Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002095939), received between Mar 15, 2018 and Mar 15, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unraveled Material” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Unraveled Material1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 15, 2018Unraveled Material; Unraveled Material

    THE CUSTOMER REPORTED THAT, UPON UNCAPPING AN EXISTING INTRAVENOUS LINE (IV) IN THE RIGHT ANTECUBITAL (AC) VEIN DURING A RIGHT HEART CATHETERIZATION WITH A RIGHT AC VEIN APPROACH, THE PHYSICIAN NOTED THAT THERE WAS NO BLOOD RETURN. THEY THEN PROCEEDED TO ATTEMPT TO PUT THE COPE STAINLESS STEEL MANDRIL WIRE GUIDE THROUGH THE IV WITH THE ASSISTANCE OF FLUOROSCOPIC IMAGING, BUT THE TIP OF THE WIRE APPEARED TO HAVE BUNCHED UP. WHEN REMOVING THE WIRE, THE PHYSICIAN REPORTED THAT IT FELT AS IF THE WIRE WAS COMING OUT, BUT THE TIP DID NOT APPEAR TO BE MOVING ON THE FLUOROSCOPIC IMAGE. WHEN THE WIRE WAS FINALLY REMOVED, IT WAS DISCOVERED THAT THE TIP HAD UNRAVELED. THE CUSTOMER REPORTED THAT THE WIRE GUIDE HAD BEEN KEPT FLUSHED AND WET WITH HEPARINIZED SALINE. THE CUSTOMER CONFIRMED THAT NO PATIENT ADVERSE EVENTS RESULTED FROM THE PRODUCT PROBLEM, AND NO ADDITIONAL PROCEDURES WERE REQUIRED AS A RESULT; HOWEVER, THE PHYSICIAN DID ELECT TO PURSUE FEMORAL ACCESS FOR THE RIGHT HEART CATHETERIZATION, AND THE IV LINE WAS REMOVED. ADDITIONAL INFORMATION HAS BEEN REQUESTED FORM THE CUSTOMER, BUT NONE HAS YET BEEN PROVIDED. THE DEVICE HAS BEEN RETURNED FOR EVALUATION; HOWEVER, AS OF THE DATE OF THI

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7343785 ↗