Cook TPN G07960Recalls & reported problems
Cook Incorporated · Cook TPN, Single Lumen Catheter Set With 3.5Fr Peel-Away, Introducer Set · Class 2 · Implant
All models of Cook TPN (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002079601), received between Apr 17, 2015 and May 12, 2020 — classified as 5 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break4
- Fitting Problem2
- Adverse Event Without Identified Device or Use Problem1
- Fluid/Blood Leak1
- Hole In Material1
- Leak/Splash1
- Material Puncture/Hole1
- Material Split, Cut or Torn1
Patient problems
- No Consequences Or Impact To Patient3
- No Code Available2
- No Known Impact Or Consequence To Patient2
- Reinfusion1
- Surgical procedure, additional1
- Swelling1
- Unspecified Infection1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 12, 2020Break; Fitting Problem; Break; Fitting Problem
A DEVICE FAILURE ANALYSIS WAS CONDUCTED ON 01JUN2020 FOR THE RETURNED DEVICE AND FOUND THAT IN ADDITION TO THE BREAK IN THE LINE, THE SILICONE SLEEVE HAS COME LOOSE FROM THE WINGED HUB, WHICH WAS NOT REPORTED BY THE CUSTOMER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10048335 ↗MalfunctionMay 4, 2020Fitting Problem; Material Split, Cut or Torn; Fitting Problem; Material Split, Cut or Torn
NO ADDITIONAL PATIENT/EVENT INFORMATION HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10020769 ↗InjuryAug 20, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PROCEDURE PERFORMED WAS A LONG TERM CVC PLACEMENT IN THE SUBCLAVIAN VEIN. NO SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 8911930 ↗InjuryDec 21, 2018Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED A REDO SINGLE LUMEN TPN CATHETER SET WAS USED TO REPAIR A COOK CATHETER DELIVERING TPN AND IRON INFUSIONS. THE COOK CATHETER WAS SAID TO HAVE "POPPED" ON (B)(6) 2018 DURING A SALINE FLUSH, WHICH CAUSED A HOLE. THE LINE WAS LOCKED WITH HEPARIN AND A CULTURE WAS TAKEN. THE CULTURE TESTED POSITIVE FOR INFECTION AND ANTIBIOTIC TREATMENT WAS GIVEN. AS REPORTED, THE PATIENT DID NOT REQUIRE AN ADDITIONAL PROCEDURE TO REPLACE THE LINE AS THE LINE WAS REPAIRED WITH THE REPAIR KIT.
Patient problems: Unspecified Infection
MAUDE report 8192167 ↗InjuryJul 19, 2017Break; Break
INTERNATIONAL CUSTOMER REPORTED THAT A PEDIATRIC PATIENT RECEIVED A REDO SINGLE LUMEN TPN CATHETER ON AN UNSPECIFIED DATE FOR PARENTERAL NUTRITION. THE NURSING STAFF EARLIER CHECKED THE CATHETER AND THERE WAS NO ABNORMALITIES. THE FATHER FELT RESISTANCE AND HEARD A "CLAC" WHILE HOLDING THE CHILD. THE NURSING STAFF WAS ADVISED OF THIS BY THE FATHER AND CLAMPED THE CATHETER WITH A NON-STERILE CLAMP AND USED COMPRESSES WITH CHLORHEXIDINE. THE ACTIONS TAKEN BY THE CLINICAL STAFF FOLLOWING THIS EVENT ARE NOT KNOWN. ADDITIONAL INFORMATION WAS REQUESTED; NO FURTHER INFORMATION HAS BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Patient problems: No Code Available
MAUDE report 6725744 ↗MalfunctionJul 19, 2017Leak/Splash; Leak/Splash
INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT RECEIVED A REDO SINGLE LUMEN TPN CATHETER IMPLANTED ON (B)(6) 2017 FOR AN UNSPECIFIED INDICATION. THE PATIENT UNDERWENT INFUSION OF CONTRAST MEDIA DURING A THERAPY PROCEDURE ON (B)(6) /2017. LEAKAGE OF CONTRAST FLUID FROM THE CENTRAL VENOUS CATHETER WAS OBSERVED UNDER X-RAY VIEWING DURING THE PROCEDURE. THE LEAK WAS OBSERVED AROUND THE DEVICE HUB AND RIGHT JUGULAR VEIN. THE CUSTOMER REPORTS THAT "NO TRACTION OR OTHER TRAUMATIC ACTION WERE DONE." THERE ARE NO REPORTED ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL PROCEDURES OR OTHER INTERVENTION WERE REQUIRED. NO FURTHER INFORMATION WAS PROVIDED. THE DEVICE IS REPORTEDLY AVAILABLE FOR EVALUATION; HOWEVER IT HAS NOT BEEN RECEIVED BY THE MANUFACTURER AS OF THE DATE OF THIS REPORT. THE MINISTRY OF HEALTH FORM WAS RECEIVED FROM THE NATIONAL COMPETENT AUTHORITY (ITALY) AND CONTAINED THE FOLLOWING INFORMATION: "CVC POSITIONED ON THE (B)(6) 2017. ON THE (B)(6) 2017 AN X-RAY TO THE THORAX WAS PERFORMED. AT RADIOSCOPIC EXAMINATION DURING THE ADMINISTRATION OF AROUND 1.5ML OF MDC THROUGH A CVC WE OBSERVED A SPILLAGE OF MDC AROUND THE CATHETER BOTH BY THE INSERTION AT THE RIGHT JUGULAR VEIN LEVEL AN
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6725654 ↗MalfunctionJun 27, 2017Break; Break
CUSTOMER REPORTED THAT A PATIENT RECEIVED A REDO SINGLE LUMEN TPN CATHETER ON AN UNSPECIFIED DATE. THE CUSTOMER REPORTED THAT THE CENTRAL LINE IS "ABOUT TO BREAK". PEDIATRIC SURGERY WAS CALLED AND THE LINE WAS REPAIRED AT BEDSIDE. THERE WAS NO MEDICAL OR SURGICAL INTERVENTION TO ADDRESS THIS EVENT. THE CIRCUMSTANCES AND HANDLING CONDITIONS THAT LEAD UP TO THE EVENT ARE NOT KNOWN. THE DEVICE IS STILL IN CLINICAL USE, ACCORDINGLY IT IS NOT AVAILABLE FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6670050 ↗InjuryJun 9, 2017Break; Break
INTERNATIONAL CUSTOMER REPORTED THAT A REDO SINGLE LUMEN TPN CATHETER " BROKE DOWN BECAUSE OF TRACTION." AN ATTEMPT TO REPAIR THE DEVICE WAS NOT SUCCESSFUL; THE TIP OF THE DEVICE RUPTURED. THE PATIENT WILL REQUIRE AN EXPLANT AND REPLACEMENT OF THE DEVICE. IT IS NOT KNOWN IF THE EXPLANT AND REPLACEMENT WAS PERFORMED. NO FURTHER EVENT OR PATIENT DETAILS WERE PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6629478 ↗InjuryApr 17, 2015Fluid/Blood Leak; Hole In Material
AFTER CONNECTING AN INFUSION PUMP TO A LONG TERM CVC LINE ON (B)(6) 2015 ON A (B)(6) FEMALE PT. THE PHYSICIAN NOTICED SWELLING WHERE THE LINE IS TUNNELED THE NEXT MORNING. THE INFUSION WENT TO THE TISSUES. THE PHYSICIAN TOOK THE LINE OUT AND EXAMINED THE CATHETER. THERE WERE NO VISIBLE HOLES IN THE DEVICE WHEN THE CATHETER WAS FLUSHED. IT WORKED WELL WITH NO SIGNS OF LEAKAGE. THE PHYSICIAN THEN PUT A CLAMP AT THE TIP OF THE CATHETER TO CLOSE IT AND INJECTED SALINE IN THE CATHETER. THIS TIME THEY SAW THAT THE SALINE COMING OUT FROM 2 SMALL HOLES ON THE CATHETER. THE SWELLING HAS BEEN ABSORBED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT UNDERWENT GENERAL ANESTHESIA FOR PLACEMENT OF ANOTHER CVD LINE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Reinfusion; Surgical procedure, additional; Swelling
MAUDE report 4713513 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page