Melker G06246Recalls & reported problems
Cook Incorporated · Melker Uncuffed Emergency Cricothyrotomy Catheter Set · Class 2
All models of Melker (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002062467), received between Nov 9, 2016 and Jan 5, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Deformation Due to Compressive Stress1
Patient problems
- Death1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2026Deformation Due to Compressive Stress
IT WAS REPORTED THAT THE CATHETER FROM A MELKER EMERGENCY CRICOTHYROTOMY CATHETER SET KINKED. THE DEVICE WAS REQUIRED FOR AN EMERGENCY CRICOTHYROTOMY PROCEDURE PERFORMED OUTSIDE OF THE HOSPITAL. DURING THE PROCEDURE, A CRICOTHYROID MEMBRANE PUNCTURE WAS PERFORMED WITHOUT COMPLICATION. GAS ASPIRATION WAS CONFIRMED. THE WIRE GUIDE WAS INSERTED WITHOUT RESISTANCE, AND A WIDE INCISION WAS MADE AROUND THE WIRE. THE DILATOR WAS ADVANCED OVER THE WIRE WITHOUT DIFFICULTY AND REACHED THE GUARD WITHOUT EFFORT. THE DILATOR WAS THEN REMOVED. THE DILATOR AND CATHETER WERE ASSEMBLED TOGETHER AND WERE ADVANCED OVER THE WIRE. AT THIS TIME, THE CATHETER APPEARED "TOO FLEXIBLE" AND WAS "CRUSHED" AGAINST THE PATIENT'S SKIN AND THE DILATOR AND DID NOT ADVANCE. FINALLY, THE CATHETER WAS INSERTED WITH SOME EFFORT FROM TWISTING; HOWEVER, THE TWISTING MOVEMENT ORIENTED THE CATHETER TOWARDS THE "VAS" PREVENTING ANY VENTILATION. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23957434 ↗InjuryAug 20, 2018Break; Break
A LETTER WAS RECEIVED ON (B)(6) 2018 FROM (B)(6) HOSPITAL (B)(6) INFORMING THE MANUFACTURER OF AN INCIDENT THAT OCCURRED ON (B)(6) 2018. IT WAS REPORTED A PATIENT UNDERWENT A CRICOTHYROTOMY PROCEDURE PERFORMED BY A DOCTOR FROM (B)(6) TEAM IN AN EMERGENCY VEHICLE. THE PATIENT WAS IN CARDIOPULMONARY ARREST DUE TO A QUINCKE EDEMA FOR 15-MINUTES PRIOR TO THE INITIATION OF PROCEDURE. THE PATIENT PASSED NOT BECAUSE OF A FAILED PROCEDURE BUT BECAUSE VITALS WERE COMPROMISED FROM CARDIOPULMONARY ARREST. A (B)(6) EMERGENCY CRICOTHYROTOMY CATHETER SET WAS THE . DEVICE UTILIZED IN THE CRICOTHYROTOMY. AS REPORTED, THE GUIDE FOR MOUNTING THE CANNULA BROKE WHEN REMOVING THE DILATOR. THE CRICOTHYROIDOTOMY WAS UNABLE TO BE PERFORMED RELATE TO EDEMA. WHEN THE EDEMA REGRESSED, THE PATIENT WAS INTUBATED BY ORAL TRACHEA ROUTE. ADDITIONAL INFORMATION HAS BEEN REQUESTED INCLUDING IF AN AUTOPSY REPORT IS AVAILABLE, ADDITIONAL SEQUENCE OF EVENT DETAILS, THE LENGTH OF TIME BEFORE ORAL TRACHEAL INTUBATION WAS ACHIEVED, AND MEDICATIONS AND OTHER DEVICES USED DURING THIS EVENT. HOWEVER, THE REPORTER STATED THAT THE PATIENT DEATH WAS NOT LINKED WITH USE OF THE DEVICE. THEREFORE, THEY WOULD NOT COMMUNICATE ANY M
Patient problems: Death; Patient Problem/Medical Problem
MAUDE report 7799908 ↗MalfunctionNov 9, 2016Break; Break
IT WAS REPORTED WHILE PERFORMING A PROCEDURE THE BLADE OF THE CATHETER SET BROKE. ADDITIONAL INFORMATION HAS BEEN REQUESTED AS OF THE DATE OF THIS REPORT NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6089618 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page