Skip to content

HemosprayRecalls & reported problems across all models

Cook Incorporated · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

4 open recalls name at least one Hemospray model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
535
Deaths reported
2
Injuries reported
64
Recalls
4

We found 2 models of Hemospray by Cook Incorporated in the FDA's device registry (GUDID). Together they are named in 535 adverse event reports in the FDA's MAUDE database, received between Jun 28, 2018 and Sep 30, 2026 — classified as 2 deaths, 64 injuries, 469 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (334 reports).

4 FDA recall records name at least one of these models by its device identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 3 reportsDec 2023: 3 reportsJan 2024: 3 reportsFeb 2024: 2 reportsMar 2024: 7 reportsApr 2024: 1 reportsMay 2024: 9 reportsJun 2024: 4 reportsJul 2024: 7 reportsAug 2024: 3 reportsSep 2024: 18 reportsOct 2024: 7 reportsNov 2024: 19 reportsDec 2024: 6 reportsJan 2025: 6 reportsFeb 2025: 6 reportsMar 2025: 6 reportsApr 2025: 5 reportsMay 2025: 8 reportsJun 2025: 7 reportsJul 2025: 8 reportsAug 2025: 11 reportsSep 2025: 6 reportsOct 2025: 4 reportsNov 2025: 5 reportsDec 2025: 12 reportsJan 2026: 10 reportsFeb 2026: 5 reportsMar 2026: 10 reportsApr 2026: 6 reportsMay 2026: 8 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 14 reportsOct 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

Hemospray recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 20, 2026Open, Classified

    Due to issue with endoscope adherence and inability to spray powder. Firm will provide an updated Instruction for Use (IFU).

    Z-3177-2026 · Root cause: Labeling design · names 2 of these models · Recall details · FDA record ↗

  2. Aug 29, 2024Open, Classified

    Specific lots of product have been manufactured with nonconforming raw materials.

    Z-0036-2025 · Root cause: Nonconforming Material/Component · names 2 of these models · Recall details · FDA record ↗

  3. May 22, 2023Open, Classified

    There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.

    Z-1979-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. May 22, 2023Open, Classified

    There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.

    Z-1978-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported Hemospray problems

Device problems

  • Activation Failure334
  • Failure to Fire103
  • Obstruction of Flow81
  • Complete Blockage48
  • Fracture46
  • Improper or Incorrect Procedure or Method26
  • Device Handling Problem25
  • Material Twisted/Bent11

Patient problems

  • No Clinical Signs, Symptoms or Conditions308
  • No Consequences Or Impact To Patient124
  • Gastrointestinal Hemorrhage55
  • No Patient Involvement23
  • No Known Impact Or Consequence To Patient13
  • Pain9
  • Device Embedded In Tissue or Plaque7
  • Hemostasis5

Hemospray models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Hemospray adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Cook Incorporated products

All Cook Incorporated devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page