Skip to content

Cook G10643Recalls & reported problems

Cook Incorporated · Pericardiocentesis Set · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002106437), received between Jun 14, 2017 and Mar 28, 2018 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Mar 6, 2020Class IITerminated

    The printed expiration dates are incorrect.

    Z-1565-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Remove1
  • Failure to Advance1
  • Leak/Splash1

Patient problems

  • No Known Impact Or Consequence To Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 28, 2018Failure to Advance

    IT WAS REPORTED THAT THE DEVICE WAS USED FOR PERICARDIAL DRAINAGE AND THE PHYSICIAN WAS UNABLE TO ADVANCE THE CATHETER OVER THE WIRE GUIDE. HOWEVER, THE PHYSICIAN REMOVED THE CATHETER AND WIRE TOGETHER AS A UNIT SINCE THERE WAS INABILITY TO ADVANCE THE CATHETER. THE WIRE GUIDE WAS REPLACED WITH ANOTHER MANUFACTURER'S WIRE GUIDE, THEN THE SAME CATHETER COULD ADVANCE OVER THE REPLACEMENT AND COULD BE PLACED FOR DRAINAGE SUCCESSFULLY. IT WAS FURTHER NOTED THAT THE COMPLAINT WIRE GUIDE WAS ELONGATED AND THE USER MAY HAVE TRIED TO REMOVE THE CATHETER WHILE THE WIRE GUIDE REMAINED IN THE PATIENT. THERE WERE NO REPORTED ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE PRODUCT PROBLEM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7378149 ↗
  • MalfunctionDec 12, 2017Difficult to Remove; Difficult to Remove

    THE CUSTOMER REPORTED THAT THE WIRE GUIDE IN THE PERICARDIOCENTESIS CATHETER SET WAS PLACED INTO THE PATIENT'S PERICARDIAL SAC AT THE APEX, AND THE CATHETER WAS ADVANCED OVER THE WIRE GUIDE. THEN, THE USER ATTEMPTED TO REMOVE THE WIRE GUIDE; HOWEVER, THE WIRE GUIDE GOT CAUGHT WITHIN THE CATHETER AND COULD NOT BE REMOVED. THEN, THE COIL OF THE WIRE GUIDE GOT ELONGATED. THE WIRE GUIDE AND CATHETER WERE REMOVED FROM THE PATIENT TOGETHER, AND ANOTHER DEVICE WAS USED INSTEAD TO COMPLETE THE PROCEDURE WITHOUT PROBLEMS. THE DEVICE IS REPORTEDLY AVAILABLE FOR EVALUATION; HOWEVER, AS OF THE DATE OF THIS REPORT, NO DEVICE HAS YET BEEN RECEIVED FOR EVALUATION. ADDITIONAL INFORMATION RECEIVED IDENTIFIED THAT PER THE PHYSICIAN, "REGARDING ADVANCEMENT/REMOVAL OF THE WIRE GUIDE, THERE WAS CLEAR DIFFERENCE BETWEEN THE COMPLAINT DEVICE AND REPLACEMENT DEVICE. I THINK THERE WAS PROBLEM WITH THE WIRE GUIDE OR THE CATHETER".

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7107548 ↗
  • MalfunctionJun 14, 2017Leak/Splash; Leak/Splash

    INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT WAS UNDERGOING A PERICARDIOCENTESIS USING A PERICARDIOCENTESIS CATHETER SET. WHEN HE APPLIED NEGATIVE PRESSURE USING A SYRINGE AFTER PLACING THE CATHETER, SOME AIR CAME IN TO THE SYSTEM. BLOOD LEAKAGE WAS OBSERVED BETWEEN THE CATHETER AND CONNECTER. THE PHYSICIAN REMOVED AND REPLACED THE CATHETER TO COMPLETE THE PROCEDURE. THERE ARE NO REPORTED ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6641400 ↗

FDA clearance and classification

The FDA regulates product code GBX (catheter, irrigation) under 21 CFR 878.4200 as a Class I device. 21 CFR 878.4200 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Cook G10643

Has the Cook G10643 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-1565-2020 (Class II), began on Mar 6, 2020: The printed expiration dates are incorrect. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Cook G10643?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jun 14, 2017 and Mar 28, 2018. In those reports, the most common specific device problems are “Difficult to Remove” (1), “Failure to Advance” (1) and “Leak/Splash” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Cook G10643?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Cook G10643 FDA approved?

Devices of this type (product code GBX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.

Who makes the Cook G10643, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Cook, model G10643). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page