Skip to content

Cook G11916Recalls & reported problems

Cook Incorporated · Centimeter Sizing Catheter · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Cook AuRous centimeter sizing catheter is a measuring device doctors use to size blood vessels before stent graft procedures. The 38 reports mostly describe cracked, ruptured or separating catheter tips, and sterile packaging found contaminated with foreign matter or unsealed, with many cases involving no patient contact. Reports mention 1 death and 2 injuries. One open recall, started in April 2026, covers marker bands that may crack or break apart, which can lengthen procedures or injure vessels.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
1
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 00827002119161), received between Mar 7, 2017 and Aug 8, 2025 — classified as 1 death, 2 injuries, 34 malfunctions.

The most frequently reported specific device problem is “Device Contaminated During Manufacture or Shipping” (9 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 4 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 2, 2026Open, Classified

    Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

    Z-1934-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Device Contaminated During Manufacture or Shipping9
  • Crack4
  • Material Rupture3
  • Material Separation3
  • Break2
  • Burst Container or Vessel2
  • Contamination /Decontamination Problem2
  • Fluid/Blood Leak2

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • No Known Impact Or Consequence To Patient12
  • No Consequences Or Impact To Patient5
  • No Patient Involvement4
  • Death1
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 8, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    AS REPORTED, DURING THE SCREENING OF AN AUROUS CENTIMETER VESSEL SIZING CATHETER, A FOREIGN BODY WAS FOUND IN THE PACKAGING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22752816 ↗
  • MalfunctionJul 17, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    AS REPORTED, DURING AN ENDOLUMINAL ISOLATION PROCEDURE IN THE THORACIC AORTA FOR A PATIENT WITH AORTIC DISSECTION, AN UNSPECIFIED BLACK FOREIGN MATTER WAS DISCOVERED INSIDE THE PACKAGING OF THE 'AUROUS CENTIMETER VESSEL SIZING CATHETER'. THE ISSUE WAS DISCOVERED WHEN THE USER OPENED THE PACKAGE AND NOTED THAT THERE WAS FALLEN HAIR ON THE OPERATING TABLE AND UPON INSPECTION, THE USER DISCOVERED THERE WAS STILL FOREIGN MATTER INSIDE THE PACKAGE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY REPLACING THE CATHETER. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ADVERSE EFFECTS DUE TO THIS EVENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE OF THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22531330 ↗
  • MalfunctionJul 16, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    AS REPORTED, DURING AN ENDOLUMINAL ISOLATION PROCEDURE IN THE THORACIC AORTA FOR A PATIENT WITH AORTIC DISSECTION, A HAIR-LIKE FOREIGN MATTER WAS FOUND UPON INSPECTION IN THE STERILE PACKAGING OF THE 'AUROUS CENTIMETER VESSEL SIZING CATHETER'. THE DEVICE DID NOT MAKE PATIENT CONTACT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY REPLACING THE CATHETER. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ADVERSE EFFECTS DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22523785 ↗
  • MalfunctionJul 3, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT A FOREIGN BODY WAS DISCOVERED INSIDE THE PACKAGING OF THE 'AUROUS CENTIMETER VESSEL SIZING CATHETER'. THE ISSUE WAS NOTED DURING DEVICE INSPECTION WITHIN A DISTRIBUTION FACILITY, THEREFORE, THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22394387 ↗
  • MalfunctionJul 3, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT A FOREIGN BODY WAS DISCOVERED INSIDE THE PACKAGING OF THE 'AUROUS CENTIMETER VESSEL SIZING CATHETER'. THE ISSUE WAS NOTED DURING DEVICE INSPECTION WITHIN A DISTRIBUTION FACILITY, THEREFORE, THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22393938 ↗
  • MalfunctionJun 25, 2025Unsealed Device Packaging; Unsealed Device Packaging

