Cook G06339Recalls & reported problems
Cook Incorporated · Thal-Quick Abscess Drainage Set · Class 1 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002063396), received between Mar 28, 2018 and Jan 13, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Detachment of Device or Device Component1
- Device Contamination with Chemical or Other Material1
- Entrapment of Device1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
UPON THE NURSE OPENING THE PACKAGE OF A COOK THAL-QUICK ABSCESS DRAINAGE SET, IT WAS DISCOVERED THAT THE DEVICE "FELL APART." NO PATIENT CONTACT WAS MADE. ADDITIONAL INFORMATION HAS BEEN REQUESTED REGARDING HOW THE DEVICE "FELL APART," BUT HAS NOT BEEN PROVIDED AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13254708 ↗MalfunctionMar 28, 2018Break; Entrapment of Device; Device Contamination with Chemical or Other Material; Material Deformation
IT WAS REPORTED THAT THE DILATOR OF THE THAL-QUICK ABSCESS DRAINAGE SET BECAME STUCK IN THE DRAINAGE CATHETER. THIS DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER SET. THERE WERE NO INJURIES OR ADDITIONAL PROCEDURES. THE DEVICE WAS RETURNED AND EXAMINED, REVEALING THAT IT WAS ACTUALLY THE DILATOR WITH THE GUIDEWIRE LODGED INSIDE OF IT, ALONG WITH AN UNIDENTIFIED HARDENED FOREIGN MATTER. IT WAS ALSO FOUND THAT THE SAFETY WIRE WAS BROKEN, AND THE COILS WERE ELONGATED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7377082 ↗
Other CATHETER, IRRIGATION devices
- Navarre Opti-Drain NOD8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION901 reports
- Navarre NNU8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION736 reports
- BLAKE 2227Ethicon Inc. · CATHETER, IRRIGATION331 reports
- Navarre NNU6LPTBard Access Systems, Inc. · CATHETER, IRRIGATION211 reports
- Navarre Opti-Drain NOD10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION178 reports
- Navarre NNU10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION127 reports
- BLAKE 2233Ethicon Inc. · CATHETER, IRRIGATION114 reports
- Vasoshield VH-5001Maquet Cardiovascular LLC · Catheter, Irrigation1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page