Skip to content

Cook G05777Recalls & reported problems

Cook Incorporated · Thal-Quick Chest Tube Set · Class 1 · Implant

Adverse event reports
12
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00827002057777), received between Sep 21, 2017 and Nov 15, 2024 — classified as 3 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Unraveled Material” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unraveled Material3
  • Break2
  • Device Contaminated During Manufacture or Shipping2
  • Difficult to Remove2
  • Material Separation2
  • Crack1
  • Detachment of Device or Device Component1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Foreign Body In Patient1
  • Insufficient Information1
  • Internal Organ Perforation1
  • No Code Available1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2024Unraveled Material; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20694498 ↗
  • MalfunctionSep 3, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20128806 ↗
  • MalfunctionJul 10, 2024Unraveled Material; Unraveled Material

    IN ADDITIONAL INFORMATION RECEIVED ON 17JUL2024, IT WAS REPORTED THAT THE WIRE GUIDE WAS NOT WITHDRAWN OR MANIPULATED THROUGH THE NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19712489 ↗
  • MalfunctionMar 27, 2024Difficult to Remove; Unraveled Material; Difficult to Remove; Unraveled Material

    IT WAS CLARIFIED THAT THE DRAIN WAS NOT DAMAGED. THE DRAIN PORTION OF THE DEVICE WAS DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18988000 ↗
  • InjuryMar 26, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE GUIDEWIRE OF A THAL-QUICK CHEST TUBE SET SEPARATED. DURING PLACEMENT OF A THORACIC DRAIN, THE GUIDEWIRE BECAME STUCK IN THE DRAINAGE CATHETER AND COULD NOT BE REMOVED. THE USER ATTEMPTED TO REMOVE THE GUIDE AND DRAIN TOGETHER WHEN THE GUIDE BROKE OFF INSIDE THE DRAIN. A THORACIC SURGEON PERFORMED A SURGICAL EXCISION AT THE BEDSIDE IN ORDER TO REMOVE THE PIECE OF GUIDEWIRE THAT HAD BROKEN OFF WHEN THE GUIDE WAS REMOVED. NO OTHER ADVERSE EFFECTS HAVE BEEN REPORTED FOR THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 18982808 ↗
  • MalfunctionJan 19, 2024Material Frayed; Difficult to Remove; Material Frayed; Difficult to Remove

    IT WAS REPORTED THAT THE WIRE GUIDE FROM TWO THAL-QUICK CHEST TUBE SETS UNRAVELED DURING A CHEST TUBE INSERTION PROCEDURE. AFTER PLACING THE CATHETER OVER THE WIRE GUIDE AND INTO THE PATIENT, THE WIRE GUIDE WAS DIFFICULT TO REMOVE. AS A RESULT, BOTH THE WIRE GUIDE AND CATHETER WERE REMOVED FROM THE PATIENT. AT THIS TIME IT WAS NOTED THE WIRE GUIDE HAD "FRAYED". THE CATHETER WAS THEN REINSERTED TO COMPLETED THE PROCEDURE. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 18547157 ↗
  • InjuryFeb 23, 2023Use of Device Problem; Use of Device Problem

    NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Internal Organ Perforation

    MAUDE report 16428503 ↗
  • MalfunctionDec 16, 2022Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS INITIALLY REPORTED THAT A WIRE GUIDE FROM A COOK THAL-QUICK CHEST TUBE SET WAS COILED AND KINKED WHEN IT WAS OPENED. A NEW COOK THAL-QUICK CHEST TUBE SET WAS USED TO COMPLETE THE PROCEDURE. THE DEVICE WAS RETURNED TO COOK FOR INVESTIGATION AND FOREIGN MATTER WAS PRESENT ON THE WIRE GUIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15997012 ↗
  • MalfunctionApr 6, 2022Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT WHEN OPENING A COOK THAL-QUICK CHEST TUBE SET IN THE INTENSIVE CARE UNIT (ICU), 'RUST COLOURED' DISCOLORATION ON THE DISTAL END OF THE GUIDE WIRE LOCATED UNDER THE RED INTRODUCER WAS IDENTIFIED. THE DEVICE DID NOT MAKE PATIENT CONTACT. THE PROCEDURE WAS COMPLETED WITHOUT INCIDENT USING ANOTHER COOK THAL-QUICK CHEST TUBE DRAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14025762 ↗
  • InjuryFeb 14, 2019Material Separation; Material Separation

    IT WAS REPORTED IN ADDITIONAL INFORMATION RECEIVED ON 24FEB2019 THAT A 10ML SYRINGE WAS ATTACHED TO THE NEEDLE AT THE TIME OF THE EVENT. A LUER LOCK WAS NOT BEING USED. ALTHOUGH IT WAS UNCLEAR HOW MUCH OF THE NEEDLE WAS EXPOSED OUTSIDE OF THE BODY, IT WAS ESTIMATED THAT MAYBE IT WAS A COUPLE OF CENTIMETERS. THE PATIENT IS CURRENTLY RECOVERING AFTER SURGERY.

    Patient problems: No Code Available

    MAUDE report 8340961 ↗
  • MalfunctionSep 26, 2017Crack; Crack

    THE INTERNATIONAL CUSTOMER REPORTED THAT, DURING A CHEST DRAIN INSERTION PROCEDURE, THE OPERATOR OF THE DEVICE CHECKED THE TIP OF THE DILATOR CONTAINED IN THE THAL-QUICK CHEST TUBE SET AND DISCOVERED THAT THE DISTAL TIP OF THE DEVICE WAS SNAPPED. THE OPERATOR PERFORMED THE CHECK BY TAPPING THE TIP WITH THEIR FINGER. THE TIP HAD NOT DISENGAGED, BUT THERE WAS A CRACK WHICH RAN THE CIRCUMFERENCE OF THE DILATOR. THE BROKEN DILATOR WAS RETAINED, ANOTHER SET WAS OPENED, AND THE PROCEDURE CONTINUED. THE PRODUCT IS REPORTEDLY AVAILABLE TO BE RETURNED; HOWEVER, AS OF THE DATE OF THIS REPORT, NO PRODUCT HAS YET BEEN RECEIVED FOR EVALUATION. THE DILATOR DID NOT COME INTO CONTACT WITH THE PATIENT; ACCORDINGLY, NO PATIENT ADVERSE EVENTS WERE EXPERIENCED.

    Patient problems: No Patient Involvement

    MAUDE report 6894592 ↗
  • MalfunctionSep 21, 2017Break; Break

    THE INTERNATIONAL CUSTOMER REPORTED THAT THE DILATOR PORTION OF A THAL-QUICK CHEST TUBE SET BROKE DURING A DILATION PROCEDURE. AFTER THE BREAK WAS IDENTIFIED, THE OPERATOR OPTED TO USE A DILATOR FROM A DIALYSIS CATHETER INSTEAD, AND WAS ABLE TO COMPLETE THE DRAINAGE PROCEDURE SUCCESSFULLY. THE PATIENT REPORTEDLY DID NOT REQUIRE ANY ADDITIONAL PROCEDURES, AND DID NOT EXPERIENCE AND ADVERSE EVENTS AS A RESULT OF THE ISSUE. THE CIRCUMSTANCES SURROUNDING THE USAGE AND HANDLING OF THE DEVICE ARE NOT KNOWN. THE PRODUCT WAS RETURNED; HOWEVER, AS OF THE DATE OF THIS REPORT, THE INVESTIGATION IS STILL PENDING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6882062 ↗

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.