Quick summary

This Cook Cope mandril wire guide is used to help guide catheters during procedures. Reports mostly describe the wire unraveling or breaking, sometimes leaving a fragment in the patient; there are 41 reports, no deaths, and 17 injuries. One open recall began in August 2022 because packaging seals may be breached, potentially compromising sterility. There were 0 reports in the most recent 12 months, compared with 7 in the previous 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
41
Deaths reported
0
Injuries reported
17
Recalls naming it
1

Summary

The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 00827002051836), received between Jul 12, 2016 and Sep 12, 2025 — classified as 17 injuries, 24 malfunctions.

The most frequently reported specific device problem is “Material Separation” (20 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 16, 2022Open, Classified

    Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

    Z-1830-2022 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Material Separation20
  • Unraveled Material15
  • Break5
  • Difficult to Remove2
  • Fracture2
  • Material Fragmentation2
  • Material Twisted/Bent2
  • Entrapment of Device1

Patient problems

  • No Known Impact Or Consequence To Patient13
  • Foreign Body In Patient9
  • No Clinical Signs, Symptoms or Conditions8
  • No Code Available8
  • Device Embedded In Tissue or Plaque3
  • Insufficient Information2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 12, 2025Unraveled Material; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23041973 ↗
  • MalfunctionSep 12, 2025Unraveled Material; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23041817 ↗
  • MalfunctionSep 12, 2025Unraveled Material; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23041739 ↗
  • MalfunctionMay 21, 2025Unraveled Material; Unraveled Material

    IT WAS REPORTED THAT THE COPE MANDRIL WIRE GUIDE UNRAVELED. THE GUIDEWIRE FROM A NEWLY OPENED SET WAS INTRODUCED THROUGH A COMPETITOR NEEDLE FOR PERCUTANEOUS LEFT PORTAL VENOUS ACCESS. UPON INSERTION, THE TIP OF THE WIRE APPEARED TO HAVE PARTIALLY DETACHED AND ELONGATED LIKE A STRING. THE WIRE GUIDE WAS REPLACED WITH A NEW ONE TO SUCCESSFULLY COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING EVENT DETAILS HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22062533 ↗
  • MalfunctionMay 1, 2025Break; Break

    IT WAS REPORTED THAT A COPE MANDRIL WIRE GUIDE BROKE WHILE IN A PATIENT DURING AN UNKNOWN PROCEDURE. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 21943004 ↗
  • MalfunctionFeb 25, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF A COPE MANDRIL WIRE GUIDE "CAME LOOSE" AND SEPARATED FROM THE MANDRIL. THE DEVICE WAS REQUIRED FOR ACCESS TO THE FEMORAL ARTERY. THE WIRE WAS NOT ALTERED FROM ITS ORIGINAL CONDITION PRIOR TO USE. THE SEPARATION OCCURRED UPON REMOVAL FROM THE ARTERY. NO PORTION OF THE DEVICE WAS LEFT IN THE BODY OF THE PATIENT. THE PROCEDURE WAS COMPLETED BY USING ANOTHER OF THE SAME DEVICE. IT IS UNKNOWN IF ANY RESISTANCE WAS ENCOUNTERED DURING INSERTION OF THE WIRE, OR IF THE WIRE WAS REMOVED THROUGH A NEEDLE OR OTHER INSTRUMENT. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21459606 ↗
  • MalfunctionNov 7, 2024Unraveled Material; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20634496 ↗
  • InjuryJun 12, 2023Material Separation; Unraveled Material; Material Separation; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 17107931 ↗
  • InjuryMay 8, 2023Break; Fracture; Unraveled Material; Break; Fracture; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 16886741 ↗
  • InjuryAug 19, 2021Material Separation; Material Twisted/Bent; Material Separation; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A PERMACATH INSERTION, THE TIP OF THE COPE MANDRIL WIRE GUIDE BROKE OFF INTO THE RIGHT SUBCLAVIAN VEIN OF AN UNKNOWN PATIENT. A FOREIGN BODY RETRIEVAL WAS PERFORMED TO SECURE THE TIP. AN ADDITIONAL PUNCTURE AT ANOTHER SITE WAS NEEDED TO RETRIEVE THE BROKEN PIECE OF WIRE GUIDE. NO HARM TO THE PATIENT HAS BEEN REPORTED. ADDITIONAL INFORMATION REGARDING EVENT AND PATIENT DETAILS HAS BEEN REQUESTED, BUT IS CURRENTLY UNAVAILABLE. THE RETRIEVED TIP OF THE WIRE GUIDE IS NOT AVAILABLE FOR RETURN, AS IT WAS FINE AND WAS LOST.

