Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002541702), received between Oct 26, 2017 and May 15, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected2
- Failure to Align1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2018Failure to Align; Device Operates Differently Than Expected; Failure to Align; Device Operates Differently Than Expected
DURING A BIOPSY PROCEDURE, THE PHYSICIAN USED A COOK CAPTURA SERRATED MAX FORCEPS-SPIKE. ONE BRANCH [CUP] COULD NOT BE OPENED. NO MAXIMUM TENSION OR OPENING WAS POSSIBLE. THERE WAS NO REPORTABLE INFORMATION AT THAT TIME. THE DEVICE WAS EVALUATED ON 04/20/2018. THE CUPS WOULD OPEN AND CLOSE, HOWEVER UPON THE CUPS CLOSING, THE TEETH WOULD NOT COME TOGETHER FULLY. WHEN APPLYING EXCESSIVE PRESSURE TO THE HANDLE THE FORCEPS, THE TEETH WOULD COME CLOSE TOGETHER, BUT NEVER CLOSE FULLY. THE FORCEPS CUPS WAS VISUALLY INSPECTED USING MAGNIFICATION AND THE CUPS EXHIBIT A SMALL GAP BETWEEN THE TEETH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7516153 ↗MalfunctionOct 26, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
DURING A COLONOSCOPY, A PHYSICIAN USED A CAPTURA SERRATED MAX FORCEPS - SPIKE. THE USER WAS IN THE TRANSVERSE COLON. A POLYPECTOMY WAS ATTEMPTED TO BE PERFORMED. THE FORCEPS WAS OPENED TO GRAB THE POLYP, BUT THE FORCEPS WOULD NOT CLOSE. THE COMPLAINT DEVICE WAS REMOVED AND A NEW CAPTURA SERRATED MAX FORCEPS- SPIKE WAS OPENED AND USED TO COMPLETE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6979150 ↗
Other Endoscopic grasping/cutting instrument, non-powered, exempt devices
- Captura G56076Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt60 reports
- Captura G56133Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt45 reports
- Single Use Grasping Forceps FG-253SXOlympus Medical Systems Corp. · Endoscopic grasping/cutting instrument, non-powered, exempt27 reports
- Raptor 00711177US Endoscopy · Endoscopic grasping/cutting instrument, non-powered, exempt27 reports
- Captura G31576Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt19 reports
- Captura G56059Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt18 reports
- Captura G31577Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt17 reports
- Captura Pro G50697Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page