Torcon NB G47032Recalls & reported problems
Cook Incorporated · Torcon NB Advantage Catheter · Class 2
All models of Torcon NB (79)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002470323), received between May 29, 2025 and May 29, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING AN UPPER EXTREMITY SHOULDER EMBOLIZATION, THE HUB OF A TORCON NB ADVANTAGE ANGIOGRAPHIC CATHETER SEPARATED UPON THE SECOND POWER INJECTION AT 750 PSI. THE CATHETER HUB WAS FLUSHED WITH HEPARINIZED SALINE. ACCESS WAS OBTAINED IN THE FEMORAL ARTERY TO TARGET THE LEFT UPPER EXTREMITY, AND THE CATHETER WAS ADVANCED THROUGH ANOTHER MANUFACTURER¿S 5-FRENCH SHEATH. THE ANATOMY WAS NOT STENOSED, TORTUOUS, OR CALCIFIED, AND RESISTANCE WAS NOT ENCOUNTERED UPON INSERTION OR ADVANCEMENT OF THE CATHETER. THE PATIENT WAS REPORTEDLY RADIATED TWICE BEFORE THE USER NOTICED THAT THE HUB HAD SEPARATED. THE CATHETER WAS PULLED OUT OF THE PATIENT, AND ANOTHER CATHETER WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THE PATIENT WAS HOSPITALIZED FOR A PRE-SCHEDULED SURGERY AFTER THE EMBOLECTOMY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22105735 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page