Shiley 6CN75HRecalls & reported problems
Covidien LP · Adult Flexible Tracheostomy Tube with TaperGuard Cuff Disposable Inner Cannula · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Covidien Shiley tracheostomy tube is a breathing tube placed through the neck into the windpipe. Its 240 reports mostly describe tubes coming out of the airway and air leaking from the cuff or connections. The reports include 3 deaths and 36 injuries, and reports fell from 16 in the prior 12 months to 5 in the last 12 months. One recall, open since March 2023, covers connectors made narrower than specified, which can make connections insecure.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 240
- Deaths reported
- 3
- Injuries reported
- 36
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 240 adverse event reports that list this exact device (DI 20884521176321), received between Jun 5, 2018 and Mar 10, 2026 — classified as 3 deaths, 36 injuries, 201 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (87 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 24, 2023Open, Classified
A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.
Z-1357-2023 · Root cause: Component change control · FDA record ↗
Most reported problems
Device problems
- Gas/Air Leak87
- Use of Device Problem78
- Defective Component26
- Material Split, Cut or Torn19
- Appropriate Device Problem Term/Code Not Available16
- Connection Problem14
- Labelling, Instructions for Use or Training Problem10
- Fluid/Blood Leak8
Patient problems
- Unintended Extubation116
- No Clinical Signs, Symptoms or Conditions84
- Insufficient Information18
- No Consequences Or Impact To Patient15
- Extubate10
- Unspecified Tissue Injury8
- Hypoventilation7
- Low Oxygen Saturation7
Recent adverse event reports
The 25 most recent of 240 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2026Labelling, Instructions for Use or Training Problem
IT WAS REPORTED THAT DURING TESTING AND MAINTENANCE, THE PACKAGING CHANGE FOR THE FLEXIBLE ADULT TRACHEOSTOMY TUBES WITH DISPOSABLE INNER CANNULAS POSED A SAFETY THREAT DUE TO THE PRODUCT SIZING NOT BEING CLEARLY IDENTIFIED. THE UPDATED PRODUCT PACKAGING AND REFERENCE CODE CHANGES RESULTED IN THE SIZE OF THE TRACHEOSTOMY TUBE NOT BEING VISIBLE TO THE USER. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24563220 ↗MalfunctionDec 2, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING TRACH CHANGE DUE TO HOLE ON THE ANTERIOR SIDE. A NEW TRACH WAS PLACED WITHOUT INCIDENT. AFTER A FEW DAYS A LEAK WAS NOTED AND TUBE WAS REPLACED. IT WAS OBSERVED THAT IT WAS THE SAME ISSUE THAT OF THE HOLE IN THE ANTERIOR SIDE.
Patient problems: Unintended Extubation
MAUDE report 23694713 ↗MalfunctionNov 21, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT POST-OPERATIVELY, THE TRACH TUBE THAT WAS REMOVED DURING REPLACEMENT HAD A HOLE FOUND ON THE ANTERIOR SIDE OF THE CUFF. THE REPLACEMENT TRACH TUBE HAD THE SAME ISSUE BUT WAS RETAINED FROM THE PATIENT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23617866 ↗MalfunctionNov 3, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE CUFF ON THE PATIENT'S TRACHEOSTOMY TUBE WAS DEFLATING AND NOT HOLDING PRESSURE DURING USE. THE PATIENT EXPERIENCED GURGLING SOUNDS INDICATIVE OF AN AIR LEAK. THE SPOUSE WAS ABLE TO ATTEND TO THE PATIENT; THEREFORE, NO HARM OCCURRED TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23452469 ↗MalfunctionNov 3, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE CUFF ON THE PATIENT'S TRACHEOSTOMY TUBE WAS DEFLATING AND NOT HOLDING PRESSURE DURING USE. THE PATIENT EXPERIENCED GURGLING SOUNDS INDICATIVE OF AN AIR LEAK. THE SPOUSE WAS ABLE TO ATTEND TO THE PATIENT, THEREFORE NO HARM OCCURRED TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23452319 ↗MalfunctionOct 15, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING PRE-USAGE TESTING, BOTH TRACH TUBE'S CUFF BLEW. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23303192 ↗MalfunctionOct 15, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING PRE-USAGE TESTING, BOTH TRACH TUBE'S CUFF BLEW. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23302471 ↗MalfunctionAug 18, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE CUFFS OF THE TWO TRACHEOSTOMY TUBE HAD NOT HOLDING AIR SHORTLY AFTER BEING PLACED. REPLACEMENT OF THE TUBE WAS DONE.
