Cook G47308Recalls & reported problems
Cook Incorporated · Centimeter Sizing Catheter · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002473089), received between Mar 26, 2026 and Mar 26, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Crack” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Apr 2, 2026Open, Classified
Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
Z-1935-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Crack2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 26, 2026Crack
IT WAS REPORTED THAT THE "WALL" OF AN AUROUS CENTIMETER SIZING CATHETER "CRACKED" BELOW THE HUB OF THE CATHETER DURING A TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) PROCEDURE. THE USER NOTED A "WEIRD RATING", WHEN THE PRESSURE MONITOR WAS TURNED ON. THE USER NOTED BODILY FLUID LEAKING, AND THE CRACK WAS IDENTIFIED. THE HUB WAS NOT WIPED WITH ANY CLEANING SOLUTION. A POWER INJECTOR WAS NOT USED; RATHER MANUAL HAND INJECTION WAS PERFORMED. THE PROCEDURE WAS COMPLETED BY USING A DIFFERENT UNSPECIFIED CATHETER. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24703520 ↗MalfunctionMar 26, 2026Crack
IT WAS REPORTED THAT THE "WALL" OF AN AUROUS CENTIMETER SIZING CATHETER "CRACKED" BELOW THE HUB OF THE CATHETER DURING AN IVC FILTER PLACEMENT PROCEDURE. THE CATHETER MADE PATIENT CONTACT. THE HUB WAS NOT WIPED WITH ANY CLEANING SOLUTION. THE CATHETER WAS CONNECTED TO AN UNSPECIFIED TUBING OR SYRINGE. AN UNSPECIFIED POWER INJECTOR WAS USED AT 900 PSI. THE CATHETER WAS NOTED TO BE LEAKING DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITHOUT ISSUE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24703413 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.