Hemospray G56572Recalls & reported problems

Cook Incorporated · HEMOSPRAY ENDOSCOPIC HEMOSTAT · Class 2

All models of Hemospray (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Cook Hemospray endoscopic hemostat is a powder spray used to stop bleeding during endoscopy. The 446 reports mostly describe the device failing to activate or release its powder, and some describe blockages and a section of the device not remaining inside the patient. The reports include 1 death, 63 injuries and 382 malfunctions. There are 3 open recalls, covering labeling design, nonconforming raw materials in some lots, and powder sticking to the scope.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
439
Deaths reported
1
Injuries reported
61
Recalls naming it
3

Summary

The FDA's MAUDE database contains 439 adverse event reports that list this exact device (DI 00827002565722), received between Jul 5, 2018 and Sep 29, 2026 — classified as 1 death, 61 injuries, 377 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (272 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 3 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 7 reportsApr 2024: 1 reportsMay 2024: 5 reportsJun 2024: 4 reportsJul 2024: 7 reportsAug 2024: 3 reportsSep 2024: 16 reportsOct 2024: 6 reportsNov 2024: 16 reportsDec 2024: 5 reportsJan 2025: 4 reportsFeb 2025: 5 reportsMar 2025: 5 reportsApr 2025: 4 reportsMay 2025: 6 reportsJun 2025: 6 reportsJul 2025: 7 reportsAug 2025: 10 reportsSep 2025: 5 reportsOct 2025: 4 reportsNov 2025: 5 reportsDec 2025: 12 reportsJan 2026: 9 reportsFeb 2026: 5 reportsMar 2026: 10 reportsApr 2026: 5 reportsMay 2026: 6 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 20, 2026Class IIOpen, Classified

    Due to issue with endoscope adherence and inability to spray powder. Firm will provide an updated Instruction for Use (IFU).

    Z-3177-2026 · Root cause: Labeling design · Recall details · FDA record ↗

  2. Aug 29, 2024Class IIOpen, Classified

    Specific lots of product have been manufactured with nonconforming raw materials.

    Z-0036-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  3. May 22, 2023Class IIOpen, Classified

    There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.

    Z-1978-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure272
  • Failure to Fire82
  • Obstruction of Flow69
  • Complete Blockage39
  • Fracture38
  • Device Handling Problem23
  • Improper or Incorrect Procedure or Method21
  • Material Twisted/Bent12

Patient problems

  • No Clinical Signs, Symptoms or Conditions248
  • No Consequences Or Impact To Patient99
  • Gastrointestinal Hemorrhage51
  • No Patient Involvement16
  • No Known Impact Or Consequence To Patient11
  • Device Embedded In Tissue or Plaque7
  • Pain7
  • Hemostasis5

Recent adverse event reports

The 25 most recent of 439 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Activation Failure

    During an unknown endoscopic hemostasis procedure, the physician used a Cook Hemospray Endoscopic Hemostat.\\It was reported that the Hemospray would not dispense \"[unable to spray - subject of report]\". \\At the time of this report, our attempts to collect additional information regarding patient outcome have been unsuccessful. While the complainant did not specify if the patient experienced any adverse effects or required additional medical procedures due to this event, the information able to be collected does not reasonably suggest the patient was adversely impacted. Additional attempts to gather this information will be made.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941349 ↗
  • InjurySep 29, 2026Obstruction of Flow; Activation Failure

    During an Endoscopic gastrointestinal hemostasis procedure, the physician used a Cook Hemospray Endoscopic Hemostat.\\It was reported that there was bleeding from the duodenum, and endoscopic clipping was performed.\Rebleeding occurred 2 to 3 days later (b)(6)2026), and Hemospray was intended to be used for hemostasis; however, the powder could not be dispensed \"[unable to spray]\".\As bleeding could not be controlled at the facility, the patient was urgently transferred to another hospital, where hemostasis was achieved by angiographic intervention.\\A section of the device did not remain inside the patient¿s body. The patient did require an additional procedure \"[angiographic intervention]\" due to this occurrence. Our attempts to collect additional information regarding patient outcome were unsuccessful.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 50941288 ↗
  • MalfunctionSep 29, 2026Failure to Disconnect

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936676 ↗
  • InjuryAug 28, 2026Material Twisted/Bent; Activation Failure

