HiWire G36277Recalls & reported problems
Cook Incorporated · HiWire Hydrophilic Wire Guide · Class 2
All models of HiWire (16)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002362772), received between Jul 19, 2017 and Jul 19, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Separation1
Patient problems
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 19, 2017Material Separation; Material Separation
IT WAS REPORTED THAT A HIWIRE HYDROPHILIC WIRE GUIDE, WAS USED DURING AN UNSPECIFIED PROCEDURE. THE ACCESS SITE WAS REPORTED TO BE THE LEFT UPPER ARM (ARTERIAL LIMB OF AV GRAFT). THE HIWIRE WAS COMPLETELY SOAKED IN SALINE PRIOR TO BEING PLACED IN THE PATIENT. THE HIWIRE WAS THEN INSERTED THRU A 21G MICRONEEDLE AND WHEN REMOVING THE HIWIRE, SOME OF THE COATING CAME OFF AND BUNCHED UP ON THE WIRE. A SMALL PARTICLE OF THE COATING WAS LEFT REMAINING INSIDE THE PATIENT. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THERE WERE NO ADVERSE EFFECTS ON THE PATIENT REPORTED DUE TO THIS OCCURRENCE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6724414 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.