    Patient problems: Foreign Body In Patient

    MAUDE report 12345962 ↗
  • MalfunctionMay 24, 2021Material Separation; Material Separation

    IT WAS REPORTED AN UNKNOWN PATIENT WITH TORTUOUS ANATOMY REQUIRED A COPE MANDRIL WIRE GUIDE FOR AN ATHERECTOMY. DURING THE PROCEDURE IN (B)(6) 2019, THE USER REMOVED THE WIRE GUIDE THROUGH A CATHETER AND DISCOVERED THE END OF THE WIRE GUIDE HAD "CAME UNDONE". THE USER DID NOT OBSERVE KINKING OF THE WIRE GUIDE. THE PROCEDURE WAS COMPLETED WITH ANOTHER UNKNOWN WIRE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11870888 ↗
  • MalfunctionMar 4, 2021Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Insufficient Information

    MAUDE report 11416852 ↗
  • MalfunctionDec 22, 2020Unraveled Material; Unraveled Material

    IT WAS REPORTED AN UNKNOWN PATIENT REQUIRED A COPE MANDRIL WIRE GUIDE FOR A PERIPHERALLY INSERTED CENTRAL CATHETER (PICC) LINE PLACEMENT IN THE RIGHT ARM. DURING THE PROCEDURE, THE WIRE BECAME LODGED IN THE PATIENT'S SCAR TISSUE. WHEN THE OPERATOR ATTEMPTED TO REMOVE THE WIRE BY PULLING IT BACK, THE WIRE UNRAVELED. AFTER SEVERAL ATTEMPTS AT REMOVAL, THE WIRE WAS "RETRIEVED". AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 11061393 ↗
  • InjuryOct 12, 2020Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Device Embedded In Tissue or Plaque; No Code Available

    MAUDE report 10662452 ↗
  • InjuryAug 27, 2020Material Separation; Unraveled Material; Material Separation; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Device Embedded In Tissue or Plaque; No Code Available

    MAUDE report 10462630 ↗
  • MalfunctionJun 16, 2020Unraveled Material; Unraveled Material

    IT WAS REPORTED A COPE MANDRIL WIRE GUIDE WAS UTILIZED IN AN UNKNOWN PROCEDURE. WHEN BEING WITHDRAWN THROUGH THE MICROPUNCTURE NEEDLE, THE WIRE GUIDE "SHEARED." NO OTHER ADVERSE EFFECTS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10157414 ↗
  • MalfunctionMay 8, 2020Unraveled Material; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10039250 ↗
  • InjuryDec 3, 2019Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Foreign Body In Patient; No Code Available

    MAUDE report 9401281 ↗
  • InjuryDec 3, 2019Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 9400801 ↗
  • InjuryNov 19, 2019Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 9340818 ↗
  • MalfunctionOct 2, 2019Unraveled Material; Material Twisted/Bent; Unraveled Material; Material Twisted/Bent

    SEE H10.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9146943 ↗
  • InjuryMay 28, 2019Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 8645394 ↗
  • MalfunctionMay 24, 2019Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8640177 ↗
  • InjuryMay 21, 2019Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Code Available

    MAUDE report 8628128 ↗
  • InjuryMay 21, 2019Material Fragmentation; Material Fragmentation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Code Available

    MAUDE report 8627983 ↗

Other WIRE, GUIDE, CATHETER devices