Amplatz G28781Recalls & reported problems
Cook Incorporated · Amplatz Stiff Wire Guide · Class 2
All models of Amplatz (42)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002287815), received between Sep 21, 2018 and Apr 19, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Unraveled Material” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Unraveled Material2
- Peeled/Delaminated1
- Structural Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2024Unraveled Material; Unraveled Material
AS REPORTED, DURING AN INFERIOR VENA CAVA (IVC) FILTER RETRIEVAL PROCEDURE, THE TIP OF AN AMPLATZ STIFF SUPPORT WIRE GUIDE UNRAVELED. ACCESS WAS OBTAINED AT THE INTERNAL JUGULAR (IJ) VEIN TO TARGET THE INFERIOR VENA CAVA (IVC). THE PATIENT'S ANATOMY WAS NOT STENOSED, TORTUOUS, CALCIFIED OR SCARRED. THE WIRE WAS NOT ALTERED FROM ITS ORIGINAL CONDITION PRIOR TO USE AND RESISTANCE WAS NOT FELT UPON INSERTION OF THE WIRE. UPON REMOVAL OF THE WIRE THROUGH AN UNSPECIFIED SHEATH, THE "DISTAL FLOPPING TIP OF THE WIRE CASING UNCOILED". THE WIRE WAS REMOVED FROM THE STERILE FIELD AND ANOTHER WIRE WAS USED TO COMPLETE THE PROCEDURE. THE WIRE WAS NOT MANIPULATED BACKWARDS THROUGH A NEEDLE OR OTHER METAL INSTRUMENT DURING THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THE OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19142167 ↗MalfunctionJun 26, 2020Peeled/Delaminated; Unraveled Material; Peeled/Delaminated; Unraveled Material
UPON REVIEW OF INITIAL DFAF, PERFORMED (B)(6) 2020, THE WIRE GUIDE WAS FOUND TO BE SEPARATED AND ELONGATED. THE SAFETY WIRE WAS ALSO EXPOSED. NO DELAMINATION WAS MENTIONED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10201746 ↗MalfunctionSep 21, 2018Structural Problem
THE COMPLAINANT REPORTED THAT THE PROCEDURE WAS FOR ILIAC STENTING FROM A CONTRALATERAL PUNCTURE SITE, BUT THE END OF THE AMPLATZ STIFF SUPPORT WIRE GUIDE WAS ¿VERY FLOPPY.¿ IT WAS FURTHER NOTED THAT THE ANCHOR SOLDER WAS SEPARATED FROM THE INNER MANDREL WIRE WHICH RESULTED IN COIL ELONGATION. THE PRODUCT PROBLEM DID NOT RESULT IN ANY REPORTED SERIOUS INJURIES TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7897948 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page