Cook TPN G08074Recalls & reported problems

Cook Incorporated · Cook TPN, Double Lumen Catheter Set With 13Fr Peel-Away, Introducer Set · Class 2 · Implant

All models of Cook TPN (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002080744), received between Aug 3, 2017 and Aug 3, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Detachment Of Device Component1
  • Kinked1
  • Split1

Patient problems

  • No Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 3, 2017Break; Detachment Of Device Component; Kinked; Split

    THE INFORMATION RECEIVED INDICATED THAT THE TIP OF THE DILATOR WAS DAMAGED DURING AN INSERTION ATTEMPT. THE REPORTER INDICATED THAT DUE TO THE EXTREME OBESITY OF PATIENT OR CURVATURE OF VEIN AND OBSTACLE OF CLAVICLE, THE APPROPRIATE TWISTING MOTION WAS USED, AT THIS POINT THE INTRODUCER KINKED AND THE TIP OF THE DILATOR WAS DAMAGED. EVALUATION OF THE RETURNED DEVICE INDICATED THE DILATOR HAD BEEN SPLIT BEGINNING AT THE TIP AND EXTENDING 3MM PROXIMALLY. THE DEVICE HAD BEEN SPLIT SO THAT ALMOST HALF THE TIP HAD DETACHED AT AN ANGLE, THE SECOND DILATOR HAD A 1MM SPLIT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PROCEDURE WAS COMPLETED WITH THE USE OF A NEW SET. ACCORDING TO THE INITIAL REPORTER THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE AND THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT.

    Patient problems: No Information

    MAUDE report 6764243 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices