Skip to content

Turbo-Ject G23904Recalls & reported problems

Cook Incorporated · Turbo-Ject PICC Set Power Injectable Polyurethane Peripherally Inserted Central Venous Catheter · Class 2 · Implant

All models of Turbo-Ject (40)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00827002239043), received between Jul 25, 2017 and May 13, 2018 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Crack” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Crack2
  • Leak/Splash2
  • Fracture1
  • Split1
  • Sticking1
  • Unraveled Material1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • No Code Available2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED A LEAK WAS NOTICED BELOW THE HUB OF A TURBO-JECT® DOUBLE LUMEN OVER-THE-WIRE POWER-INJECTABLE PICC WHILE A PATIENT WAS BEING ISSUED CHEMOTHERAPY. THE PICC LINE WAS REMOVED AND REPLACED WITH ANOTHER LIKE DEVICE TO CONTINUE TREATMENT. IT WAS REPORTED THAT WHEN IT WAS REMOVED THE NURSE FLUSHED BOTH LUMENS AND COULD NOT SEE A LEAK.

    Patient problems: No Code Available

    MAUDE report 7509705 ↗
  • InjuryOct 11, 2017Crack; Leak/Splash; Crack; Leak/Splash

    THE INTERNATIONAL CUSTOMER REPORTED THAT, WHILE THE TURBO-JECT® DOUBLE LUMEN OVER-THE-WIRE POWER-INJECTABLE PICC (PERIPHERALLY INSERTED CENTRAL CATHETER) WAS CONNECTED TO A PUMP FOR AN INFUSION, A LEAK/CRACK IN THE MIDDLE OF THE PURPLE HUB WAS OBSERVED. THE CIRCUMSTANCES SURROUNDING THE USAGE AND HANDLING OF THE DEVICE WERE NOT REPORTED. A NEW PICC LINE HAD TO BE INSERTED. ADDITIONAL INFORMATION HAS BEE REQUESTED FROM THE CUSTOMER. THE PRODUCT IS REPORTEDLY AVAILABLE FOR RETURN; HOWEVER, AS OF THE DATE OF THIS REPORT, NO PRODUCT HAS YET BEEN RECEIVED FOR EVALUATION.

    Patient problems: No Code Available

    MAUDE report 6941592 ↗
  • MalfunctionSep 18, 2017Fracture

    THE CUSTOMER REPORTED THAT DURING A PROCEDURE TO PLACE A PICC LINE, THE WIRE TIP SHEARED AND NEARLY FRACTURED DURING MANIPULATION. NO PARTS OF THE DEVICE WERE LEFT IN THE PATIENT, AND THERE WERE NO INJURIES OR ADDITIONAL PROCEDURES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6872918 ↗
  • MalfunctionAug 30, 2017Crack; Crack

    THE INTERNATIONAL CUSTOMER REPORTED THAT THE PURPLE HUB OF THE DEVICE APPEARED TO BE CRACKED, AND HAD A HOLE WHERE THE PORT CONNECTS TO THE LINE. NO PATIENT ADVERSE EVENTS WERE REPORTED BY THE CUSTOMER, AND THE CIRCUMSTANCES SURROUNDING THE USAGE AND HANDLING OF THE DEVICE ARE UNKNOWN. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE CUSTOMER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6831759 ↗
  • MalfunctionJul 28, 2017Sticking; Unraveled Material; Sticking; Unraveled Material

    INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT WAS UNDERGOING PLACEMENT OF A TURBO-JECT® DOUBLE LUMEN OVER-THE-WIRE POWER-INJECTABLE PICC FOR AN UNSPECIFIED INDICATION. THE CUSTOMER REPORTED OBSERVING UPON THE REMOVAL OF THE DEVICE THAT THE WIRE HAD BECOME STUCK AND UNRAVELLED. THE CUSTOMER OBTAINED A REPLACEMENT DEVICE AND COMPLETED THE PROCEDURE. THERE ARE NO REPORTED ADVERSE PATIENT CONSEQUENCES RELATED TO THIS EVENT. THE DEVICE IS REPORTEDLY AVAILABLE FOR EVALUATION; HOWEVER IT HAS NOT BEEN RECEIVED BY THE MANUFACTURER AS OF THE DATE OF THIS REPORT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6752298 ↗
  • MalfunctionJul 25, 2017Split; Split

    INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT WAS UNDERGOING PLACEMENT OF A TURBO-JECT® DOUBLE LUMEN OVER-THE-WIRE POWER-INJECTABLE PICC FOR AN UNSPECIFIED INDICATION. THE CUSTOMER REPORTED OBSERVING THAT THE DISTAL END OF THE CATHETER HAD A SPLIT. THE CUSTOMER OBTAINED A REPLACEMENT DEVICE AND COMPLETED THE PROCEDURE. THERE ARE NO REPORTED ADVERSE PATIENT CONSEQUENCES RELATED TO THIS EVENT. THE DEVICE IS REPORTEDLY AVAILABLE FOR EVALUATION; HOWEVER IT HAS NOT BEEN RECEIVED BY THE MANUFACTURER AS OF THE DATE OF THIS REPORT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6741547 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page