Captura G56074Recalls & reported problems
Cook Incorporated · Captura Biopsy Forceps With Spike · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00827002560741), received between May 17, 2016 and Sep 18, 2020 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected3
- Device Damaged by Another Device1
- Device Handling Problem1
- Difficult to Open or Close1
- Failure to Align1
- Improper or Incorrect Procedure or Method1
- Material Protrusion/Extrusion1
- Peeled/Delaminated1
Patient problems
- No Consequences Or Impact To Patient6
- Foreign Body In Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2020Device Damaged by Another Device; Difficult to Open or Close; Material Protrusion/Extrusion; Device Damaged by Another Device; Difficult to Open or Close; Material Protrusion/Extrusion
DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN SELECTED A COOK CAPTURA BIOPSY FORCEPS WITH SPIKE. AFTER OPENING THE FORCEPS, IT WAS NOT POSSIBLE TO CLOSE THE CUPS. THE HINGES WERE NOT ALIGNED WITH THE SHAFT OF THE FORCEPS, BUT WERE STICKING OUT. THE FORCEPS COULD NOT BE PULLED BACK INTO THE WORKING CHANNEL OF THE ENDOSCOPE WITHOUT DAMAGING THE ENDOSCOPE. THE COMPLETE ENDOSCOPE PLUS THE FORCEPS WERE TAKEN OUT CAREFULLY. THE OUTSIDE DISTAL PORTION OF THE FORCEPS WAS CUT OFF TO TAKE EVERYTHING OUT OF THE ENDOSCOPE A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10555604 ↗MalfunctionMar 19, 2019Peeled/Delaminated; Improper or Incorrect Procedure or Method; Device Handling Problem; Peeled/Delaminated; Improper or Incorrect Procedure or Method; Device Handling Problem
DURING AN ENDOSCOPY, THE PHYSICIAN USED A COOK CAPTURA BIOPSY FORCEPS WITH SPIKE. THE SHEATH CAME APART [FLAKED OFF] WHILE IT WAS BEING PUT DOWN THE ENDOSCOPE. A NEW DEVICE WAS OPENED AND USED TO COMPLETE THE PROCEDURE. CLARIFICATION WAS RECEIVED 07-MAR-2019 STATING THAT IT WAS PART OF THE GRAY SHEATH ON THE BIOPSY FORCEPS THAT FLAKED OFF. THE PIECE WAS RETRIEVED WITH THE BIOPSY FORCEPS. THE DEVICE WAS RECEIVED FOR EVALUATION AND THE SHEATH WAS FOUND TO BE CRACKED AND MISSING SOME PIECES. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient
MAUDE report 8434237 ↗MalfunctionJun 8, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE, THE PHYSICIAN USED A COOK CAPTURA BIOPSY FORCEPS WITH SPIKE. THE FORCEPS WERE ADVANCED DOWN THE ENDOSCOPE. THEY WERE OPENED BUT THEN WOULD NOT CLOSE. THE PHYSICIAN KEPT WORKING WITH THE FORCEPS AND WAS ABLE TO GET THEM CLOSED TO REMOVE FROM THE ENDOSCOPE. ANOTHER OF THE SAME DEVICE WAS USED TO CONTINUE THE PROCEDURE. THERE WAS NO HARM TO THE PATIENT. ON (B)(6) 2017, THE DEVICE WAS RECEIVED FOR EVALUATION. IT WAS FOUND THAT THE JAWS WOULD OPEN BUT WOULD NOT CLOSE AT ALL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6623430 ↗MalfunctionJun 2, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
IN PREPARATION FOR AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK CAPTURA BIOPSY FORCEPS WITH SPIKE. THE USER OPENED THE FORCEPS AND IT WOULD NOT CLOSE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6608524 ↗MalfunctionMay 2, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
DURING AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED THREE (3) COOK CAPTURA FORCEPS WITH SPIKE. THE FORCEPS BROKE WHEN THEY WERE OPENED AND WOULD NOT CLOSE. DURING THE BEGINNING OF THE PROCEDURE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6538864 ↗MalfunctionMay 17, 2016Failure to Align; Failure to Align
DURING THREE (3) SEPARATE COLONOSCOPY PROCEDURES, THE PHYSICIAN USED A COOK CAPTURA BIOPSY FORCEP WITH SPIKE. THE FORCEP WAS INTRODUCED INTO THE COLONOSCOPE AND WHEN THE ASSISTANT OPENED THE FORCEP, THE DISTAL METAL BIOPSY CUP DETACHED FROM THE SPRING COIL AND EXTENDED ON THE DRIVE WIRE. THE DEVICE WAS REMOVED WITH NO HARM TO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. FOUR (4) USED DEVICES FROM THE SAME LOT WERE RETURNED. IT WAS CONFIRMED ON 4/22/16 FROM THE SUPPLIER THAT ONE OF THE DEVICES RETURNED HAD MISALIGNED CUPS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5661205 ↗
Other Endoscopic grasping/cutting instrument, non-powered, exempt devices
- Captura G56076Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt60 reports
- Captura G56133Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt45 reports
- Single Use Grasping Forceps FG-253SXOlympus Medical Systems Corp. · Endoscopic grasping/cutting instrument, non-powered, exempt27 reports
- Raptor 00711177US Endoscopy · Endoscopic grasping/cutting instrument, non-powered, exempt27 reports
- Captura G31576Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt19 reports
- Captura G56059Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt18 reports
- Captura G31577Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt17 reports
- Captura Pro G50697Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page