Cook G14337Recalls & reported problems
Cook Incorporated · UPJ Occlusion Balloon Catheter Set · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002143371), received between Nov 27, 2018 and Mar 27, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Unsealed Device Packaging” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Unsealed Device Packaging3
- Contamination /Decontamination Problem2
- Device Contaminated During Manufacture or Shipping1
- Device Contamination with Chemical or Other Material1
- Fluid/Blood Leak1
- Inflation Problem1
- Material Rupture1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2025Material Rupture; Material Rupture
CORRECTION OF INITIAL AWARE DATE: 12MAR2025 NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21707247 ↗MalfunctionJul 2, 2024Unsealed Device Packaging; Unsealed Device Packaging
IT WAS REPORTED THE UPJ OCCLUSION BALLOON CATHETER'S PACKAGE WAS FOUND UNSEALED PRIOR TO AN UNSPECIFIED PROCEDURE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19661107 ↗MalfunctionAug 23, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17607305 ↗MalfunctionFeb 28, 2022Unsealed Device Packaging; Unsealed Device Packaging
AS REPORTED, THE PRIMARY PACKAGING OF THREE UPJ OCCLUSION BALLOON CATHETERS WERE FOUND TO BE "DETACHED FROM THE SEAL," BREAKING THE STERILE BARRIER. NO DAMAGE WAS OBSERVED TO THE OUTER PACKAGING. NO PATIENT WAS AFFECTED BY THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13620202 ↗MalfunctionDec 6, 2021Unsealed Device Packaging; Unsealed Device Packaging
THERE IS NO NEW PATIENT OR EVENT INFORMATION TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12935216 ↗MalfunctionApr 6, 2021Obstruction of Flow; Obstruction of Flow
THERE IS NO NEW PATIENT OR EVENT INFORMATION TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11618395 ↗MalfunctionJan 15, 2021Contamination /Decontamination Problem
AS REPORTED, DURING A PERCUTANEOUS NEPHROLITHOTRIPSY, AN UNKNOWN WHITE SUBSTANCE WAS FOUND IN A UPJ OCCLUSION BALLOON CATHETER. AFTER INSERTING THE BALLOON CATHETER INTO THE PATIENT, THE PHYSICIAN INFLATED THE BALLOON AND INSERTED A SYRINGE INTO THE INJECTION PORT OF THE INFLATION/INJECTION ADAPTER FOR PYELOGRAPHY. HOWEVER, THE PHYSICIAN COULD NOT INJECT UROGRAFIN CONTRAST MEDIUM, AS IT WOULD NOT FLOW THROUGH THE PORT OF THE ADAPTER. ONCE THE DEVICE WAS REMOVED FROM THE PATIENT, THE PHYSICIAN REMOVED THE SCREW OF THE INJECTION PORT OF THE ADAPTER AND CONFIRMED A WHITE BURR-LIKE SUBSTANCE WAS STUCK/CLOGGED IN THE DEVICE. ANOTHER PRODUCT OF THE SAME REFERENCE NUMBER WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11177822 ↗MalfunctionJan 7, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem
ADDITIONAL INFORMATION WAS RECEIVED 12JAN2021: WHILE ATTEMPTING TO ADJUST THE ADAPTOR MANY TIMES AND TIGHTENING IT STRONGLY, A PIECE OF THE ADAPTER BROKE OFF. WHILE REPLACING THE ADAPTOR, THE USER DISCOVERED THE FOREIGN SUBSTANCE IN THE PROXIMAL END OF THE BALLOON CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11132766 ↗MalfunctionNov 27, 2018Fluid/Blood Leak; Inflation Problem; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT HYDRONEPHROSIS WAS INTENTIONALLY INDUCED WITH THE BALLOON TO MAKE PERCUTANEOUS RENAL PUNCTURE EASIER FOR PERCUTANEOUS NEPHROLITHOTRIPSY (PNL) PROCEDURE. THE UPJ OCCLUSION BALLOON CATHETER WAS ADVANCED OVER A WIRE GUIDE AND WHEN THE BALLOON REACHED THE URETEROPELVIC JUNCTION (UPJ), THE PHYSICIAN ATTEMPTED TO INFLATE THE BALLOON AFTER CONNECTING A SYRINGE CONTAINED IN THE DEVICE SET TO THE DEVICE CONNECTOR. HOWEVER, THE BALLOON DID NOT INFLATE AND FLUID LEAKED FROM THE OPPOSITE END. IT WAS FURTHER NOTED THAT A DRIED WHITE SUBSTANCE WAS FOUND AROUND THE OBTURATOR AND THE BALLOON CATHETER. IN ADDITION, THE OBTURATOR/WIRE GUIDE WAS NOTED TO BE BROKEN AND STRETCHED OUT. THERE WERE NO ADVERSE EFFECTS REPORTED DUE TO THE EVENT AND NO REPORT OF AN ADDITIONAL PROCEDURE. A NEW PRODUCT WAS USED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8108596 ↗
FDA clearance and classification
The FDA regulates product code EYB (catheter, ureteral, gastro-urology) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Cook G14337
Has the Cook G14337 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002143371), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Cook G14337?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Nov 27, 2018 and Mar 27, 2025. In those reports, the most common specific device problems are “Unsealed Device Packaging” (3), “Contamination /Decontamination Problem” (2) and “Device Contaminated During Manufacture or Shipping” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Cook G14337?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Cook G14337, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand Cook, model G14337). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, URETERAL, GASTRO-UROLOGY devices
- MEDLINE DYND13508Medline Industries, Inc. · CATHETER, URETERAL, GASTRO-UROLOGY1 recall2 reports
- MEDLINE DYND13510Medline Industries, Inc. · CATHETER, URETERAL, GASTRO-UROLOGY1 recall1 report
- MEDLINE DYND13514Medline Industries, Inc. · CATHETER, URETERAL, GASTRO-UROLOGY1 recall1 report
- MEDLINE DYND13518Medline Industries, Inc. · CATHETER, URETERAL, GASTRO-UROLOGY1 recall1 report
- Medline Industries, Inc. DYNJ0040839Medline Industries, Inc. · CATHETER, URETERAL, GASTRO-UROLOGY1 recall0 reports
- MEDLINE DYND13520Medline Industries, Inc. · CATHETER, URETERAL, GASTRO-UROLOGY1 recall0 reports
- MEDLINE DYND13515Medline Industries, Inc. · CATHETER, URETERAL, GASTRO-UROLOGY1 recall0 reports
- MEDLINE DYND13614Medline Industries, Inc. · CATHETER, URETERAL, GASTRO-UROLOGY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page