Captura G53638Recalls & reported problems

Cook Incorporated · Captura Serrated Large Forcep - Spike · Class 2

All models of Captura (37)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002536388), received between Apr 25, 2017 and Sep 27, 2018 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Operates Differently Than Expected2
  • Difficult to Open or Close1

Patient problems

  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 27, 2018Difficult to Open or Close; Difficult to Open or Close

    DURING A BIOPSY PROCEDURE, THE PHYSICIAN USED A COOK CAPTURA SERRATED LARGE FORCEP - SPIKE. THE HINGE BROKE ON THE BIOPSY CABLE [SIC]. THEY COULD NOT CLOSE IT, SO THEY HAD TO REMOVE IT FROM THE ENDOSCOPE WITH THE BIOPSY CABLE [CUPS] OPEN. THEY SUCCESSFULLY COMPLETED THE PROCEDURE WITH ANOTHER DEVICE OF THE SAME TYPE. THE PHOTO PROVIDED BY THE CUSTOMER SHOWS ONE (1) CUP IN THE CLOSED POSITION AND ONE (1) CUP IN THE OPEN POSITION. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7914013 ↗
  • MalfunctionJan 5, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    DURING AN GASTRODUODENOSCOPY WITH A BIOPSY IN THE DUODENUM, THE PHYSICIAN USED A COOK CAPTURA SERRATED LARGE FORCEP - SPIKE. THE FORCEPS WERE USED WITH A GASTROSCOPE IN THE DUODENUM WHERE THEY WANTED TO TAKE A BIOPSY. THE FORCEPS WERE OPENED, BUT COULD NOT BE CLOSED. SEVERAL ATTEMPTS WERE MADE TO CLOSE THE CUPS WITHOUT SUCCESS. THE PHYSICIAN HAD TWO OPTIONS TO REMOVE THE FORCEPS; EITHER PULL BACK THE GASTROSCOPE WITH THE OPENED FORCEPS OUTSIDE OF THE GASTROSCOPE, "WITH HIGH RISK OF PATIENT INJURY IN THE DUODENUM", OR TO PULL BACK THE OPENED FORCEPS INTO THE WORKING CHANNEL. THE FORCEPS WERE PULLED BACK INTO THE WORKING CHANNEL TO PREVENT PATIENT INJURY. THE GASTROSCOPE WAS THEN REMOVED. THERE WAS DAMAGE TO THE WORKING CHANNEL OF THE GASTROSCOPE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7169214 ↗
  • MalfunctionApr 25, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    DURING A GASTRIC BIOPSY PROCEDURE, THE PHYSICIAN USED A COOK CAPTURA SERRATED LARGE FORCEP - SPIKE. THE DEVICE PASSED THROUGH THE WORKING CHANNEL OF THE ENDOSCOPE WITHOUT PROBLEMS AND WAS OPENED, BUT WHEN DOCTOR TRIED TO CLOSE THE FORCEPS TO OBTAIN MUCOSAL BIOPSY THE DEVICE COULDN´T CLOSE. THEY NEEDED TO CUT THE SHEATH IN THE HANDLE LENGTH AND EXTRACT THE GASTROSCOPE OUT OF THE PATIENT IN ORDER TO RETRIEVE THE DEVICE FROM THE WORKING CHANNEL. ON 03/31/2017, AN EVALUATION RESPONSE LETTER WAS RECEIVED FROM THE APPROVED SUPPLIER. THE SUPPLIER EVALUATION FOUND THAT THE LINK WIRE TO DRIVE WIRE SOLDER JOINT WAS BROKEN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6520473 ↗

Other Endoscopic grasping/cutting instrument, non-powered, exempt devices

All Endoscopic grasping/cutting instrument, non-powered, exempt devices →