Roadrunner G09608Recalls & reported problems

Cook Incorporated · Roadrunner, PC Wire Guide Nimble, · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002096080), received between Sep 26, 2018 and Jun 4, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 19, 2022Open, Classified

    Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

    Z-1707-2022 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Difficult to Advance3
  • Break1
  • Difficult to Remove1
  • Material Split, Cut or Torn1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2024Break; Unraveled Material; Break; Unraveled Material

    AS REPORTED, DURING AN UNKNOWN PROCEDURE, FOUR ROADRUNNER NIMBLE HYDROPHILIC WIRE GUIDES ¿BROKE AT THE DISTAL END DURING USE.¿ REPORTEDLY, THE WIRES BECAME PROGRESSIVELY ¿MORE DAMAGED¿ AT THE DISTAL END. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19463429 ↗
  • MalfunctionMay 9, 2022Difficult to Advance; Difficult to Advance

    AS REPORTED, DURING A PROCEDURE INVOLVING A THROMBECTOMY VIA RIGHT GROIN APPROACH, AN UNSPECIFIED SHEATH WOULD NOT ADVANCE OVER A ROADRUNNER NIMBLE HYDROPHILIC WIRE GUIDE. THE WIRE WAS REPLACED WITH AN UNKNOWN STIFF WIRE TO COMPLETE THE PROCEDURE. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14338345 ↗
  • MalfunctionJan 27, 2022Difficult to Advance; Difficult to Advance

    AS REPORTED, THIRTY ROADRUNNER NIMBLE HYDROPHILIC WIRE GUIDES WERE STICKY AND DIFFICULT TO ADVANCE. THIS OCCURRED DURING FISTULAGRAMS, THROMBECTOMIES, AND DURING CATHETER INSERTIONS. THERE HAS BEEN NO REPORT THAT ANY PATIENT REQUIRED ADDITIONAL PROCEDURES OR EXPERIENCED ANY ADVERSE EFFECTS FROM THESE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13367772 ↗
  • InjuryJan 27, 2022Difficult to Advance; Difficult to Advance

    NO ADDITIONAL INFORMATION HAS BEEN RECEIVED SINCE THE PREVIOUS REPORT WAS SUBMITTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 13367686 ↗
  • MalfunctionSep 26, 2018Difficult to Remove; Material Split, Cut or Torn

    THE COMPLAINANT REPORTED THAT THE ROADRUNNER NIMBLE HYDROPHILIC WIRE GUIDE WOULD NOT COME OUT OF THE HOLDER, AND IT WAS SPLIT AFTER IT WAS FINALLY REMOVED. IT WAS FURTHER NOTED THAT A SECTION OF THE POLYMER MATERIAL WAS MISSING FROM THE DISTAL END OF THE WIRE WHICH RESULTED IN EXPOSURE OF THE INNER MANDRIL WIRE. THIS PRODUCT ISSUE OCCURRED PRIOR TO PATIENT CONTACT, AND IT DID NOT RESULT IN ANY REPORTED IMPACT ON THE PROCEDURE.

    Patient problems: No Patient Involvement

    MAUDE report 7909524 ↗