Skip to content

Roadrunner G07518Recalls & reported problems

Cook Incorporated · Roadrunner, PC Wire Guide The Firm · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002075184), received between Jan 26, 2018 and May 9, 2024 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Device Contaminated During Manufacture or Shipping” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 4 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 19, 2022Open, Classified

    Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

    Z-1709-2022 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Device Contaminated During Manufacture or Shipping4
  • Contamination /Decontamination Problem3
  • Material Separation1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Patient Involvement2
  • No Code Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 9, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19278574 ↗
  • MalfunctionMay 9, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19278438 ↗
  • MalfunctionMay 8, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT ONE UNIT OF A ROADRUNNER THE FIRM HYDROPHILIC WIRE GUIDE WAS NOTED TO BE DIRTY UPON ARRIVAL AND INSPECTION AT THE DISTRIBUTOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19272138 ↗
  • MalfunctionMay 8, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    AS REPORTED BY THE DISTRIBUTOR, TWO ROADRUNNER THE FIRM HYDROPHILIC WIRE GUIDES WERE FOUND TO HAVE "DIRTY GUIDEWIRES" WHEN THE PRODUCT ARRIVED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19271769 ↗
  • MalfunctionApr 24, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY A DISTRIBUTOR, FOREIGN MATTER RESEMBLING A HAIR WAS NOTED IN THE PACKAGES OF TWO ROADRUNNER THE FIRM HYDROPHILIC WIRE GUIDES. THERE WAS NO PATIENT INVOLVEMENT, AS THE FOREIGN MATTER WAS FOUND WHEN THE PRODUCTS ARRIVED AT THE FACILITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19174244 ↗
  • MalfunctionMay 29, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Patient Involvement

    MAUDE report 10101464 ↗
  • MalfunctionMay 4, 2020Peeled/Delaminated; Peeled/Delaminated

    AS REPORTED, DURING AN UNSPECIFIED PROCEDURE TO REMOVE A CLOT, A ROADRUNNER THE FIRM HYDROPHILIC WIRE GUIDE PEELED. THE DEVICE WAS USED WITH AN UNKNOWN BALLOON. AFTER INFLATION, THE USER REMOVED THE BALLOON, AT WHICH POINT THE WIRE GUIDE'S "PEELING" CAME OUT WITH THE BALLOON. ANOTHER DEVICE OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10020766 ↗
  • MalfunctionJul 8, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Patient Involvement

    MAUDE report 8768955 ↗
  • InjuryJan 26, 2018Material Separation; Material Separation

    IT WAS REPORTED THAT DURING ENDOVASCULAR SURGERY, THE ROADRUNNER THE FIRM HYDROPHILIC WIRE GUIDE CAME OUT OF THE LUMEN OF ANOTHER MANUFACTURER'S 5 FR CATHETER. WHEN THE PHYSICIAN TRIED TO REPOSITION THE WIRE GUIDE, THE TIP WAS OBSERVED TO BE BROKEN. THE PROCEDURE WAS CONVERTED TO OPEN SURGERY AS THE LESION COULD NOT BE RE-CANALIZED AND THE BROKEN TIP PORTION REMAINED INSIDE OF THE PATIENT'S ANATOMY. A BY-PASS WAS PERFORMED AND THE TIP PORTION WAS UNABLE TO BE RETRIEVED. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Code Available

    MAUDE report 7221917 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.