Melker G30488Recalls & reported problems
Cook Incorporated · Melker Cuffed Emergency Cricothyrotomy Catheter Set · Class 2
All models of Melker (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002304888), received between Mar 22, 2018 and Mar 22, 2018 — classified as 1 death.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected1
Patient problems
- Death1
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathMar 22, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
THE CUSTOMER REPORTED THAT OXYGEN COULD BE DELIVERED TO THE PATIENT THROUGH THE MELKER EMERGENCY CUFFED CRICOTHYROTOMY CATHETER, BUT IT COULD NOT BE REMOVED THROUGH THE DEVICE. THE APERTURE OF THE CATHETER WAS REPORTEDLY SO SMALL THAT VENTILATION COULD NOT BE ACHIEVED, ALTHOUGH OXYGENATION DID OCCUR BECAUSE THE AIR COULD BE FORCED IN. AUTO-PEEP (POSITIVE END EXPIRATORY PRESSURE) RESULTANTLY DEVELOPED. THE PATIENT WAS TAKEN BACK TO THE OR TO EXCHANGE THE DEVICE EMERGENTLY FOR A TRACH, BUT ULTIMATELY THE PATIENT EXPIRED. THE MANUFACTURER GLOBAL PRODUCT MANAGER HAS ATTEMPTED TO GAIN ADDITIONAL INFORMATION ON 16MAR2018 AND WAS REDIRECTED TO THE USER FACILITY RISK MANAGEMENT DEPARTMENT. THE MANUFACTURER GLOBAL PRODUCT MANAGER WAS ABLE TO SPEAK WITH THE NURSE PRESENT FOR THE PROCEDURE ON 19MAR2018, SHE DID NOT PROVIDE ADDITIONAL INFORMATION BUT HAS INDICATED SHE WOULD HAVE THE PHYSICIAN RETURN THE MANUFACTURER'S CALL. THE CUSTOMER HAS INDICATED THAT THEY ARE UNWILLING TO PROVIDE PATIENT-RELATED INFORMATION. THE DEVICE IS REPORTEDLY UNAVAILABLE FOR RETURN AND EVALUATION.
Patient problems: Death; No Code Available
MAUDE report 7362896 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.