    During an endoscopic hemostasis procedure, the physician used a Cook Hemospray Endoscopic Hemostat. It was reported that the physician opened the device. The powder failed to deploy. The customer believes that the failure was due to the catheter being kinked in the scope.\\A section of the device did not remain inside the patient¿s body. The procedure was finished with a clip. According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 50507480 ↗
  • InjuryAug 12, 2026Activation Failure

    During an endoscopic gastrointestinal hemostasis procedure, the physician used a Cook Hemospray Endoscopic Hemostat. It was reported that the trigger button was pressed to deliver powder to the target site during endoscopic gastrointestinal hemostasis; however, no powder was emitted. The procedure was completed using an alternative product.A section of the device did not remain inside the patient¿s body. The patient did not require any additional procedures due to this occurrence. According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 50225648 ↗
  • MalfunctionAug 7, 2026Obstruction of Flow; Activation Failure

    During an endoscopic hemostasis procedure, the physician used a Cook Hemospray Endoscopic Hemostat. It was reported that during use, the device failed to dispense spray. A second device of the same model was opened and functioned as intended.A section of the device did not remain inside the patient¿s body. The patient did not require any additional procedures due to this occurrence. According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50179725 ↗
  • InjuryJun 5, 2026Complete Blockage; Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE IN THE DUODENUM, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE POWDER SPRAYING WAS INITIALLY SUCCESSFUL BUT BECAME IMPOSSIBLE DURING THE PROCEDURE. THE CATHETER WAS REPLACED DUE TO SUSPECTED CLOGGING; HOWEVER, SPRAYING STILL FAILED, AND HEMOSTASIS COULD NOT BE ACHIEVED. THE PROCEDURE WAS COMPLETED USING CLIPS AND HEMOSTATIC FORCEPS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. HEMOSTASIS WAS ACHIEVED USING AN ALTERNATE MODALITY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 25411056 ↗
  • MalfunctionJun 1, 2026Material Twisted/Bent; Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE FOR A BLEED IN THE DIVERTICULUM, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT WHILE APPLYING THE POWDER, THE POWDER WOULD NOT COME OUT. A KINK WAS NOTED IN THE CATHETER, AND THE MEDICAL STAFF UNKINKED IT. THEN AT THE LUER LOCK THERE WAS A BIG POOF OF POWDER, AND THE DEVICE RAN OUT OF CO2 [UNABLE TO SPRAY - SUBJECT OF REPORT]. THE SECOND CATHETER WAS NOT USED; THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25354681 ↗
  • MalfunctionMay 20, 2026Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE FOR A GI BLEED, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE TECH WENT TO ACTIVATE THE HEMOSPRAY AND IT DID NOT PROPEL THE POWDER [UNABLE TO SPRAY - SUBJECT OF REPORT]. THEN THEY WENT TO USE A SECOND HEMOSPRAY AND THAT ONE FAILED IN THE EXACT SAME WAY. THE PROCEDURE WAS FINISHED WITHOUT THE USE OF THE HEMOSPRAY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25233722 ↗
  • MalfunctionMay 19, 2026Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE FOR A GI BLEED, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT A HEMOSPRAY WAS OPENED TO USE WITH THE PHYSICIAN BUT WOULD NOT WORK [UNABLE TO SPRAY - SUBJECT OF REPORT]. THEY DOUBLE CHECKED TO MAKE SURE CONNECTIONS WERE CORRECT AND THEY WERE. A SECOND HEMOSPRAY WAS OPENED TO FINISH THE PROCEDURE A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25216408 ↗
  • MalfunctionMay 18, 2026Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT A RIGHT HEMICOLECTOMY WAS PERFORMED FOR TRANSVERSE COLON CANCER FOR A PATIENT ON (B)(6) 2026. ON (B)(6) 2026 POSTOPERATIVE BLEEDING WAS OBSERVED AT THE ANASTOMOTIC SITE. WHEN ATTEMPTING TO APPLY HEMOSPRAY, POWDER COULD NOT BE SPRAYED FROM THE JUNCTION BETWEEN THE HANDLE AND THE CATHETER. THE DEVICE WAS REPLACED WITH ANOTHER OF THE SAME SPECIFICATION, AND HEMOSTASIS WAS SUCCESSFULLY ACHIEVED A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25201996 ↗
  • MalfunctionMay 14, 2026Obstruction of Flow; Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE DEVICE WAS UNABLE TO OPERATE [UNABLE TO SPRAY - SUBJECT OF REPORT] DESPITE FOLLOWING THE INSTRUCTIONS FOR USE. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25176700 ↗
  • MalfunctionMay 14, 2026Obstruction of Flow; Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE DEVICE WAS UNABLE TO OPERATE [UNABLE TO SPRAY - SUBJECT OF REPORT] DESPITE FOLLOWING THE INSTRUCTIONS FOR USE. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25176004 ↗
  • InjuryMay 4, 2026Activation Failure

