Beacon G31216Recalls & reported problems
Cook Incorporated · Beacon Tip Centimeter Sizing Catheter · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002312166), received between Mar 30, 2017 and Mar 30, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Material Separation” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K173289); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Apr 2, 2026Open, Classified
Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
Z-1936-2026 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Material Separation1
Patient problems
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 30, 2017Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM (AAA) WITH THE COOK ALPHA ABDOMINAL SYSTEM USING THE BEACON TIP AUROUS CENTIMETER VESSEL SIZING CATHETER X. THE PROCEDURE WAS COMPLETED ACCORDING TO PLAN. DURING THE POST-OP CT, SOMETHING STRANGE WAS OBSERVED INSIDE THE LUMEN OF THE STENTGRAFT. IT CAN BE SEEN IN ALL PROJECTIONS BUT SAGITTAL AND AXILLARY ARE MOST CLEAR. IT STARTS AT THE UPPER LEVEL OF THE BARE STENT (ZIMB) AND GOES INSIDE THE LUMEN DOWNWARDS. IT BIFURCATES AND CONTINUES DOWN EACH COMMON ILIAC. IT WAS DETERMINED THAT A FRAGMENT OF THE DEVICE, THE PEEL AWAY SHEATH, WAS LEFT IN THE PATIENT. IT WAS REPORTED THAT THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS AT THE TIME OF THE PROCEDURE, HOWEVER THE PATIENT IS GOING TO REQUIRE AN ADDITIONAL PROCEDURE TO REMOVE THE DEVICE PORTION. THIS WAS NOT IDENTIFIED AT THE TIME OF THE PROCEDURE. THE PLAN IS NOW TO TAKE THE PATIENT BACK TO THE HOSPITAL AND SEE IF THEY CAN VISUALIZE AND POSSIBLY EXTRACT IT. THIS IS STILL IN PLANNING AND AT THE TIME OF THIS REPORT, NO SECONDARY PROCEDURE HAS YET BEEN SCHEDULED OR COMPLETED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6446475 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K173289. One submission often covers many models, so check the product and lot details in each record.
- Z-1902-2025 · May 15, 2025 · Open, ClassifiedBeacon Tip 5.0 Fr Angiographic catheters are designed to facilitate diagnostic and therapeutic procedures. Catheters are available in various configurations. Configurations include
Other Catheter, intravascular, diagnostic devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page