Patient problems: Unintended Extubation
MAUDE report 22836009 ↗MalfunctionJul 30, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THREE TRACHEOSTOMY TUBES WERE NOT HOLDING AIR DURING PREPARATION FOR USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22668767 ↗MalfunctionJul 30, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THREE TRACHEOSTOMY TUBES WERE NOT HOLDING AIR DURING PREPARATION FOR USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22668674 ↗MalfunctionJul 30, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THREE TRACHEOSTOMY TUBES WERE NOT HOLDING AIR DURING PREPARATION FOR USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22668666 ↗MalfunctionJul 21, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE TRACH TUBE'S CUFF WILL NOT STAY INFLATED AFTER INSERTION. THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Insufficient Information
MAUDE report 22559939 ↗MalfunctionJun 30, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE CUFF OF THE TWO TRACHEOSTOMY TUBE HAD LEAK PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22367402 ↗MalfunctionMay 19, 2025Insufficient Device Problem Information; Appropriate Device Problem Term/Code Not Available; Insufficient Device Problem Information; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT ROUTINE TRACH CHANGE DONE WITH SAME TYPE AND SIZE OF TRACH AS PREVIOUS. AFTER THE TRACH CHANGE, THE PATIENT EITHER BECOMES IMMEDIATELY SHORT OF BREATH ONCE OFF THE VENT (USING THE SPEAKING VALVE) OR NOTICES INTERMITTENT SHORTNESS OF BREATH THROUGHOUT THE DAY WHILE ON HIS SPEAKING VALVE. ALSO NOTICES SOME INTERMITTENT DIFFICULTIES WITH SPEAKING WHILE ON THE VENT, DEPENDING ON POSITIONING OF THE TRACH. THIS WAS ALSO ALLEVIATED WHEN THE TRACH TUBE WAS PUSHED UP TOWARDS HIS CHIN.
Patient problems: Insufficient Information
MAUDE report 22043830 ↗InjuryMay 12, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT ROUTINE TRACH CHANGE WAS DONE WITH SAME TYPE AND SIZE OF TRACH AS PREVIOUS. AFTER THE TRACH CHANGE, THE PATIENT EITHER BECAME IMMEDIATELY SHORT OF BREATH ONCE OFF THE VENT (USING THE SPEAKING VALVE) OR NOTICED AN INTERMITTENT SHORTNESS OF BREATH THROUGHOUT THE DAY WHILE ON HIS SPEAKING VALVE. SOME INTERMITTENT DIFFICULTIES WITH SPEAKING WHILE ON THE VENT WAS ALSO NOTICED, DEPENDING ON POSITIONING OF THE TRACH. THIS WAS ALLEVIATED WHEN THE TRACH TUBE WAS PUSHED UP TOWARDS HIS CHIN.