    DURING A GASTROSCOPY PROCEDURE IN THE STOMACH FOR A BLEEDING ULCER, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT WHEN THE TECHNICIAN PRESSED THE SPRAY BUTTON NO POWDER FIRED [UNABLE TO SPRAY - SUBJECT OF REPORT]. SCLEROTHERAPY WAS PERFORMED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. SCLEROTHERAPY WAS PERFORMED TO COMPLETE THE PROCEDURE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 25073472 ↗
  • MalfunctionApr 29, 2026Complete Blockage; Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE FOR A GI BLEED, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT WHEN SPRAYED THE HEMOSPRAY CLOGGED DOWN THE CATHETER, 2ND CATHETER ALSO DID NOT WORK, THE PROCEDURE WAS COMPLETED BY USING ANOTHER OF THE SAME DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25032527 ↗
  • InjuryApr 16, 2026Complete Blockage; Activation Failure

    DURING A HEMOSTASIS PROCEDURE FOR A BLEEDING GASTRIC ULCER, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO SPRAY USING THE FIRST CATHETER; HOWEVER, THE CATHETER BECAME OCCLUDED. A SECOND CATHETER WAS THEN USED, BUT NO POWDER WAS DISCHARGED, AND THE DEVICE WAS CONSIDERED MALFUNCTIONING. SPRAYING WAS NOT POSSIBLE FROM THE INITIAL TRIGGER ACTIVATION. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. APC [ARGON PLASMA COAGULATION] WAS PERFORMED, FOLLOWED BY APPLICATION OF PURASTAT, AND THE PROCEDURE WAS COMPLETED. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 24898723 ↗
  • MalfunctionApr 15, 2026Activation Failure

    DURING A HEMOSTASIS PROCEDURE FOR A GASTRIC BLEED, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE DEVICE WOULD NOT SPRAY AND THAT IT WAS AS IF THE BUTTON WOULDN'T WORK. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24889364 ↗
  • MalfunctionApr 11, 2026Complete Blockage; Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE IN THE DUODENUM, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE SPRAYING WAS PERFORMED USING THE FIRST CATHETER, BUT THE CATHETER BECAME CLOGGED. WHEN ATTEMPTING TO USE THE SECOND CATHETER AFTER SWITCHING, THE CO2 GAS HAD BEEN DEPLETED AND SPRAYING COULD NOT BE PERFORMED. SINCE THE CATHETER WAS NOT CLOGGED, GAS DEPLETION WAS CONFIRMED. UPON CHECKING THE REMAINING AMOUNT OF POWDER, APPROXIMATELY TWO THIRDS WAS STILL LEFT. AS HEMOSTASIS HAD ALREADY BEEN ACHIEVED WITH THE FIRST SPRAY, THE PROCEDURE WAS COMPLETED WITHOUT ADDITIONAL INTERVENTIONS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24852105 ↗
  • InjuryApr 11, 2026Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE FOR A DUODENAL ULCER, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT POWDER SPRAYING WAS ATTEMPTED AT THE TARGET SITE, NO POWDER WAS DISCHARGED. AS HEMOSTASIS REMAINED INSUFFICIENT, THE DEVICE WAS SWITCHED TO A CLIP, AND THE PROCEDURE WAS COMPLETED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. AN ALTERNATE MODALITY WAS USED TO COMPLETE THE PROCEDURE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 24846031 ↗
  • InjuryMar 30, 2026Obstruction of Flow; Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE FOR A RECTAL ULCER, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE HEMOSPRAY PRESSURE WAS WEAK FROM THE OUTSET. IN THE INITIAL PHASE, THE POWDER COULD BE SPRAYED DESPITE THE WEAK PRESSURE, BUT IT SUBSEQUENTLY CEASED TO BE DELIVERED. AS HEMOSTASIS REMAINED INSUFFICIENT, THE DEVICE WAS SWITCHED TO A NON-COOK PRODUCT (NEXPOWDER), AND THE PROCEDURE WAS COMPLETED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. AN ALTERNATE MODALITY WAS USED TO COMPLETE THE PROCEDURE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 24737345 ↗
  • MalfunctionMar 30, 2026Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE IN THE COLON, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE HEMOSPRAY WAS UNABLE TO SPRAY, POWDER DEPLOYMENT WAS UNSUCCESSFUL. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND ADEQUATE HEMOSTASIS WAS ACHIEVED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24735614 ↗
  • MalfunctionMar 19, 2026Obstruction of Flow; Activation Failure