    AS REPORTED, DURING DEVICE INSPECTION UPON ARRIVAL TO A DISTRIBUTION FACILITY, AN AUROUS CENTIMETER VESSEL SIZING CATHETER'S STERILE PACKAGE WAS FOUND TO BE UNSEALED. THERE WAS NO PATIENT INVOLVEMENT, AS THIS WAS NOTED WITHIN A DISTRIBUTION FACILITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22331628 ↗
  • MalfunctionDec 27, 2024Break; Break

    AS REPORTED, THE TIP OF AN AUROUS CENTIMETER VESSEL SIZING CATHETER SEPARATED DURING ANGIOGRAPHY. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21026606 ↗
  • MalfunctionDec 13, 2024Unsealed Device Packaging; Unsealed Device Packaging

    AS REPORTED, UPON INSPECTION OF EIGHT AUROUS CENTIMETER VESSEL SIZING CATHETERS AT A HOSPITAL, THE SEAL WAS MISSING FROM ONE END OF THE PACKAGES. THERE WERE NO ADVERSE EFFECTS TO ANY PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20924150 ↗
  • MalfunctionNov 18, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20705316 ↗
  • MalfunctionJun 17, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19552156 ↗
  • MalfunctionNov 17, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    AS REPORTED BY THE DISTRIBUTOR, A HAIR WAS FOUND INSIDE THE PACKAGING OF A AUROUS CENTIMETER VESSEL SIZING CATHETER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18160436 ↗
  • MalfunctionSep 7, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17697701 ↗
  • MalfunctionMay 7, 2022Material Puncture/Hole; Material Rupture; Material Puncture/Hole; Material Rupture

    AS REPORTED, DURING AN UNKNOWN PROCEDURE, AN AUROUS CENTIMETER VESSEL SIZING CATHETER DEVELOPED A HOLE. THERE WAS REPORTEDLY NO TRAUMA TO THE CATHETER DURING PLACEMENT; HOWEVER, A HOLE WAS BLOWN IN THE CATHETER UPON THE FIRST USE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 14321747 ↗
  • MalfunctionJun 4, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED, DURING THE DISTRIBUTOR INSPECTION OF AN AUROUS CENTIMETER VESSEL SIZING CATHETER, AN UNKNOWN PARTICLE WAS NOTED INSIDE OF THE PRIMARY PACKAGING OF THE PRODUCT. THIS DEVICE DID NOT MAKE PATIENT CONTACT.

    Patient problems: No Patient Involvement

    MAUDE report 10118678 ↗
  • MalfunctionJan 17, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Patient Involvement

    MAUDE report 9602309 ↗
  • DeathJul 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Death

    MAUDE report 8822684 ↗
  • MalfunctionApr 26, 2019Separation Problem; Separation Problem

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8556167 ↗
  • MalfunctionDec 18, 2018Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT THE HUB OF THE AUROUS CENTIMETER VESSEL SIZING CATHETER HAD A CRACK ON IT, AND WHEN THE AND WHEN THE USER CONNECTED IT TO THE INJECTOR, "IT SPRAYED EVERYWHERE". ANOTHER DEVICE WAS USED FOR THE PROCEDURE, AND THERE WERE NO INJURIES OR ADDITIONAL PROCEDURES REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8176912 ↗
  • MalfunctionDec 18, 2018Crack

    IT WAS REPORTED THAT THE HUB OF THE AUROUS CENTIMETER VESSEL SIZING CATHETER CRACKED WHEN THE USER ATTEMPTED TO INJECT FLUID THROUGH IT DURING A PULMONARY EMBOLISM PROCEDURE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE, AND THERE WERE NO INJURIES OR ADDITIONAL PROCEDURES REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8176899 ↗
  • MalfunctionDec 18, 2018Crack

    IT WAS REPORTED THAT THE HUBS OF TWO OF THE AUROUS CENTIMETER VESSEL SIZING CATHETERS (OTHER DEVICE IN MDR 1820334-2018-03846) HAD CRACKED VERTICALLY, JUST AFTER THE CATHETER WAS CONNECTED TO THE POWER INJECTOR BUT BEFORE THE INJECTION TOOK PLACE. THERE HAD BEEN NO VISIBLE CRACKS DURING PREPARATION. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED, BUT THERE WERE NO INJURIES OR ADDITIONAL PROCEDURE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8176871 ↗
  • MalfunctionDec 18, 2018Crack