Patient problems: Dyspnea; Respiratory Failure; Unintended Extubation
MAUDE report 22002031 ↗MalfunctionFeb 25, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE CUFFS OF THE FOUR TRACHEOSTOMY TUBES WERE NOT INFLATING AND HAD A LEAK. WHILE THE OTHER CUFF "RUPTURED". THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21455449 ↗MalfunctionFeb 25, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE CUFFS OF THE FOUR TRACHEOSTOMY TUBES WERE NOT INFLATING AND HAD A LEAK. WHILE THE OTHER CUFF "RUPTURED". THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21455438 ↗MalfunctionFeb 25, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE CUFFS OF THE FOUR TRACHEOSTOMY TUBES WERE NOT INFLATING AND HAD A LEAK. WHILE THE OTHER CUFF "RUPTURED". THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21455434 ↗MalfunctionFeb 5, 2025Use of Device Problem; Defective Component; Use of Device Problem; Defective Component
IT WAS REPORTED THAT A 7.5MM SHIL CUFFED TRACH CANNULA EXPERIENCED A CUFF INFLATION/DEFLATION ISSUE AND HAD A HOLE IN THE CUFF. THE ISSUE WAS DISCOVERED IN THE OR WHEN THE SURGEON WAS PLACING THE TRACH. THE PRODUCT WAS REPLACED BY A MEDTRONIC PRODUCT. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21315961 ↗InjuryJan 30, 2025Use of Device Problem; Appropriate Device Problem Term/Code Not Available; Use of Device Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING USE OF A 7.5MM SHIL CUFFED TRACH THE PATIENT WAS IN RESPIRATORY DISTRESS DUE TO THE CUFF NOT PROPERLY INFLATING. THERE WAS A HOLE FOUND IN THE BODY OF THE TRACH IN THE MIDDLE OF THE CUFF. THE EVENT OCCURRED IN A HOSPITAL SETTING, AND EVERY TIME THE PATIENT COUGHED, THE CUFF WOULD INFLATE. A RESPIRATORY THERAPIST REMOVED THE TRACH AND IDENTIFIED THE HOLE. A NEW TRACH WAS PLACED WITHOUT ANY ISSUES.
Patient problems: Adult Respiratory Distress Syndrome; Cough; Respiratory Insufficiency; Unintended Extubation
MAUDE report 21280408 ↗MalfunctionJan 29, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT A TRACHEOSTOMY DEVICE PERSISTENTLY LOST CUFF PRESSURE DURING USE. THE CUFF WAS TESTED AND FOUND TO BE LOSING AIR IMMEDIATELY UPON INFLATION. A REPLACEMENT OF THE TUBE WAS PERFORMED.
Patient problems: Unintended Extubation
MAUDE report 21263274 ↗MalfunctionOct 1, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, DURING USE, THE PATIENT SUSTAINED IRRITATION DESCRIBED AS A 'FLAP' AT THE BASE OF THE TRACH. ISSUE OCCURRED AFTER USING A NEW MODEL OF THE TRACHEOSTOMY TUBE. THE PATIENT SWITCHED BACK TO THE OLDER MODEL.
Patient problems: Inflammation; Skin Inflammation/ Irritation; Unintended Extubation
MAUDE report 20343625 ↗MalfunctionJun 28, 2024Disconnection; Connection Problem; Disconnection; Connection Problem
ACCORDING TO THE REPORTER, DURING USE, THE CANNULA WAS EASILY POPPING OFF FROM THE PATIENT'S CONNECTOR. RE-CANNULATION AND REPLACEMENT OF THE TUBE WAS NOT REQUIRED BECAUSE A TRACHEOSTOMY TIE WAS USED TO SECURE THE TUBE UNTIL THE NEXT ROUTINE TRACHEOSTOMY CHANGE WAS REQUIRED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19633799 ↗MalfunctionJun 27, 2024Disconnection; Disconnection
ACCORDING TO THE REPORTER, DURING USE, THE CANNULA WAS EASILY POPPED OFF FROM THE PATIENT'S CONNECTOR. RE-CANNULATION AND REPLACEMENT OF THE TUBE WAS NOT REQUIRED BECAUSE A TRACHEOSTOMY TIE WAS USED TO SECURE THE TUBE UNTIL THE NEXT ROUTINE TRACHEOSTOMY CHANGE WAS REQUIRED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19624971 ↗MalfunctionJun 27, 2024Disconnection; Disconnection
ACCORDING TO THE REPORTER, DURING USE, THE CANNULA WAS EASILY POPPED OFF FROM THE PATIENT'S CONNECTOR. RE-CANNULATION AND REPLACEMENT OF THE TUBE WAS NOT REQUIRED BECAUSE A TRACHEOSTOMY TIE WAS USED TO SECURE THE TUBE UNTIL THE NEXT ROUTINE TRACHEOSTOMY CHANGE WAS REQUIRED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19624914 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
- Shiley 6FENCovidien LP · Tube tracheostomy and tube cuff1 recall106 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page