Beacon Tip Torcon NB G09312Recalls & reported problems
Cook Incorporated · Beacon Tip Torcon NB Advantage Catheter · Class 2
All models of Beacon Tip Torcon NB (69)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002093126), received between Jan 18, 2019 and Apr 23, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K173289); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 15, 2025Open, Classified
Cook Medical identified that catheters supplied in the affected device lots may experience tip separation.
Z-1902-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component2
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ADDITIONAL INFORMATION WAS RECEIVED 17MAY2024. THE INTENDED ACCESS SITE WAS THE FEMORAL ARTERY, BUT THE DEVICE DID NOT MAKE IT THAT FAR AS THE CATHETER HUB BROKE OFF BEFORE ENTERING THE PATIENT. THE HUB OF THE CATHETER BROKE "CLEAN OFF" OF THE CATHETER BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR ADVERSE EFFECTS DUE TO THE PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19162714 ↗MalfunctionJan 18, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN EMBOLIZATION PROCEDURE FOR GASTROINTESTINAL BLEEDING, THE HUB OF A BEACON TIP TORCON NB ADVANTAGE ANGIOGRAPHIC CATHETER SEPARATED FROM THE DEVICE. THE HUB WAS REPORTED TO SEPARATE WHILE THE DEVICE WAS BEING ADVANCED ONTO ANOTHER MANUFACTURER'S 0.035 GUIDE WIRE WITH LITTLE FORCE. THE DEVICE DID NOT MAKE PATIENT CONTACT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER UNKNOWN CATHETER. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8262259 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K173289. One submission often covers many models, so check the product and lot details in each record.
- Z-1936-2026 · Apr 2, 2026 · Open, ClassifiedBeacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page