    IT WAS REPORTED THAT THE HUBS OF TWO OF THE AUROUS CENTIMETER VESSEL SIZING CATHETERS (OTHER DEVICE IN MDR 1820334-2018-03847) HAD CRACKED VERTICALLY, JUST AFTER THE CATHETER WAS CONNECTED TO THE POWER INJECTOR BUT BEFORE THE INJECTION TOOK PLACE. THERE HAD BEEN NO VISIBLE CRACKS DURING PREPARATION. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED, BUT THERE WERE NO INJURIES OR ADDITIONAL PROCEDURE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8175368 ↗
  • MalfunctionDec 17, 2018Break; Material Puncture/Hole

    THE COMPLAINANT REPORTED THAT A HOLE "BLEW OUT" AT THE PROXIMAL END OF THE AUROUS CENTIMETER VESSEL SIZING CATHETER DURING THE FINAL RUN OF THE PROCEDURE. THIS PART OF THE CATHETER WAS OUTSIDE OF THE BODY WHEN IT HAPPENED. THE PROCEDURE WAS FOR AN AAA (ABDOMINAL AORTIC ANEURYSM) REPAIR VIA THE GROINS AND ABDOMEN. THIS WAS A 6-7 HOUR PROCEDURE, AND THE CATHETER HAD BEEN INSERTED AND REMOVED SEVERAL TIMES. THERE WAS SUSPICION FOR A KINK AT THE LOCATION OF THE HOLE. IT WAS ALSO NOTED THAT A POWER INJECTOR WAS USED AT 1000 PSI. THIS PRODUCT PROBLEM DID NOT RESULT IN ANY REPORTED SERIOUS INJURIES TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8170056 ↗
  • MalfunctionOct 24, 2018Leak/Splash; Leak/Splash

    THE FOLLOWING COMPLAINT WAS RECEIVED VIA MEDWATCH MW5079740. IT WAS REPORTED THAT DURING A "VASCULAR ANGIOGRAPHY AND INTERVENTION" PROCEDURE, THE TUOHY-BORST LARGE BORE CLEAR PLASTIC SIDEARM ADAPTER LEAKED. ACCORDING TO THE INITIAL REPORTER, AN UNKNOWN 0.014 WIRE GUIDE WAS USED, AND THE COMPLAINT DEVICE TIGHTENED WELL AROUND THE UNKNOWN WIRE. WHEN ISOVUE 250 CONTRAST WAS INJECTED VIA THE LUER LOCK PORT, HOWEVER, CONTRAST SPRAYED AROUND THE WIRE AND INTO THE PHYSICIAN'S EYES. THERE HAVE BEEN NO ADVERSE EVENTS REPORTED REGARDING THIS COMPLAINT. THE COMPLAINT DEVICE IS NOT EXPECTED TO BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION, AND THERE HAS BEEN NO CONTACT INFORMATION PROVIDED TO REQUEST ADDITIONAL INFORMATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7999694 ↗
  • MalfunctionOct 8, 2018Fluid/Blood Leak; Fluid/Blood Leak

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7942465 ↗
  • MalfunctionSep 14, 2018Material Separation; Material Separation

    IT WAS REPORTED, THE PHYSICIAN CONFIRMED AIR ENTRY IN THE AUROUS CENTIMETER VESSEL SIZING CATHETER WHEN PERFORMING FINAL CONFIRMATORY ANGIOGRAPHY DURING THE STENT GRAFT PLACEMENT ALTHOUGH, IT COULD BE USED WITH NO PROBLEM BEFORE THE FINAL ANGIOGRAPHY. HE STOPPED USING THE DEVICE. THE DEVICE WAS REPLACED WITH ANOTHER AUROUS CENTIMETER VESSEL SIZING CATHETER TO COMPLETE THE PROCEDURE. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7878632 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.