Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002097513), received between Jul 22, 2022 and Jul 22, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K173289); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- May 15, 2025Class IOpen, Classified
Cook Medical identified that catheters supplied in the affected device lots may experience tip separation.
Z-1902-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 22, 2022Break; Break
AS REPORTED, UPON SUCCESSFUL COMPLETION OF A FISTULA DE-CLOTTING PROCEDURE IN THE UPPER ARM, THE TIP OF A BEACON TIP TORCON NB ADVANTAGE ANGIOGRAPHIC CATHETER SEPARATED AND REMAINS IN THE PATIENT. PER THE USER, THE VESSEL IS TOO SMALL TO RETRIEVE THE DEVICE, WHICH REPORTEDLY WILL BE UNABLE TO MIGRATE DUE TO THE SMALL SIZE OF THE VESSEL. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Foreign Body In Patient
MAUDE report 15085921 ↗
FDA clearance and classification
Cleared through Special 510(k) K173289 on Nov 15, 2017 (decision: Substantially Equivalent). Submitted by Cook Incorporated as “Beacon Tip Torcon NB Advantage Catheter, Slip-Cath Beacon Tip Catheter and Shuttle Select Slip-Cath Catheter, Beacon Tip Centimeter Sizing Catheter, Beacon Tip Cava Vessel Sizing Catheter, Beacon Tip White Vessel Sizing Catheter”. The FDA review took 30 days. FDA record ↗
The FDA regulates product code DQO (catheter, intravascular, diagnostic) under 21 CFR 870.1200 as a Class II device. 21 CFR 870.1200 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K173289. One submission often covers many models, so check the product and lot details in each record.
- Z-1936-2026 · Apr 2, 2026 · Open, ClassifiedBeacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20
Common questions about the Beacon Tip Torcon NB G09751
Has the Beacon Tip Torcon NB G09751 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1902-2025 (Class I), began on May 15, 2025: Cook Medical identified that catheters supplied in the affected device lots may experience tip separation. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Beacon Tip Torcon NB G09751?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jul 22, 2022. That report lists device problem “Break” (1); patient problem “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Beacon Tip Torcon NB G09751?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Beacon Tip Torcon NB G09751 FDA approved?
It was cleared through Special 510(k) K173289 on Nov 15, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Beacon Tip Torcon NB G09751, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand Beacon Tip Torcon NB, model G09751). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page