Melker G06246Recalls & reported problems

Cook Incorporated · Melker Uncuffed Emergency Cricothyrotomy Catheter Set · Class 2

All models of Melker (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002062467), received between Nov 9, 2016 and Jan 5, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Deformation Due to Compressive Stress1

Patient problems

  • Death1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 5, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE CATHETER FROM A MELKER EMERGENCY CRICOTHYROTOMY CATHETER SET KINKED. THE DEVICE WAS REQUIRED FOR AN EMERGENCY CRICOTHYROTOMY PROCEDURE PERFORMED OUTSIDE OF THE HOSPITAL. DURING THE PROCEDURE, A CRICOTHYROID MEMBRANE PUNCTURE WAS PERFORMED WITHOUT COMPLICATION. GAS ASPIRATION WAS CONFIRMED. THE WIRE GUIDE WAS INSERTED WITHOUT RESISTANCE, AND A WIDE INCISION WAS MADE AROUND THE WIRE. THE DILATOR WAS ADVANCED OVER THE WIRE WITHOUT DIFFICULTY AND REACHED THE GUARD WITHOUT EFFORT. THE DILATOR WAS THEN REMOVED. THE DILATOR AND CATHETER WERE ASSEMBLED TOGETHER AND WERE ADVANCED OVER THE WIRE. AT THIS TIME, THE CATHETER APPEARED "TOO FLEXIBLE" AND WAS "CRUSHED" AGAINST THE PATIENT'S SKIN AND THE DILATOR AND DID NOT ADVANCE. FINALLY, THE CATHETER WAS INSERTED WITH SOME EFFORT FROM TWISTING; HOWEVER, THE TWISTING MOVEMENT ORIENTED THE CATHETER TOWARDS THE "VAS" PREVENTING ANY VENTILATION. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23957434 ↗
  • InjuryAug 20, 2018Break; Break

    A LETTER WAS RECEIVED ON (B)(6) 2018 FROM (B)(6) HOSPITAL (B)(6) INFORMING THE MANUFACTURER OF AN INCIDENT THAT OCCURRED ON (B)(6) 2018. IT WAS REPORTED A PATIENT UNDERWENT A CRICOTHYROTOMY PROCEDURE PERFORMED BY A DOCTOR FROM (B)(6) TEAM IN AN EMERGENCY VEHICLE. THE PATIENT WAS IN CARDIOPULMONARY ARREST DUE TO A QUINCKE EDEMA FOR 15-MINUTES PRIOR TO THE INITIATION OF PROCEDURE. THE PATIENT PASSED NOT BECAUSE OF A FAILED PROCEDURE BUT BECAUSE VITALS WERE COMPROMISED FROM CARDIOPULMONARY ARREST. A (B)(6) EMERGENCY CRICOTHYROTOMY CATHETER SET WAS THE . DEVICE UTILIZED IN THE CRICOTHYROTOMY. AS REPORTED, THE GUIDE FOR MOUNTING THE CANNULA BROKE WHEN REMOVING THE DILATOR. THE CRICOTHYROIDOTOMY WAS UNABLE TO BE PERFORMED RELATE TO EDEMA. WHEN THE EDEMA REGRESSED, THE PATIENT WAS INTUBATED BY ORAL TRACHEA ROUTE. ADDITIONAL INFORMATION HAS BEEN REQUESTED INCLUDING IF AN AUTOPSY REPORT IS AVAILABLE, ADDITIONAL SEQUENCE OF EVENT DETAILS, THE LENGTH OF TIME BEFORE ORAL TRACHEAL INTUBATION WAS ACHIEVED, AND MEDICATIONS AND OTHER DEVICES USED DURING THIS EVENT. HOWEVER, THE REPORTER STATED THAT THE PATIENT DEATH WAS NOT LINKED WITH USE OF THE DEVICE. THEREFORE, THEY WOULD NOT COMMUNICATE ANY M

    Patient problems: Death; Patient Problem/Medical Problem

    MAUDE report 7799908 ↗
  • MalfunctionNov 9, 2016Break; Break

    IT WAS REPORTED WHILE PERFORMING A PROCEDURE THE BLADE OF THE CATHETER SET BROKE. ADDITIONAL INFORMATION HAS BEEN REQUESTED AS OF THE DATE OF THIS REPORT NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6089618 ↗

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page