Reuter G08159Recalls & reported problems
Cook Incorporated · Disposable Reuter Tip Deflecting Wire Guide Heparin Coated · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002081598), received between Nov 8, 2018 and Nov 8, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Split, Cut or Torn1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 8, 2018Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED, A MALE PATIENT OF UNKNOWN AGE UNDERWENT A CARDIAC CATHETERIZATION PROCEDURE IN WHICH A REUTER DISPOSABLE TIP DEFLECTING WIRE GUIDE WAS TO BE USED. A KINKED AND SPLIT AREA WAS OBSERVED WHEN THE DEVICE WAS REMOVED FROM ITS PACKAGING. THE DEVICE DID NOT MAKE PATIENT CONTACT. NO PATIENT ADVERSE EFFECT HAS BEEN REPORTED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8052821 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page