Skip to content

COOK G31590Recalls & reported problems

Cook Incorporated · FUSION OMNI ERCP CATHETER · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002315907), received between Oct 2, 2017 and Oct 2, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Protrusion/Extrusion” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Material Protrusion/Extrusion1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2017Material Protrusion/Extrusion; Material Protrusion/Extrusion

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A FUSION OMNI ERCP CATHETER. ZIPPING THE CATHETER WAS VERY DIFFICULT. THE CATHETER WAS DAMAGED. THE DEVICE WAS EVALUATED ON 09/07/2017. THE STIFFENING WIRE HAS PENETRATED OUT OF THE CATHETER NEAR THE DISTAL END OF THE DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6907277 ↗

Other CATHETER, CHOLANGIOGRAPHY devices

All CATHETER, CHOLANGIOGRAPHY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page