Dolphin BT G36102Recalls & reported problems
Cook Incorporated · Dolphin BT Ciaglia Balloon-Assisted Tracheostomy Introducer Set · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002361027), received between Nov 16, 2017 and Aug 7, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Fluid/Blood Leak1
- Material Rupture1
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2018Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRACHEOSTOMY, A DOLPHIN BT CIAGLIA BALLOON-ASSISTED PERCUTANEOUS TRACHEOSTOMY INTRODUCER SET WAS USED. THE BALLOON OF THE DEVICE BURST WHILE INFLATING AT A PRESSURE OF LESS THAN 11 ATMOSPHERES (BALLOON BURST AT 10ATM) AND THEN BECAME DIFFICULT TO REMOVE. A NEW DOLPHIN SET WAS THEN USED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7757868 ↗MalfunctionNov 16, 2017Fluid/Blood Leak; Material Rupture; Fluid/Blood Leak; Material Rupture
THE INTERNATIONAL CUSTOMER REPORTED THAT, WHEN THE OPERATOR INCREASED THE SODIUM CHLORIDE (NACL) PRESSURE IN THE BALLOON TO APPROXIMATELY 11 PSI, THE BALLOON PORTION OF THE DOLPHIN BT CIAGLIA BALLOON-ASSISTED PERCUTANEOUS TRACHEOSTOMY INTRODUCER DEVICE RUPTURED AFTER APPROXIMATELY 15 SECONDS. NACL LEAKED OUT OF THE BALLOON. THE OPERATOR PROCEEDED TO INSERT THE TRACHEAL CANNULA AS ORIGINALLY INTENDED, AND IT OCCURRED WITH NO PROBLEMS. AS A RESULT, THE PROCEDURE OCCURRED WITH NO PATIENT ADVERSE EVENTS OR ADDITIONAL MEDICAL INTERVENTION ON THE FIRST TRY. THE DEVICE IS REPORTEDLY AVAILABLE FOR RETURN; HOWEVER, AS OF THE DATE OF THIS REPORT, NO DEVICE HAS YET BEEN RECEIVED FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7039657 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page