Cook TPN G29464Recalls & reported problems
Cook Incorporated · Cook TPN, Double Lumen Catheter Set With 10Fr Peel-Away, Introducer Set · Class 2 · Implant
All models of Cook TPN (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002294646), received between Jun 2, 2016 and Aug 31, 2017 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Hole In Material” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Hole In Material3
- Bent1
- Break1
- Device Slipped1
- Fracture1
- Inadequacy of Device Shape and/or Size1
- Kinked1
- Leak/Splash1
Patient problems
- No Code Available2
- No Known Impact Or Consequence To Patient2
- No Patient Involvement1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2017Fracture; Fracture
THE TIP OF THE WIRE BROKE WHEN THE SURGEON WAS TAKING THE PRODUCT OUT OF THE PACKAGING . PRIOR TO PATIENT CONTACT. IT WAS REPORTED BY THE INTERNATIONAL CUSTOMER THAT, DURING THE REMOVAL OF THE PRODUCT FROM THE PACKAGING, THE TIP OF THE WIRE BROKE. THE CUSTOMER HAS CONFIRMED THAT THIS WAS PRIOR TO PATIENT CONTACT; ACCORDINGLY, NO ADVERSE EVENTS WERE EXPERIENCED BY THE PATIENT.
Patient problems: No Patient Involvement
MAUDE report 6834637 ↗MalfunctionAug 3, 2017Bent; Break; Kinked; Inadequacy of Device Shape and/or Size; Device Slipped; Unraveled Material; Physical Resistance
THE CUSTOMER REPORTED THAT "THE WIRE WAS VERY SHORT, MAKING IT VERY EASY TO DISPLACE AND NEARLY CAUSING IT TO SLIP OUT OF THE PUNCTURE POINT". IN ADDITION, THE CUSTOMER REPORTED THAT "THE MATERIAL OF THE WIRE WAS VERY FRAGILE, RESULTING IN KINKING AND UNRAVELING DURING INSERTION, AND THAT THE TIP OF THE DILATOR WAS BENDING AND BREAKING WHENEVER RESISTANCE WAS ENCOUNTERED". NO PIECES OF THE DEVICE WERE LEFT IN THE PATIENT, AND THERE WERE NO INJURIES OR ADDITIONAL MEDICAL INTERVENTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6764909 ↗MalfunctionMay 30, 2017Hole In Material; Hole In Material
INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT RETURNED AT AN UNSPECIFIED TIME FOLLOWING INITIAL IMPLANT FOR A CATHETER EXCHANGE DUE TO A HOLE FOUND IN THE LINE (1820334-2017-01283.) THE PHYSICIAN EXCHANGED THE CATHETER AND OBSERVED A SIMILAR HOLE ON THE REPLACEMENT DEVICE (1820334-2017-01266.) THERE ARE NO REPORTED ADVERSE PATIENT CONSEQUENCES RELATED TO THIS EVENT. NO UNINTENDED SECTION OF THE DEVICE REMAINED WITHIN THE PATIENT'S BODY. THE DEVICE WAS RECEIVED BY THE MANUFACTURER FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6597783 ↗InjuryMay 30, 2017Hole In Material; Hole In Material
INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT HAD A REDO DOUBLE LUMEN TPN CATHETER IMPLANTED. THE PATIENT RETURNED AT AN UNSPECIFIED TIME FOLLOWING INITIAL IMPLANT FOR A CATHETER EXCHANGE DUE TO A HOLE FOUND IN THE LINE. THE PHYSICIAN EXCHANGED THE CATHETER AND OBSERVED A SIMILAR HOLE ON THE REPLACEMENT DEVICE (1820334-2017-01266). THERE ARE NO REPORTED ADVERSE PATIENT CONSEQUENCES RELATED TO THIS EVENT. NO UNINTENDED SECTION OF THE DEVICE REMAINED WITHIN THE PATIENT'S BODY. THE DEVICE WAS RECEIVED BY THE MANUFACTURER FOR EVALUATION.
Patient problems: No Code Available
MAUDE report 6597760 ↗InjuryJun 2, 2016Hole In Material; Leak/Splash; Hole In Material; Leak/Splash
DURING THE FLUSHING OF THE LINE THE PATIENT FELT A POP FOLLOWED BY SOME PAIN IN THE NECK. A LINOGRAM SHOWED LEAKING OF CONTRAST FROM THE LINE DUE TO A SMALL HOLE IN THE LINE. THE LINE WAS REMOVED AND A PICC LINE INSERTED INSTEAD. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 5694366 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page