Skip to content

Quick-Core G08790Recalls & reported problems

Cook Incorporated · Quick-Core, Biopsy Needle Set · Class 2

All models of Quick-Core (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This is a Cook Quick-Core biopsy needle set. Most of the 83 reports describe needle parts that are difficult to remove or unscrew, sometimes requiring tools, and some led to a replacement needle being used. There are 4 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
83
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 83 adverse event reports that list this exact device (DI 00827002087903), received between Nov 14, 2016 and Oct 11, 2023 — classified as 4 injuries, 79 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (68 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Difficult to Remove68
  • Device Operates Differently Than Expected5
  • Device Contaminated During Manufacture or Shipping4
  • Adverse Event Without Identified Device or Use Problem2
  • Separation Failure2
  • Device Operational Issue1
  • Mechanical Jam1
  • Mechanical Problem1

Patient problems

  • No Known Impact Or Consequence To Patient40
  • No Patient Involvement18
  • No Clinical Signs, Symptoms or Conditions14
  • No Consequences Or Impact To Patient8
  • Hematoma2
  • No Code Available2
  • Fluid Discharge1

Recent adverse event reports

The 25 most recent of 83 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 11, 2023Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17916512 ↗
  • MalfunctionSep 8, 2022Difficult to Remove; Difficult to Remove

    IN ADDITIONAL INFORMATION RECEIVED ON 15SEP2022, IT WAS CONFIRMED THAT THE ISSUE WAS WITH THE COAXIAL INTRODUCER NEEDLE. CONTACT WITH THE PATIENT WAS MADE. AFTER THE BIOPSY CANNULA WAS INSERTED INTO THE PATIENT, IT WAS FOUND THAT THE NEEDLE CORE COULD NOT BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15384204 ↗
  • MalfunctionJul 27, 2022Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THE NEEDLE ASSEMBLY OF A QUICK-CORE COAXIAL BIOPSY NEEDLE SET WAS UNABLE TO BE UNSCREWED PRIOR TO A LUNG PUNCTURE PROCEDURE THAT OCCURRED (B)(6) 2022. THE USER HAD TO USE TOOLS TO UNSCREW THE TROCAR. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NEW LIKE DEVICE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15113604 ↗
  • MalfunctionJul 27, 2022Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT DIFFICULTY WAS EXPERIENCED WHEN ATTEMPTING TO REMOVE THE TROCAR FROM A QUICK-CORE COAXIAL BIOPSY NEEDLE SET PRIOR TO A SUPRACLAVICULAR LYMPH NODE PUNCTURE ON (B)(6) 2022. THE PUNCTURE NEEDLE WAS REPLACED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15113450 ↗
  • MalfunctionJul 27, 2022Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THE NEEDLE ASSEMBLY OF A QUICK-CORE COAXIAL BIOPSY NEEDLE SET WAS UNABLE TO BE UNSCREWED PRIOR TO A LUNG PUNCTURE PROCEDURE THAT OCCURRED ON (B)(6) 2022. THE USER HAD TO USE TOOLS TO UNSCREW THE TROCAR. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NEW, LIKE DEVICE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15113328 ↗
  • MalfunctionJun 1, 2022Difficult to Remove; Difficult to Remove

    IT WAS REPORTED A MALE PATIENT REQUIRED A QUICK-CORE COAXIAL BIOPSY NEEDLE SET FOR A PERCUTANEOUS LUNG BIOPSY PROCEDURE. PRIOR TO PATIENT CONTACT, THE INNER STYLET WAS DIFFICULT TO REMOVE FROM THE COAXIAL NEEDLE. ANOTHER SIMILAR DEVICE WAS USED INSTEAD TO SUCCESSFULLY COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14565639 ↗
  • MalfunctionJan 26, 2022Difficult to Remove; Difficult to Remove

    PRIOR TO PATIENT CONTACT, IT WAS DISCOVERED THAT THE STYLET OF THE COOK QUICK-CORE COAXIAL BIOPSY NEEDLE SET WAS SCREWED TOO TIGHT IN THE OUTER PART OF THE NEEDLE AND WAS UNABLE TO BE TAKEN APART. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13361300 ↗
  • MalfunctionOct 1, 2021Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE INNER STYLETS OF TWO QUICK-CORE COAXIAL BIOPSY NEEDLE SETS COULD NOT BE REMOVED FROM THE COAXIAL NEEDLES. THIS OCCURRED PRIOR TO PATIENT CONTACT. THE DEVICES WERE INTENDED TO BE USED FOR A NEEDLE BIOPSY IN THE PATIENT'S LIVER. THE FIRST TWO DEVICES THE USER ATTEMPTED TO USE EXPERIENCED THIS ISSUE, SO A THIRD WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE EFFECTS TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12560491 ↗
  • MalfunctionSep 28, 2021Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE INNER STYLET OF A QUICK-CORE COAXIAL BIOPSY NEEDLE SET COULD NOT BE REMOVED FROM THE COAXIAL NEEDLE. THIS OCCURRED PRIOR TO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12540636 ↗
  • MalfunctionSep 24, 2021Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12529534 ↗
  • MalfunctionAug 3, 2021Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12266560 ↗
  • MalfunctionJul 30, 2021Difficult to Remove; Difficult to Remove

    IT WAS REPORTED A QUICK-CORE COAXIAL BIOPSY NEEDLE SET WAS REQUIRED FOR AN UNKNOWN PROCEDURE. PRIOR TO PATIENT CONTACT, THE STYLET WAS SCREWED TOO TIGHT IN THE OUTER PART OF THE NEEDLE AND COULD NOT BE SEPARATED. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12256505 ↗
  • MalfunctionJul 16, 2021Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12181106 ↗
  • MalfunctionJun 10, 2021Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11975193 ↗
  • MalfunctionApr 20, 2021Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Fluid Discharge

    MAUDE report 11697570 ↗
  • MalfunctionFeb 22, 2021Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11361727 ↗
  • MalfunctionDec 21, 2020Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Patient Involvement

    MAUDE report 11049645 ↗
  • MalfunctionDec 15, 2020Difficult to Remove; Difficult to Remove

    IT WAS REPORTED AN UNKNOWN PATIENT REQUIRED A QUICK-CORE COAXIAL BIOPSY NEEDLE SET FOR AN UNKNOWN PROCEDURE. PRIOR TO PATIENT CONTACT, THE OPERATOR COULD NOT SEPARATE THE STYLET FROM THE OUTER PART OF THE NEEDLE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11012817 ↗
  • MalfunctionNov 6, 2020Difficult to Remove; Difficult to Remove

    IT WAS REPORTED A MALE PATIENT REQUIRED A QUICK-CORE COAXIAL BIOPSY NEEDLE SET FOR A LUNG BIOPSY PROCEDURE. PRIOR TO PATIENT CONTACT, THE STYLET COULD NOT BE SEPARATED FROM THE NEEDLE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10802726 ↗
  • MalfunctionNov 3, 2020Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10777898 ↗
  • MalfunctionOct 23, 2020Difficult to Remove; Difficult to Remove

    IT WAS REPORTED AN UNKNOWN PATIENT REQUIRED A QUICK-CORE COAXIAL BIOPSY NEEDLE SET FOR AN UNKNOWN PROCEDURE. PRIOR TO PATIENT CONTACT, THE OPERATOR NOTED THE STYLET WAS SCREWED TOO TIGHT IN THE OUTER PART OF THE NEEDLE AND COULD NOT BE TAKEN APART. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10729445 ↗
  • MalfunctionOct 23, 2020Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10729276 ↗
  • MalfunctionOct 23, 2020Difficult to Remove; Difficult to Remove

    IT WAS REPORTED A MALE PATIENT REQUIRED A QUICK-CORE COAXIAL BIOPSY NEEDLE SET FOR A PUNCTURE BIOPSY OF A LUNG TUMOR. PRIOR TO PATIENT CONTACT, THE OPERATOR NOTED THE STYLET WAS SCREWED TOO TIGHT IN THE OUTER PART OF THE NEEDLE AND COULD NOT BE TAKEN APART. THIS FAILURE OCCURRED TWO MORE TIMES. A FOURTH DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10729229 ↗
  • MalfunctionSep 29, 2020Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10599884 ↗
  • MalfunctionSep 29, 2020Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10599780 ↗

FDA clearance and classification

The FDA regulates product code KNW (instrument, biopsy) under 21 CFR 876.1075 as a Class II device. 21 CFR 876.1075 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Quick-Core G08790

Has the Quick-Core G08790 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002087903), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Quick-Core G08790?

The FDA's MAUDE database lists 83 adverse event reports for this device, received between Nov 14, 2016 and Oct 11, 2023. In those reports, the most common specific device problems are “Difficult to Remove” (68), “Device Operates Differently Than Expected” (5) and “Device Contaminated During Manufacture or Shipping” (4); the most common patient problems are “Hematoma” (2) and “Fluid Discharge” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Quick-Core G08790?

None of the reports are classified as deaths, 4 as injuries and 79 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Quick-Core G08790, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Quick-Core, model G08790). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page