCENTESIS CATHETER SET A PIECE BROKE OFF THE DEVICE AND THE DRAINAGE HAD TO BE REMOVED COMPLETELY FROM THE BODY; IT WAS FURTHER NOTED THAT THE DISTAL WELD CONNECTION SEPARATED FROM THE COIL WHICH ALLOWED THE COIL TO ELONGATE. THERE WERE NO OTHER ADVERSE EFFECTS TO THE PATIENT REPORTED. ADDITIONAL INFORMATION HAD BEEN REQUESTED BUT NOT PROVIDED BY THE REPORTER.

    Patient problems: No Code Available

    MAUDE report 7374418 ↗
  • MalfunctionNov 30, 2017Material Frayed; Material Separation; Material Frayed; Material Separation

    THE INTERNATIONAL CUSTOMER REPORTED THAT THE GUIDEWIRE IN THE LOCK PERICARDIOCENTESIS CATHETER SET WAS FRAYING AND HAS BROKEN DURING USE. THE WIRE WAS REMOVED, AND AN EXTENDED SCAN (IMAGING) CONFIRMED THAT NO PARTS OF THE PRODUCT REMAINED IN THE PATIENT. THE CUSTOMER CONFIRMED NO ADDITIONAL PATIENT CONSEQUENCES OCCURRED. THE PRODUCT HAS BEEN RETURNED FOR EVALUATION; HOWEVER, AS OF THE DATE OF THIS REPORT, THE INVESTIGATION IS STILL PENDING.

    Patient problems: No Code Available

    MAUDE report 7072050 ↗
  • MalfunctionNov 16, 2016Material Fragmentation; Material Fragmentation

    IT WAS REPORTED "AFTER SUCCESSFULLY INSERTING THE PIG TAIL ON THE FIRST ATTEMPT, THE PHYSICIAN MET RESISTANCE WHILE TRYING TO REMOVE THE WIRE. HE RETRACTED THE PIG TAIL SLIGHTLY AND THE WIRE WAS EASILY REMOVED. AT THIS POINT THE PIG TAIL WAS REMOVED. THE INTRODUCER NEEDLE WAS REPLACED IN THE ORIGINAL PUNCTURE SITE. THE PHYSICIAN REINTRODUCED THE WIRE AND PIGTAIL A SECOND TIME. THERE WAS NO RESISTANCE TO WIRE REMOVAL ON THE SECOND ATTEMPT. ONCE THE GUIDEWIRE WAS PULLED OUT, IT STARTED TO BREAK AND SHRED. PER CUSTOMER, THEY DO NOT BELIEVE ANY PORTION BROKE OFF OR SHREDDED OFF INSIDE THE PATIENT." A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT''S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6106181 ↗

FDA clearance and classification

The FDA regulates product code GBX (catheter, irrigation) under 21 CFR 878.4200 as a Class I device. 21 CFR 878.4200 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Cook G02433

Has the Cook G02433 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002024335), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Cook G02433?

The FDA's MAUDE database lists 12 adverse event reports for this device, received between Nov 16, 2016 and Feb 26, 2026. In those reports, the most common specific device problems are “Break” (3), “Unraveled Material” (3) and “Material Frayed” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Cook G02433?

None of the reports are classified as deaths, 2 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Cook G02433 FDA approved?

Devices of this type (product code GBX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.

Who makes the Cook G02433, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Cook, model G02433). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page