Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002107410), received between Aug 28, 2019 and Feb 19, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 21, 2019Terminated
The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.
Z-0468-2020 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Material Puncture/Hole1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2026Detachment of Device or Device Component
AS REPORTED, A TORCON NB ADVANTAGE ANGIOGRAPHIC CATHETER WAS USED DURING A CARDIAC CATHETERIZATION AND AORTIC RADIOGRAPHY PROCEDURE. DURING THE PROCEDURE, AN UNSPECIFIED POWER INJECTOR WAS USED TO DELIVER CONTRAST MEDIA AT A PRESSURE OF EIGHT HUNDRED POUNDS PER SQUARE INCH (PSI), WITH A VOLUME OF FORTY CUBIC CENTIMETERS (CC) OVER TWO SECONDS. THE HUB OF THE CATHETER SEPARATED FROM THE CATHETER BODY DURING THIS PROCESS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY REPLACING THE SEPARATED CATHETER WITH A NEW UNSPECIFIED DEVICE. THE SEPARATED CATHETER WAS MANUALLY REMOVED FROM THE PATIENT BY HAND. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24385581 ↗MalfunctionJan 7, 2026Detachment of Device or Device Component
ADDITIONAL INFORMATION WAS RECEIVED ON 20FEB2026. THE PROCEDURE INVOLVED AORTIC RADIOGRAPHY, AND THE CONTRAST MEDIA PRESSURE WAS AT 40CC FOR TWO SECONDS. THE HUB OF ONE DEVICE SEPARATED, NOT TWO AS ORIGINALLY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23993734 ↗MalfunctionDec 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ADDITIONAL INFORMATION WAS RECEIVED 14JAN2026. ACCESS WAS OBTAINED IN THE RIGHT FEMORAL ARTERY. THE CATHETER WAS ADVANCED THROUGH ANOTHER MANUFACTURER¿S SHEATH, OVER ANOTHER MANUFACTURER¿S WIRE GUIDE. THE ANATOMY WAS NOT STENOSED, TORTUOUS, OR CALCIFIED. BECAUSE PART OF THE CATHETER SHAFT WAS OUTSIDE OF THE BODY WHEN THE HUB SEPARATED, THE CATHETER SHAFT WAS REMOVED BY HAND. A NEW DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23936201 ↗MalfunctionAug 28, 2019Material Puncture/Hole; Stretched; Material Puncture/Hole; Stretched
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8940565 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.