Tuloc 271661Recalls & reported problems
SOMATEX Medical Technologies GmbH · Class 1 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04250195600844), received between Apr 14, 2023 and Jul 22, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation1
- Positioning Failure1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 22, 2024Positioning Failure; Positioning Failure
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A MARKING OF BREAST LESION USING A TULOC DEVICE, A PATIENT HAD TO BE OPERATED ON TWICE, DUE TO THE DEVICE DISLOCATED AND WAS BEING STUCK IN THE INCORRECT POSITION. BETWEEN THE 1ST AND 2ND PROCEDURE APPROXIMATELY 6 WEEKS TRANSPIRED. THE WIRE COULD NOT BE REMOVED FROM THE PATIENT IN QUESTION AS IT HAD SHORTENED AFTER BEING IN THE FIRST POSITION. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19795362 ↗InjuryApr 14, 2023Material Fragmentation; Use of Device Problem; Material Fragmentation; Use of Device Problem
IT WAS REPORTED THAT ON (B)(6) 2023, AFTER A BREAST BIOPSY THE PHYSICIAN ATTEMPTED TO IMPLANT A TULOC HARPOON, THE PHYSICIAN REPORTED THAT HE HAD TO READJUST THE HARPOON AND ON THE CONTROL MAMMOGRAM IT WAS OBSERVED A SMALL CYLINDER INSIDE THE BREAST THAT HAD DETACHED FROM THE HARPOON. THE PHYSICIAN VERIFIED THE PROPER FUNCTIONING OF THE HARPOON BEFORE USING IT. DURING TRACKING HE REMOVED THE HARPOON TO READJUST ITS POSITION WITHOUT ANY PARTICULAR TECHNICAL PROBLEM. HOWEVER ON THE FOLLOW UP MAMMOGRAM IT WAS FOUND THAT A 17MM LINEA METAL WAS STUCK INSIDE THE BREAST. IT WAS NOT THERE BEFORE REMOVAL. THE HARPOON WAS CHECKED AND A SMALL CYLINDRAL PIECE WAS MISSING FROM THE END OF THE WIRE. THE DEVICE WAS RETURNED FOR EXAMINATION AND IT WAS CONFIRMED THAT A SMALL PIECE WAS MISSING FROM THE END SECTION OF THE DEPLOYER. A FOLLOW UP WITH THE PHYSICIAN REVEALED THAT THE PATIENT WAS NOT AFFECTED BY THE PROCEDURE AND IN A SUBSEQUENT SURGERY ALREADY PLANED TO REMOVE THE PATHOLOGY THE METAL CYLINDER WAS REMOVED AS WELL. NO MORE TISSUE REQUIRED TO BE REMOVED FROM THE PATIENT AND NO ADDITIONAL MEDICAL INTERVENTION REQUIRED. A FOLLOW UP MAMMOGRAM IS SCHEDULED FOR (B)(6) 2023. NO OTHER INFORMATION I
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 16742843 ↗
Other Needle, Tumor Localization devices
- Kopans G02587Cook Incorporated · NEEDLE, TUMOR LOCALIZATION5 reports
- Kopans G02554Cook Incorporated · NEEDLE, TUMOR LOCALIZATION4 reports
- Kopans G02586Cook Incorporated · NEEDLE, TUMOR LOCALIZATION4 reports
- Tuflex 271641SOMATEX Medical Technologies GmbH · Needle, Tumor Localization2 reports
- Kopans G02900Cook Incorporated · NEEDLE, TUMOR LOCALIZATION1 report
- Kopans G03795Cook Incorporated · NEEDLE, TUMOR LOCALIZATION1 report
- Kopans G02850Cook Incorporated · NEEDLE, TUMOR LOCALIZATION1 report
- Cook G19380Cook Incorporated · NEEDLE, TUMOR LOCALIZATION1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.