Turbo-Ject G23902Recalls & reported problems
Cook Incorporated · Turbo-Ject PICC Set Power Injectable Polyurethane Peripherally Inserted Central Venous Catheter · Class 2 · Implant
All models of Turbo-Ject (40)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00827002239029), received between May 5, 2017 and Sep 23, 2019 — classified as 5 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Leak/Splash2
- Break1
- Detachment Of Device Component1
- Flaked1
- Fracture1
- Material Separation1
- Split1
- Unraveled Material1
Patient problems
- No Code Available5
- No Consequences Or Impact To Patient2
- Hematoma1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2019Flaked; Flaked
NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9105907 ↗InjuryMar 29, 2019Material Separation; Material Separation
NO NEW EVENT DESCRIPTION INFORMATION TO REPORT AT THIS TIME.
Patient problems: No Code Available
MAUDE report 8464446 ↗InjuryFeb 26, 2019Fracture; Fracture
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Code Available
MAUDE report 8371801 ↗InjuryFeb 20, 2018Split; Split
THE INTERNATIONAL CUSTOMER REPORTED THAT, AT AN UNSPECIFIED TIME FOLLOWING THE USE OF THE TURBO-JECT® DOUBLE LUMEN POWER-INJECTABLE PICC (PERIPHERALLY INSERTED CENTRAL CATHETER) FOR AN UNKNOWN INDICATION, A HEMATOMA APPEARED AT THE PATIENT'S ELBOW. THE CATHETER WAS SPLIT ALONG A SEGMENT THAT SPANNED FROM 2 CM TO 5 CM FROM THE DISTAL END OF THE DEVICE. A NEW DEVICE NEEDED TO BE PLACED. THE DEVICE HAS BEEN RETURNED FOR INVESTIGATION; HOWEVER, AS OF THE DATE OF THIS REPORT, THE DEVICE INVESTIGATION IS STILL PENDING. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Hematoma; No Code Available
MAUDE report 7283570 ↗InjurySep 20, 2017Leak/Splash
THE PICC WAS INSERTED THEN LEAKAGE AT HUB WAS NOTED AND THE PICC REPLACED. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY OTHER ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 6880309 ↗InjurySep 20, 2017Break; Detachment Of Device Component
APPROXIMATELY FORTY-FOUR DAYS AFTER THE PICC LINE WAS INSERTED, THE PURPLE HUB CAP SNAPPED OFF. THE PICC LINE WAS REPLACED. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY OTHER ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 6879431 ↗MalfunctionMay 5, 2017Unraveled Material; Unraveled Material
INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT UNDERWENT PLACEMENT OF A TURBO-JECT® DOUBLE LUMEN POWER-INJECTABLE PICC FOR AN UNSPECIFIED INDICATION. THE GUIDE WIRE WAS DAMAGED DURING REMOVAL, REPORTED AS "ELONGATION (DESTRUCTION)" FOLLOWING PLACEMENT OF THE CATHETER. THE PROCEDURE WAS COMPLETED WITHOUT ANY FURTHER ISSUES. THERE ARE NO REPORTED ADVERSE PATIENT CONSEQUENCES. THE DEVICE WAS RECEIVED BY THE MANUFACTURER FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6547010 ↗MalfunctionMay 5, 2017Leak/Splash; Leak/Splash
INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT UNDERWENT PLACEMENT OF A TURBO-JECT® DOUBLE LUMEN POWER-INJECTABLE PICC ON AN UNSPECIFIED DATE. THE PATIENT UNDERWENT ROUTINE TREATMENT INCLUDING DISINFECTION, BLOOD ASPIRATION, FLUSHING AND ANTIBIOTIC TREATMENT WITHOUT INCIDENT. THE PATIENT CONTACTED THE STAFF TO REPORT THAT BLOOD WAS LEAKING FROM THE NEEDLELESS CONNECTOR FOLLOWING TREATMENT. THE NURSE REPORTS THAT THE CONNECTOR WAS DAMAGED. THE CONNECTOR WAS REMOVED FROM THE PICC LINE AND REPLACED. IT IS NOT KNOWN IF THE CONNECTOR IS THE ORIGINAL CONNECTOR OR ONE THAT WAS REPLACED WITH A DRESSING CHANGE. NO FURTHER EVENT DETAILS WERE PROVIDED. THERE ARE NO REPORTED ADVERSE PATIENT CONSEQUENCES RELATED TO THIS EVENT NOR ANY REPORT OF A NEED FOR MEDICAL OR SURGICAL INTERVENTION TO PREVENT SERIOUS INJURY. THE DEVICES IS REPORTEDLY AVAILABLE FOR EVALUATION, HOWEVER IT HAS NOT BEEN RECEIVED BY THE MANUFACTURER AS OF THE DATE OF THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6546980 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page