Kopans G02587Recalls & reported problems
Cook Incorporated · Kopans Breast Lesion Localization Needle · Class 1 · Implant
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002025875), received between Oct 5, 2017 and Aug 8, 2018 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Migration or Expulsion of Device3
- Detachment of Device or Device Component1
- Fracture1
Patient problems
- No Code Available3
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 8, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED BETWEEN THE TIME A KOPANS MODIFIED BREAST LESION LOCALIZATION NEEDLE WAS PLACED AND THE SCHEDULED PROCEDURE, THE MARKER MOVED. DUE TO THE LESION NOT CLEARLY IDENTIFIED, RESECTION WAS "IRRADICAL" AND THE SCHEDULED INTRAOPERATIVE RADIATION THERAPY (IORT) LED TO A RE-OPERATION. TRANSLATION OF FURTHER INFORMATION INDICATED: "SMALL TUMOR MASS RECOGNIZABLE BY A SMALL CALCIFICATION MARK, PATIENT HAD A LARGE COMMUTING MOM AND A LOT OF FAT AND A SMALL TUMOR, INGREDIENTS FOR DISLOCATION. THE WIRE IS PLACED WITH THE TIP THROUGH THE MASS, BUT THE BARB STILL HANGS IN THE GREASY TISSUE: M.I. THEN MORE LIKELY DISLOCATION IN THIS SINGLE BARB. WE ALMOST ALWAYS PLACE UNDER ULTRASOUND GUIDE, WHERE THE TIP OF THE NEEDLE IS PLACED JUST 1-5 MM THROUGH THE MASS AND THEN UNDER ECHO VIEW THE WIRE IS HELD IN PLACE WHILE THE NEEDLE IS RETRACTED. WHEN MASKING THE WIRE, AFTER MAMMOGRAPHICALLY DOCUMENTING CORRECT POSITION CC AND LM, THE THREAD IS ROLLED UP EXTERNALLY LOOSELY AND TAPED OFF WITH FIXOMUL, IN PATIENTS WITH SKIN THAT IS SUITABLE FOR THIS, NOT QUICK WOUNDS OR RASHES IN PLASTERS WITH TRANSPARENT PLASTIC ( 'SECOND SKIN', PRODUCT NAME ESCAPED ME). LOCALIZATIONS TAKE PLACE AT THE END OF T
Patient problems: No Code Available
MAUDE report 7764652 ↗InjuryJun 8, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED A KOPANS MODIFIED BREAST LESION LOCALIZATION NEEDLE WAS NOT IN THE CORRECT PLACE FOR THE SURGEON. THE PATIENT WAS TAKEN FROM THE OPERATING ROOM TO RADIOLOGY WHERE THE RADIOLOGIST PLACED A SECOND MARKER. AS REPORTED, THIS INCIDENT CAUSED A DELAY IN THE PROCEDURE, HOWEVER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7582398 ↗InjuryJun 8, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED WHILE USING A KOPANS MODIFIED BREAST LESION LOCALIZATION NEEDLE, THE DEVICE MIGRATED FROM THE INTENDED BREAST TUMOR LOCATION. THE RADIOLOGIST REMOVED AND REPOSITIONED THE DEVICE USING AN ULTRASOUND TO ESTABLISH CORRECT PLACEMENT. AS REPORTED, THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available; No Known Impact Or Consequence To Patient
MAUDE report 7582384 ↗InjuryDec 12, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
THE INTERNATIONAL CUSTOMER REPORTED THAT THE WIRE INCLUDED IN THE KOPANS MODIFIED BREAST LESION LOCALIZATION NEEDLE SET HAD DISLODGED FROM THE PATIENT. THE DEVICE HAD BEEN PLACED PRIOR TO A SCHEDULED LUMPECTOMY, AND THE PRODUCT ISSUE WAS NOTICED ONCE THE PATIENT REACHED THE OPERATING ROOM. AT THIS POINT, THE PATIENT WAS RESCHEDULED FOR ANOTHER PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE CUSTOMER, BUT NONE HAS YET BEEN PROVIDED. THE DEVICE IS REPORTEDLY UNAVAILABLE FOR RETURN.
Patient problems: No Code Available
MAUDE report 7107559 ↗MalfunctionOct 5, 2017Fracture; Fracture
THE INTERNATIONAL CUSTOMER REPORTED THAT THE OPERATOR OF THE KOPANS MODIFIED BREAST LESION LOCALIZATION NEEDLE WAS "RESENTING THE LUMP AND RETRIEVING THE KOPANS MARKER NEEDLE" [SIC] AND IT ALLEGEDLY REMOVED IN THREE PIECES. THE PROCEDURE WAS A BREAST LUMP RESECTION. THE CUSTOMER CONFIRMED THAT THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THE PRODUCT ISSUE, AND ALL PIECES OF THE DEVICE WERE RECOVERED FROM THE PATIENT. THE CIRCUMSTANCES SURROUNDING THE USAGE AND HANDLING OF THE DEVICE WERE NOT REPORTED. THE PRODUCT IS REPORTEDLY AVAILABLE FOR RETURN; HOWEVER, AS OF THE DATE OF THIS REPORT, NO PRODUCT HAS YET BEEN RECEIVED FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6917004 ↗
Other NEEDLE, TUMOR LOCALIZATION devices
- Kopans G02554Cook Incorporated · NEEDLE, TUMOR LOCALIZATION4 reports
- Kopans G02586Cook Incorporated · NEEDLE, TUMOR LOCALIZATION4 reports
- Tuflex 271641SOMATEX Medical Technologies GmbH · Needle, Tumor Localization2 reports
- Tuloc 271661SOMATEX Medical Technologies GmbH · Needle, Tumor Localization2 reports
- Kopans G02900Cook Incorporated · NEEDLE, TUMOR LOCALIZATION1 report
- Kopans G03795Cook Incorporated · NEEDLE, TUMOR LOCALIZATION1 report
- Kopans G02850Cook Incorporated · NEEDLE, TUMOR LOCALIZATION1 report
- Cook G19380Cook Incorporated · NEEDLE, TUMOR LOCALIZATION1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.