Skip to content

Cook G05454Recalls & reported problems

Cook Incorporated · Thal-Quick Chest Tube Set · Class 1 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002054547), received between Aug 15, 2017 and Sep 16, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation2
  • Unraveled Material2
  • Break1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Code Available1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Unraveled Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50785367 ↗
  • MalfunctionApr 15, 2024Break; Material Separation; Break; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19106798 ↗
  • MalfunctionSep 20, 2021Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12501272 ↗
  • InjuryAug 17, 2017Loose or Intermittent Connection

    THE COMPLAINANT REPORTED THAT THE PATIENT EXPERIENCED MOTOR AGITATION (RESTLESSNESS) AND MOVED A LOT. THE JOINT BETWEEN THE DRAINAGE CATHETER AND THE MULTI-SIZE CONNECTOR LOOSENED. THE PATIENT REQUIRED AN ADDITIONAL PROCEDURE TO HAVE THE CHEST TUBE REPLACED.

    Patient problems: No Code Available

    MAUDE report 6800085 ↗
  • MalfunctionAug 15, 2017Unraveled Material

    IT WAS REPORTED THE WIRE GUIDE UNRAVELED. NO SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6794269 ↗

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page