Amplatz G28781Recalls & reported problems

Cook Incorporated · Amplatz Stiff Wire Guide · Class 2

All models of Amplatz (42)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002287815), received between Sep 21, 2018 and Apr 19, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Unraveled Material” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unraveled Material2
  • Peeled/Delaminated1
  • Structural Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 19, 2024Unraveled Material; Unraveled Material

    AS REPORTED, DURING AN INFERIOR VENA CAVA (IVC) FILTER RETRIEVAL PROCEDURE, THE TIP OF AN AMPLATZ STIFF SUPPORT WIRE GUIDE UNRAVELED. ACCESS WAS OBTAINED AT THE INTERNAL JUGULAR (IJ) VEIN TO TARGET THE INFERIOR VENA CAVA (IVC). THE PATIENT'S ANATOMY WAS NOT STENOSED, TORTUOUS, CALCIFIED OR SCARRED. THE WIRE WAS NOT ALTERED FROM ITS ORIGINAL CONDITION PRIOR TO USE AND RESISTANCE WAS NOT FELT UPON INSERTION OF THE WIRE. UPON REMOVAL OF THE WIRE THROUGH AN UNSPECIFIED SHEATH, THE "DISTAL FLOPPING TIP OF THE WIRE CASING UNCOILED". THE WIRE WAS REMOVED FROM THE STERILE FIELD AND ANOTHER WIRE WAS USED TO COMPLETE THE PROCEDURE. THE WIRE WAS NOT MANIPULATED BACKWARDS THROUGH A NEEDLE OR OTHER METAL INSTRUMENT DURING THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THE OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19142167 ↗
  • MalfunctionJun 26, 2020Peeled/Delaminated; Unraveled Material; Peeled/Delaminated; Unraveled Material

    UPON REVIEW OF INITIAL DFAF, PERFORMED (B)(6) 2020, THE WIRE GUIDE WAS FOUND TO BE SEPARATED AND ELONGATED. THE SAFETY WIRE WAS ALSO EXPOSED. NO DELAMINATION WAS MENTIONED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10201746 ↗
  • MalfunctionSep 21, 2018Structural Problem

    THE COMPLAINANT REPORTED THAT THE PROCEDURE WAS FOR ILIAC STENTING FROM A CONTRALATERAL PUNCTURE SITE, BUT THE END OF THE AMPLATZ STIFF SUPPORT WIRE GUIDE WAS ¿VERY FLOPPY.¿ IT WAS FURTHER NOTED THAT THE ANCHOR SOLDER WAS SEPARATED FROM THE INNER MANDREL WIRE WHICH RESULTED IN COIL ELONGATION. THE PRODUCT PROBLEM DID NOT RESULT IN ANY REPORTED SERIOUS INJURIES TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7897948 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →