Roadrunner UniGlide G56173Recalls & reported problems

Cook Incorporated · Roadrunner UniGlide Hydrophilic Wire Guide · Class 2

All models of Roadrunner UniGlide (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00827002561731), received between Apr 3, 2018 and Jul 18, 2024 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Flaked” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 19, 2022Class IIOpen, Classified

    Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

    Z-1713-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Flaked2
  • Contamination /Decontamination Problem1
  • Defective Device1
  • Device Contaminated During Manufacture or Shipping1
  • Material Separation1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient2
  • No Code Available1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 18, 2024Flaked; Flaked

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19777767 ↗
  • MalfunctionDec 18, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18351354 ↗
  • MalfunctionAug 22, 2023Flaked; Flaked

    AS ORIGINALLY REPORTED, THE LUBRICITY OF THREE ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDES WAS INSUFFICIENT, MAKING THE WIRES DIFFICULT TO ADVANCE THROUGH CATHETERS AFTER MULTIPLE SALINE FLUSHES. THE WIRES WERE NOT ALTERED PRIOR TO USE. THE WIRES WERE FLUSHED AND WIPED DOWN WITH NORMAL SALINE TO ACTIVATE THE HYDROPHILIC COATING AND WERE KEPT HYDRATED IN A BOWL WHEN NOT IN USE. THE WIRES WERE USED NO MORE THAN TWO TIMES EACH. LATEX-FREE GLOVES WERE WORN. PER THE CUSTOMER, NO FLAKING WAS NOTED ON ANY OF THE WIRES. NOTHING WAS LEFT IN THE PATIENT, NO ADDITIONAL PROCEDURES WERE NEEDED, AND THERE WERE NO ADVERSE EFFECTS TO THE PATIENT. UPON RETURN AND EVALUATION OF UNUSED DEVICES FROM THE CUSTOMER (FROM THE SAME LOT), FLAKING OF THE HYDROPHILIC COATING WAS NOTED ONE UNUSED/PRIOR-TO USE WIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17597076 ↗
  • InjuryJun 3, 2021Peeled/Delaminated; Peeled/Delaminated

    ADDITIONAL INFORMATION WAS RECEIVED 21JUL2021. ANOTHER MANUFACTURER'S ENTRY NEEDLE WAS USED FOR ACCESS AND THE COMPLAINT WIRE WAS REMOVED THROUGH THE ENTRY NEEDLE. LATEX GLOVES WERE WORN, AND THE HYDROPHILIC COATING WAS ACTIVATED USING SALINE. THE DEVICE, WHICH WAS ONLY USED ONE TIME, WAS KEPT HYDRATED WHEN NOT IN USE.

    Patient problems: Foreign Body In Patient

    MAUDE report 11929910 ↗
  • MalfunctionJan 21, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Patient Involvement

    MAUDE report 9611248 ↗
  • MalfunctionSep 18, 2018Defective Device

    THE COMPLAINANT REPORTED THAT THE HYDROPHILIC WIRE DISTAL TIP WAS UNEVENLY COATED. THE PRODUCT WAS NOT USED, AND THERE WAS NO REPORTED IMPACT TO THE PATIENT AS A RESULT OF THE PRODUCT ISSUE. IT WAS FURTHER NOTED THAT THE POLYMER MATERIAL LOCATED AT THE DISTAL TIP WAS PARTIALLY ELONGATED, GIVING THE APPEARANCE OF A DISTORTED DISTAL TIP.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7886213 ↗
  • InjuryApr 3, 2018Material Separation; Material Separation

    IT WAS REPORTED THAT DURING A FISTULAGRAM ON (B)(6) 2018, THE TIP OF THE ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE SHEARED OFF AND WAS RETAINED IN THE PATIENT. THE PATIENT WAS BROUGHT BACK ON (B)(6) 2018 FOR A SURGICAL FOREIGN BODY RETRIEVAL OF WIRE FRAGMENT FROM THE LEFT VENTRICLE AND PULMONARY OUTFLOW TRACK. THE TIP OF THE ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE WAS SAFELY REMOVED. NO OTHER INTERVENTIONS WERE REQUIRED AND THE PATIENT WAS DISCHARGED FROM THE HOSPITAL IN STABLE CONDITION.

    Patient problems: Foreign Body In Patient; No Code Available

    MAUDE report 7393421 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →