Skip to content

Amplatz G04001Recalls & reported problems

Cook Incorporated · Amplatz Extra Stiff Wire Guide · Class 2

All models of Amplatz (42)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002040014), received between Apr 25, 2017 and Jan 31, 2023 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Protrusion/Extrusion” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Protrusion/Extrusion3
  • Detachment of Device or Device Component1
  • Difficult to Advance1
  • Unintended Movement1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • Cardiac Perforation1
  • Cardiac Tamponade1
  • Pericardial Effusion1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 31, 2023Unraveled Material; Material Protrusion/Extrusion; Unraveled Material; Material Protrusion/Extrusion

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16271011 ↗
  • MalfunctionSep 16, 2022Material Protrusion/Extrusion; Material Protrusion/Extrusion

    AS REPORTED, UPON OPENING THE PACKAGE FOR AN UNKNOWN PROCEDURE, THE MANDRIL WAS FOUND TO BE PROTRUDING FROM A KINKED AMPLATZ EXTRA-STIFF SUPPORT WIRE GUIDE. THE DEVICE DID NOT MAKE PATIENT CONTACT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15440501 ↗
  • InjuryMar 5, 2019Unintended Movement; Unintended Movement

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Cardiac Perforation; Cardiac Tamponade; Pericardial Effusion

    MAUDE report 8391455 ↗
  • MalfunctionSep 20, 2018Difficult to Advance; Material Protrusion/Extrusion

    IT WAS REPORTED THAT THE PHYSICIAN HAD THE AMPLATZ EXTRA-STIFF SUPPORT WIRE GUIDE WAS BEING USED FOR ENDOVASCULAR THERAPY IN THE SUPERFICIAL FEMORAL ARTERY WITH A CONTRALATERAL APPROACH, BUT THE WIRE WOULD NOT ADVANCE THROUGH THE SHEATH. IT WAS REPLACED WITH ANOTHER MANUFACTURER'S WIRE, AND THE PROCEDURE WAS COMPLETED. IT WAS LATER FOUND THAT THE GUIDEWIRE'S INNER MANDREL WAS PROTRUDING FROM THE COIL. THERE WERE NO INJURIES OR ADDITIONAL PROCEDURES REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7895828 ↗
  • MalfunctionApr 25, 2017Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE PHYSICIAN OBSERVED THE STAINLESS STEEL MANDRIL EXPOSED OUTSIDE THE POLYTETRAFLUOROETHYLENE (PTFE) ON THE AMPLATZ EXTRA-STIFF SUPPORT WIRE GUIDE WHILE HE WAS SHAPING THE WIRE. THIS OCCURRED PRIOR TO ANY PATIENT CONTACT. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS AND DID NOT REQUIRE ANY ADDITIONAL PROCEDURES AS A RESULT OF THIS INCIDENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6520945 ↗