Skip to content

Biodesign G23653Recalls & reported problems

Cook Biotech Incorporated · Biodesign Enterocutaneous Fistula Plug · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002236530), received between Nov 14, 2023 and Nov 14, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2023Material Separation; Material Separation

    ON (B)(6) 2023, A BIODESIGN ENTEROCUTANEOUS FISTULA PLUG WAS PLACED FOR REPAIR OF AN ENTEROCUTANEOUS FISTULA. THE PATIENT HAD THE ANAL CUTANEOUS FISTULA FOR OVER A YEAR. THE FISTULA EXITED OUT OF THE GLUTEUS. THERE WAS NO DISTAL OBSTRUCTION. THE PATIENT HAD NO HISTORY OF RECEIVING RADIATION. THE FISTULA DRAINED LESS THAN 20 ML/DAY. THE FISTULA TRACT WAS 11 CM IN LENGTH AND AROUND 2 MM IN DIAMETER. THERE WERE NO 4 MM PLUG SETS AVAILABLE AT THE IMPLANTING FACILITY. THE PHYSICIAN MADE THE DECISION TO IMPLANT A 7 MM PLUG. THE PLUG WAS SUCCESSFUL AT HEALING THE FISTULA HOWEVER, THE FLANGE WOULD NOT PASS ON ITS OWN WHICH CAUSED THE PATIENT TO EXPERIENCE PAIN. THE PAIN WAS NEAR THE PATIENT'S RECTUM AND OCCURRED WHEN PRESSURE WAS APPLIED. ON (B)(6) 2023, THE PHYSICIAN ATTEMPTED TO LAPAROSCOPICALLY REMOVE THE FLANGE. DUE TO THE LOCATION OF THE FLANGE,NEAR THE RECTUM RIGHT NEXT TO OR POSSIBLY ON THE SPHINCTER, THE FLANGE WAS REMOVED MANUALLY. THE PATIENT DID NOT EXPERIENCE ANY ADDITIONAL COMPLICATIONS OR HARM. UPON EXAMINATION OF THE EXPLANTED FLANGE, THE PHYSICIAN NOTICES THAT THE SILICONE HAD PARTIALLY SEPARATED FROM THE NITINOL RING.

    Patient problems: Pain

    MAUDE report 18134912 ↗

Other Mesh, surgical devices

All Mesh, surgical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.