Amplatz G09956Recalls & reported problems
Cook Incorporated · Amplatz Ultra Stiff Wire Guide · Class 2
All models of Amplatz (42)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002099562), received between Oct 5, 2017 and Nov 6, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove1
- Flaked1
- Material Deformation1
- Material Frayed1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 6, 2025Flaked; Flaked
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23487463 ↗MalfunctionJan 31, 2024Material Separation; Material Separation
AS REPORTED, PRIOR TO PATIENT CONTACT FOR THE DIAGNOSTIC PORTION OF AN UNSPECIFIED PROCEDURE, AN AMPLATZ ULTRA-STIFF SUPPORT WIRE GUIDE "FELT GRITTY" UPON OPENING THE PACKAGE. THE WIRE WAS NOT USED. A SECOND AMPLATZ ULTRA-STIFF SUPPORT WIRE GUIDE REPORTEDLY FELT SIMILAR BUT WAS WIPED OFF AND FELT BETTER; THEREFORE, THE SECOND WIRE WAS USED FOR THE PROCEDURE WITHOUT ISSUE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18617128 ↗MalfunctionNov 2, 2017Material Frayed; Difficult to Remove; Material Frayed; Difficult to Remove
IT WAS REPORTED THAT THE AMPLATZ ULTRA-STIFF SUPPORT WIRE GUIDES FRAYED AND GOT STUCK IN THE CATHETER HUBS AS WELL AS A CLOSURE DEVICE. THE PHYSICIAN REPORTED THE THEY ENCOUNTERED RESISTANCE DURING PERFORMANCE OF THE COMPLAINT DEVICES, AND HAD TO PULL OUT THE WIRE GUIDES AND CATHETERS TOGETHER. THE WIRE GUIDES WERE REPORTED TO HAVE BUMPY SILVER SPOTS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6997746 ↗MalfunctionOct 5, 2017Material Deformation; Material Deformation
AN AMPLATZ ULTRA-STIFF SUPPORT WIRE GUIDE WAS USED IN A LOWER EXTREMITY ANGIOGRAM. IT WAS REPORTED THAT THE WIRE WAS OVER THE AORTIC BIFURCATION, AND THE CATHETER WAS ADVANCED OVER THE WIRE WHEN A "BUMP" WAS FELT IN THE MIDDLE OF THE WIRE. THE PHYSICIAN ATTEMPTED TO CUT THE BACK END OF THE WIRE WHICH WAS UNSUCCESSFUL. BOTH PRODUCTS WERE REMOVED AND A NEW WIRE WAS ADVANCED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. UPON RETURN OF THE DEVICE, IT WAS NOTED THAT THE COIL IS TO BE SLIGHTLY EXPANDED, THE WIRE WAS SEVERELY KINKED AND NO COATING WAS PRESENT AT 24.0CM FROM THE PROXIMAL END, AND THE COIL WIRES WERE PROTRUDING DUE TO THE SEVERE DAMAGE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6918796 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page