    DURING A HEMOSTASIS PROCEDURE, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT AFTER ALL PRODUCT PREPARATION STEPS WERE COMPLETED, THE PHYSICIAN BEGAN SPRAYING HEMOSPRAY FOR HEMOSTASIS. AFTER APPROXIMATELY FIVE SPRAYS, IT WAS NOTICED THAT THE POWDER OUTPUT HAD DECREASED, IT WAS FOUND THAT THE AIR PRESSURE HAD BECOME SIGNIFICANTLY WEAK, AND THE DEVICE WAS NO LONGER ABLE TO SPRAY [UNABLE TO SPRAY-SUBJECT OF REPORT] A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24641654 ↗
  • MalfunctionMar 17, 2026Fracture

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE NURSE WAS TRYING TO ACTIVATE THE HEMOSPRAY DEVICE AND THE RED KNOB BROKE AND MADE A LOUD NOISE. THEY DESCRIBED A HAIRLINE CRACK IN THE KNOB, NO EVACUATION OF CO2 CANISTER. THE HEMOSPRAY WOULD NOT WORK AFTER THE RED KNOB BROKE. THEY OPENED ANOTHER HEMOSPRAY DEVICE TO FINISH THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24621695 ↗
  • MalfunctionMar 17, 2026Activation Failure

    DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE IN THE STOMACH, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE HEMOSPRAY WAS ABLE TO SPRAY AT FIRST; BUT SPRAYING BECAME IMPOSSIBLE DURING THE PROCEDURE. THE CATHETER WAS REPLACED; HOWEVER, SPRAYING COULD NOT BE ACHIEVED, AND HEMOSTASIS WAS NOT OBTAINED. THEREFORE, ARGON WAS USED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. ARGON WAS USED TO COMPLETE THE HEMOSTASIS PROCEDURE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 24617870 ↗
  • MalfunctionMar 11, 2026Activation Failure

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE WITH A STOMACH ULCER, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT THE HEMOSPRAY WOULD NOT SPRAY. CANISTER WAS PROPERLY ACTIVATED, CHANNEL DRIED, NO CLOGS, BUT SPRAY WOULD NOT WORK. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT ISSUE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24574841 ↗

FDA clearance and classification

The FDA regulates product code QAU (hemostatic device for endoscopic gastrointestinal use) under 21 CFR 878.4456 as a Class II device. 21 CFR 878.4456 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: Hemostatic device for intraluminal gastrointestinal use. A hemostatic device for intraluminal gastrointestinal use is a prescription device that is endoscopically applied to the upper and/or lower gastrointestinal tract and is intended to produce hemostasis via absorption of fluid or by other physical means.

Common questions about the Hemospray G56572

Has the Hemospray G56572 been recalled?

Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-3177-2026 (Class II), began on Aug 20, 2026: Due to issue with endoscope adherence and inability to spray powder. Firm will provide an updated Instruction for Use (IFU). If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Hemospray G56572?

The FDA's MAUDE database lists 439 adverse event reports for this device, received between Jul 5, 2018 and Sep 29, 2026. In those reports, the most common specific device problems are “Activation Failure” (272), “Failure to Fire” (82) and “Obstruction of Flow” (69); the most common patient problems are “Gastrointestinal Hemorrhage” (51), “Device Embedded In Tissue or Plaque” (7) and “Pain” (7). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Hemospray G56572?

1 report is classified as a death, 61 as injuries and 377 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Hemospray G56572, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Hemospray, model G56572). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Hemostatic device for endoscopic gastrointestinal use devices

All Hemostatic device for endoscopic gastrointestinal use devices →