Skip to content

Ascend AQ G32825Recalls & reported problems

Cook Incorporated · Ascend, AQ, Ureteral Dilation Balloon Catheter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002328259), received between Sep 10, 2018 and Sep 10, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Inflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Inflation Problem1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2018Inflation Problem; Inflation Problem

    IT WAS REPORTED, DURING A URETERAL DILATATION PROCEDURE, THE DOCTOR PLACED THE BALLOON OF THE ASCEND® AQ® URETERAL BALLOON CATHETER OVER THE GUIDE WIRE, CONNECTED THE INFLATION DEVICE, AND FOUND THAT THE BALLOON DID NOT INFLATE UNDER FLUOROSCOPY. THE DOCTOR REMOVED BALLOON AND SAW THAT THE BALLOON HAS BEEN COMPROMISED AS IT WAS UNABLE TO INFLATE AT ALL. THE BALLOON WAS NOT PLACED THROUGH ANY EQUIPMENT, ONLY OVER THE GUIDE WIRE. ADDITIONAL INFORMATION WAS RECEIVED 21AUG2018: THE ISSUE WITH THE DEVICE WAS NOTICED WHILE INFLATING THE BALLOON. THE BALLOON BURST LONGITUDINALLY AND ANOTHER PRODUCT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7859700 ↗

Other DILATOR, CATHETER, URETERAL devices

All DILATOR, CATHETER, URETERAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page